510(k) · Cleared
Recalled · 2010X-Coated Capiox SX18 and SX25 Hollow Fiber Oxygenator with/ Without Detachable Hardshell Reservoir
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Terumo Medical Corp. · DTZ Oxygenator, Cardiopulmonary Bypass
Cleared on Jun 5, 2000 after a 210-day review. The FDA summary names no predicate by number. 16 later devices descend from it. It has been recalled once.
- Decided
- Jun 5, 2000
- Review
- 210 days
- Class
- Class II
- Product code
- DTZ
- Recalls
- 1
- Predicates
- 0
16 later devices descend from it, 3 citing it directly.
- This deviceK993772 X-Coated Capiox SX18 and SX25 Hollow Fiber Oxygenator with/ Without…Terumo Medical Corp. · 2000
- Descendant · cites it directly · 12 of its ownK040210 Capiox RX25 Hollow Fiber Oxygenator W/WO Hardshell ReservoirTerumo Corp. · 2004
- Descendant · cites it directly · 9 of its ownK062381 Capiox RX Oxygenator with or Without ReservoirTerumo Cardiovascular Systems Corp. · 2006 · Recalled
- Descendant · 2 generations on · 7 of its ownK130333 Capix RX Hollow Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp. · 2013
- Descendant · 2 generations on · 7 of its ownK071494 Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp. · 2007 · Recalled
- Descendant · 3 generations on · 6 of its ownK130520 Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp. · 2013
Showing 5 of 16 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 16 · See all- K040210Capiox RX25 Hollow Fiber Oxygenator W/WO Hardshell ReservoirTerumo Corp. · 2004 · cites it directly12descendants
- K062381Capiox RX Oxygenator with or Without ReservoirTerumo Cardiovascular Systems Corp. · 2006 · cites it directly9descendants
- K130333Capix RX Hollow Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp. · 2013 · 2 generations on7descendants
- K071494Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp. · 2007 · 2 generations on7descendants
- K130520Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp. · 2013 · 3 generations on6descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K993772 | X-Coated Capiox SX18 and SX25 Hollow Fiber Oxygenator with/ Without Detachable… | Terumo Medical Corp. | 2000 | — |
| +1 | K040210 | Capiox RX25 Hollow Fiber Oxygenator W/WO Hardshell Reservoir | Terumo Corp. | 2004 | — |
| +1 | K062381 | Capiox RX Oxygenator with or Without Reservoir | Terumo Cardiovascular Systems Corp. | 2006 | 2 |
| +2 | K130333 | Capix RX Hollow Fiber Oxygenator with/Without Hardshell Reservoir | Terumo Cardiovascular Systems Corp. | 2013 | — |
| +2 | K071494 | Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial Filter | Terumo Corp. | 2007 | 12 |
| +3 | K130520 | Capiox FX15 and FX25 Hollow Fiber Oxygenator/Reservoir | Ashitaka Factory of Terumo Corp. | 2013 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
| 2009 | 1 | 0 |
| 2013 | 3 | 0 |
| 2014 | 1 | 0 |
| 2015 | 5 | 0 |
| 2018 | 2 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DTZ device
- #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
The FDA summary names no predicate by number.
What descends from it? (16)
See all 163 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 28, 2010 | Not classified | Z-0842-2012 | Employee error | Complaints were received of expired X-coated Capiox SX18 Hollow Fiber oxygenators. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K040210 (Terumo Corp., 2004) has gone furthest, with 12 descendants of its own.
How does its category compare?
All DTZ devicesDTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DTZ
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DTZ
124in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DTZ
1in 2025
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DTZ
976in 2025
MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 55 | 7 | 134 |
| 2008 | 3 | 88 | 5 | 105 |
| 2009 | 9 | 210 | 0 | 52 |
| 2010 | 2 | — | 1 | 119 |
| 2011 | 4 | 160 | 13 | 190 |
| 2012 | 9 | 90 | 14 | 215 |
| 2013 | 9 | 28 | 0 | 346 |
| 2014 | 5 | 28 | 2 | 764 |
| 2015 | 9 | 134 | 0 | 658 |
| 2016 | 3 | 62 | 1 | 443 |
| 2017 | 2 | — | 0 | 388 |
| 2018 | 3 | 72 | 2 (2) | 320 |
| 2019 | 5 | 90 | 1 | 489 |
| 2020 | 5 | 93 | 3 | 442 |
| 2021 | 5 | 336 | 1 (1) | 595 |
| 2022 | 2 | — | 6 | 559 |
| 2023 | 5 | 145 | 0 | 738 |
| 2024 | 3 | 124 | 0 | 535 |
| 2025 | 2 | — | 1 | 976 |
| 2026 | 3 | 137 | 0 | — |
What is it indicated for?
The X-Coated CAPIOX® SX25 and SX18 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs are used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery for periods up to 6 hours. The integral heat exchanger is used to warm or cool blood or perfusion fluid as it flows through the device. The (detachable) hardshell reservoir is used to store blood…
Read the full statement (151 words)
The X-Coated CAPIOX® SX25 and SX18 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs are used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery for periods up to 6 hours.
The integral heat exchanger is used to warm or cool blood or perfusion fluid as it flows through the device.
The (detachable) hardshell reservoir is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal.
The X-Coat is a polymer coating that is applied to blood contacting surfaces of the oxygenator to reduce the adhesion of platelets to the surfaces of the
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K993772 · printed Nov 6, 2026
Cite this page
510(k) Report. “K993772: X-Coated Capiox SX18 and SX25 Hollow Fiber Oxygenator with/ Without Detachable Hardshell Reservoir.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993772
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026