Cleared on Apr 25, 2000 after a 196-day review. The FDA summary names its predicate without a number. It has been recalled once.
- Decided
- Apr 25, 2000
- Review
- 196 days
- Class
- Class II
- Product code
- BSZ
- Recalls
- 1
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in BSZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BSZ device
- #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Datex-Ohmeda Aestiva 3000 Anesthesia System Named without a submission number | |
| * | — | Ohmeda Excel MRI Anesthesia System Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 16, 2004 | Not classified | Z-0969-04 | Other | Increased circuit pressure as a result of a misassembled Aestiva/5 Active Gas Scavenging System. |
No ancestor or descendant has a Class I recall.
How does its category compare?
All BSZ devicesBSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · BSZ
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · BSZ
145in 2017
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BSZ
5in 2025
Class I (38 since 2007) Class II/III not classified
Adverse event reports per year · BSZ
9,198in 2025
MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 175 | 1 | 132 |
| 2008 | 8 | 99 | 8 | 104 |
| 2009 | 3 | 34 | 3 | 85 |
| 2010 | 1 | — | 14 | 97 |
| 2011 | 6 | 218 | 1 | 87 |
| 2012 | 5 | 134 | 7 (6) | 89 |
| 2013 | 5 | 125 | 5 (5) | 308 |
| 2014 | 4 | 235 | 0 | 279 |
| 2015 | 4 | 134 | 7 (2) | 572 |
| 2016 | 1 | — | 1 | 2,135 |
| 2017 | 3 | 145 | 5 (1) | 2,355 |
| 2018 | 2 | — | 3 | 2,589 |
| 2019 | 0 | — | 5 (2) | 1,957 |
| 2020 | 1 | — | 7 | 2,687 |
| 2021 | 1 | — | 2 (1) | 3,123 |
| 2022 | 0 | — | 4 (3) | 4,352 |
| 2023 | 2 | — | 6 | 5,595 |
| 2024 | 0 | — | 13 (13) | 6,067 |
| 2025 | 1 | — | 5 (3) | 9,198 |
| 2026 | 2 | — | 7 (2) | — |
What is it indicated for?
The Aestiva SmarVent MRI provides the functional feature set offered by the conventional Aestiva 3000 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva 3000 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva SmartVent MRI performed to specifications when tested directly next to a 1.5 Tesla shielded MRI device in a field strength that did not exceed 300 gauss.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K993410 · printed Nov 6, 2026
Cite this page
510(k) Report. “K993410: Dates-Ohmeda Aestiva Smartvent Mri.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993410
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026