510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Aesculap Minop System

K983365510(k) TraditionalCleared Dec 18, 199885-day reviewClass II21 CFR 882.1480Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aesculap, Inc. · GWG Endoscope, Neurological

Cleared on Dec 18, 1998 after an 85-day review. The FDA summary names its predicate without a number. 12 later devices descend from it. It has been recalled 3 times.

Decided
Dec 18, 1998
Review
85 days
Class
Class II
Product code
GWG
Recalls
3
Predicates
0

12 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2002: 1 descendant cleared2019: 2 descendants cleared2014: 1 descendant cleared2017: 1 descendant cleared2016: 1 descendant cleared2018: 1 descendant cleared2020: 2 descendants cleared2023: 2 descendants cleared2025: 1 descendant cleared1995200020052010201520202025
  1. This deviceK983365 Aesculap Minop SystemAesculap, Inc. · 1998
  2. Descendant · cites it directly · 5 of its ownK132931 Apollo SystemPenumbra, Inc. · 2014
  3. Descendant · cites it directly · 4 of its ownK182211 Aurora Surgiscope SystemRebound Therapeutics Corporation · 2019
  4. Descendant · 2 generations on · 4 of its ownK152699 Apollo SystemPenumbra, Inc. · 2016
  5. Descendant · 2 generations on · 3 of its ownK191861 Aurora Surgiscope SystemRebound Therapeutics · 2019
  6. Descendant · 3 generations on · 2 of its ownK201840 Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System…Rebound Therapeutics · 2020 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2002: 12014: 12016: 12017: 12018: 12019: 22020: 22023: 22025: 1
2002high 2 (2019) · low 0 (2003)2026

Most influential descendants

Showing 5 of 12 · See all
  • K132931Apollo SystemPenumbra, Inc. · 2014 · cites it directly5descendants
  • K182211Aurora Surgiscope SystemRebound Therapeutics Corporation · 2019 · cites it directly4descendants
  • K152699Apollo SystemPenumbra, Inc. · 2016 · 2 generations on4descendants
  • K191861Aurora Surgiscope SystemRebound Therapeutics · 2019 · 2 generations on3descendants
  • K201840Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box)Rebound Therapeutics · 2020 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K983365Aesculap Minop SystemAesculap, Inc.1998—
+1K132931Apollo SystemPenumbra, Inc.2014—
+1K182211Aurora Surgiscope SystemRebound Therapeutics Corporation2019—
+2K152699Apollo SystemPenumbra, Inc.2016—
+2K191861Aurora Surgiscope SystemRebound Therapeutics2019—
+3K201840Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control…Rebound Therapeutics20201

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
201410
201610
201710
201810
201920
202020
202320
202510
FIG. 1 — Lineage of K983365, 1998–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWG device

  1. #1 K211202 1688 4K Camera System, L11 LED Light Source…score 22.5 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Aesculap Ventriculopscope
Named without a submission number
—
*—Aesculap Angled Neuroendoscopes
Named without a submission number
—
*—Codman Neuroendoscopes
Named without a submission number
—
*—Neuro Navigational Neuroview
Named without a submission number
—
*—Richard Wolf Neuro-Endoscope
Named without a submission number
—

What descends from it? (12)

See all 12

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132931Apollo SystemPenumbra, Inc.20145
K182211Aurora Surgiscope SystemRebound Therapeutics Corporation20194
K014149Paediscope, Model # P010A and PF011AAesculap, Inc.20020

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 15, 2026Class IIZ-1485-2026Under Investigation by firmThere is the potential for the length of the trocar shaft to be too long.
Jul 20, 2023Class IIZ-2518-2023Process change controlTrocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30¿ neuroendoscope, a portion of the visual field is obscured.
Mar 7, 2017Class IIZ-1814-2017Nonconforming Material/ComponentAesculap Implant Systems LLC is recalling the Minop Trocar due to the possibility it may have sharp edges on the distal end which may lead to the abrasion of the insulation when removing the electrode.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K132931 (Penumbra, Inc., 2014) has gone furthest, with 5 descendants of its own.

How does its category compare?

All GWG devices

GWG Endoscope, Neurological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GWG

2in 2025

2008: 22013: 12014: 22016: 22017: 42018: 42019: 32020: 22021: 32023: 52024: 12025: 22026 (partial year): 4
2007high 5 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWG

269in 2023

2017: 126 days median2018: 128 days median2019: 156 days median2021: 106 days median2023: 269 days median2026 (partial year): 217 days median
2007high 269 (2023) · low 106 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWG

2in 2011

2009: 12011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWG

33in 2025

2007: 32008: 52010: 12011: 42012: 72013: 32014: 192015: 372016: 252017: 122018: 162019: 102020: 192021: 212022: 122023: 92024: 92025: 33
2007high 37 (2015) · low 0 (2009)2025

MAUDE reports naming product code GWG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—03
20082—05
20090—10
20100—01
20110—24
20120—07
20131—03
20142—019
20150—037
20162—025
20174126012
20184128016
20193156010
20202—019
20213106021
20220—012
2023526909
20241—09
20252—033
202642170—

What is it indicated for?

Aesculap’s MINOP System is indicated for use in endoscope-assisted microneurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for direct visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K983365: Aesculap Minop System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K983365

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026