510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Accuflow and Accuflux Elastomeric Infusion Pump

K081905510(k) TraditionalCleared Jul 15, 200812-day reviewClass II21 CFR 880.5725Recalls 7 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Westmed, Inc. · MEB Pump, Infusion, Elastomeric

Cleared on Jul 15, 2008 as substantially equivalent to K050770 (Mckinley Medical, LLC, 2005) and 3 other predicates after a 12-day review. Its family tree reaches back 10 generations to 1990. 1 of the 48 devices in its family, counting itself, has a Class I recall. 4 later devices descend from it. It has been recalled 7 times, including a Class I recall.

Decided
Jul 15, 2008
Review
12 days
Class
Class II
Product code
MEB
Recalls
7 (1 Class I)
Predicates
4

Its longest line of descent runs back 10 generations to K896422 (1990). 43 ancestors in all, the oldest K802820 (1980). 4 later devices descend from it, 1 citing it directly. 1 of the 48 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor2002: 1 other ancestor2001: 5 other ancestors1998: 2 other ancestors1991: 1 other ancestor1987: 1 other ancestor1985: 2 other ancestors1984: 1 other ancestor1980: 1 other ancestor2003: 1 other ancestor2000: 1 other ancestor1999: 6 other ancestors1996: 1 other ancestor1993: 2 other ancestors1989: 2 other ancestors1995: 2 other ancestors1990: 1 other ancestor2016: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant clearedK896422 · Burron Ambulatory Drug Delivery System · Burron Medical, Inc. · 1990×3: 3 generations cleared close together, K980558 (1998), K982946 (1998), K984502 (1999) K020862 · I-Flow Elastomeric Pump · I-Flow Corp. · 2002 · recalledK023098 · Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion Kit · Mckinley Infuser · 2002K032642 · Beeline System · Mckinley, Inc. · 2003×2: 2 generations cleared close together, K042228 (2004), K050770 (2005) K052451 · MPS Acacia Medflo Elastomeric Pump · Mps Acacia · 200519801990200020102020
  1. Root · 10 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 6 generations between
  3. 3 generations backK042228 Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004
  4. 2 generations backK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005
  5. 1 generation backK052451 MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005
  6. This deviceK081905 Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008
  7. Descendant · cites it directly · 3 of its ownK151650 SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd. · 2016
  8. Descendant · 2 generations on · 0 of its ownK250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd. · 2026
  9. Descendant · 2 generations on · 0 of its ownK240624 InfuLifeFirst Medical Source, LLC · 2024
  10. Descendant · 2 generations on · 0 of its ownK242152 SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd. · 2024
Show all 10 generations
  1. Root · 10 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 9 generations backK980558 Painbuster Infusion SystemI-Flow Corp. · 1998
  3. 8 generations backK982946 Painbuster Infusion KitI-Flow Corp. · 1998
  4. 7 generations backK984502 Nerve Block Infusion KitI-Flow Corp. · 1999
  5. 6 generations backK020862 I-Flow Elastomeric PumpI-Flow Corp. · 2002 · Recalled
  6. 5 generations backK023098 Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser…Mckinley Infuser · 2002
  7. 4 generations backK032642 Beeline SystemMckinley, Inc. · 2003
  8. 3 generations backK042228 Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004
  9. 2 generations backK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005
  10. 1 generation backK052451 MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005
  11. This deviceK081905 Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008

Most influential descendants

Showing 4 of 4 · See all
  • K151650SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd. · 2016 · cites it directly3descendants
  • K250234SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd. · 2026 · 2 generations on0descendants
  • K240624InfuLifeFirst Medical Source, LLC · 2024 · 2 generations on0descendants
  • K242152SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K896422Burron Ambulatory Drug Delivery SystemBurron Medical, Inc.1990—
−9K980558Painbuster Infusion SystemI-Flow Corp.1998—
−8K982946Painbuster Infusion KitI-Flow Corp.1998—
−7K984502Nerve Block Infusion KitI-Flow Corp.1999—
−6K020862I-Flow Elastomeric PumpI-Flow Corp.20021
−5K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser2002—
−4K032642Beeline SystemMckinley, Inc.2003—
−3K042228Beeline Motiv, Propolis, PcaMckinley, Inc.2004—
−2K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC2005—
−1K052451MPS Acacia Medflo Elastomeric PumpMps Acacia2005—
0 (this device)K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc.2008—
+1K151650SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd.2016—
+2K250234SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd.2026—
+2K240624InfuLifeFirst Medical Source, LLC2024—
+2K242152SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
198520
198710
198920
199010
199110
199320
199520
199610
199820
199960
200010
200150
200210
200310
200410
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
202420
202610
FIG. 1 — Lineage of K081905, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEB device

  1. #1 K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)score 14.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050770Modification to Accufuser, Accufuser Plus Standard Procedure Kit
Summary page 1
…rically Powered Disposable Infusion Pump Classification Name: Elastomeric External Infusion Pump (21 CFR 880.5725) 80 MEB Predicate Device: McKinley Accufuser K050770 cleared April 14, 2005 Baxer Infuser K062457 cleared September 21, 2006…
Mckinley Medical, LLC2005—
*K062457Infusor Small Volume Elastomeric Infusion Devices
Summary page 1
…Infusion Pump (21 CFR 880.5725) 80 MEB Predicate Device: McKinley Accufuser K050770 cleared April 14, 2005 Baxer Infuser K062457 cleared September 21, 2006 MPS MedFIo K052451 cleared December 8, 2005…
Baxter Healthcare Corporation20062
*K052451MPS Acacia Medflo Elastomeric Pump
Summary page 1
…770 cleared April 14, 2005 Baxer Infuser K062457 cleared September 21, 2006 MPS MedFIo K052451 cleared December 8, 2005 I-Flow K052117 cleared September 9, 2005 Device D…
Mps Acacia2005—
*K052117I-Flow Elastomeric Pump (on-Q, Painbuster, C-Block, Homepump, Easypump)
Summary page 1
…7 cleared September 21, 2006 MPS MedFIo K052451 cleared December 8, 2005 I-Flow K052117 cleared September 9, 2005 Device Description: The AccuFlo & AccuFlux models are both disposable, non-electric infusion pu…
I-Flow Corp.20053

Also cleared against K050770: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151650SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd.20163

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 6, 2014Class IIZ-2655-2014Labeling Change ControlThe primary packaging label (inner pouch label) indicates that the package contents are Catalog Item: CT-1000-250, while the secondary packaging label (outer carton label) indicates the contents are CT-1000-100. The actual contents are a…
Dec 13, 2013Class IIZ-0921-2014Process designInternal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
Oct 20, 2011Not classifiedZ-0267-2012Process controlRedi-Flo Elastomeric Infusion Pumps may have a higher flow rate than specified (more than +15%).
Oct 20, 2011Not classifiedZ-0268-2012Process controlRedi-Flo Elastomeric Infusion Pumps may have a higher flow rate than specified (more than +15%).
Oct 20, 2011Not classifiedZ-0269-2012Process controlRedi-Flo Elastomeric Infusion Pumps may have a higher flow rate than specified (more than +15%).
Jul 28, 2011Not classifiedZ-0081-2012Nonconforming Material/ComponentThe product may have a higher flow rate than specified resulting in over administration of drug solutions to the patients.
Jul 9, 2011Class IZ-0161-2013Nonconforming Material/ComponentPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being reclassified by the Agency as a Class I. The recall was initiated because First Medical Source has…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 7 companies; a new generation was cleared every 9 months on median.
  • 10 of the 43 ancestors have a recall of any class.
  • Of its descendants, K151650 (Epic Medical Pte., Ltd., 2016) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K050770.

How does its category compare?

All MEB devices

MEB Pump, Infusion, Elastomeric: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · MEB

2in 2024

2007: 52008: 22009: 12013: 12014: 12016: 12019: 12024: 22026 (partial year): 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEB

42in 2007

2007: 42 days median
2007high 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEB

17in 2018

2007: 12008: 32009: 482010: 22011: 8, 1 Class I2013: 2, 1 Class I2018: 17
2007high 48 (2009) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MEB

1,408in 2025

2007: 1102008: 4442009: 7982010: 1,8272011: 2,2652012: 1,9762013: 1,9562014: 2,6952015: 2,9802016: 2,1472017: 1,6532018: 1,4732019: 9742020: 2,7322021: 1,8022022: 1,2882023: 1,2482024: 1,4242025: 1,408
2007high 2,980 (2015) · low 110 (2007)2025

MAUDE reports naming product code MEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075421110
20082—3444
20091—48798
20100—21,827
20110—8 (1)2,265
20120—01,976
20131—2 (1)1,956
20141—02,695
20150—02,980
20161—02,147
20170—01,653
20180—171,473
20191—0974
20200—02,732
20210—01,802
20220—01,288
20230—01,248
20242—01,424
20250—01,408
20261—0—

What is it indicated for?

AccuFlo &AccuFlux devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications. It is the responsibility of the user to ensure that the medication is prepared and administered accordance with the drug manufacturer's package insert. The devices deliver controlled amounts of medication directly to the intraoperative site for pain management and or antibiotic…

Read the full statement (113 words)

AccuFlo &AccuFlux devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications. It is the responsibility of the user to ensure that the medication is prepared and administered accordance with the drug manufacturer's package insert. The devices deliver controlled amounts of medication directly to the intraoperative site for pain management and or antibiotic administration. The devices infuse the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes.

The devices are also intended for controlled delivery of local anesthetics in close proximity to nerves for post operative regional anesthesia and pain management. Routes of administration may be intraoperative or percutaneous.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K081905: Accuflow and Accuflux Elastomeric Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K081905

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026