510(k) · Cleared
Recalled · Class II · 2011I-Stat Portable Clinical Analyzer, Model 300
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
I-Stat Corp. · CGA Glucose Oxidase, Glucose
Cleared on Oct 5, 2000 after a 156-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 18 later devices descend from it. It has been recalled 13 times.
- Decided
- Oct 5, 2000
- Review
- 156 days
- Class
- Class II
- Product code
- CGA
- Recalls
- 13
- Predicates
- 0
18 later devices descend from it, 6 citing it directly.
- This deviceK001387 I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp. · 2000
- Descendant · cites it directly · 8 of its ownK103195 I-Stat 1 Wireless AnalyzerAbbott Point of Care, Inc. · 2011
- Descendant · cites it directly · 7 of its ownK061597 Epoc Blood Analysis SystemEpocal, Inc. · 2006 · Recalled
- Descendant · 2 generations on · 5 of its ownK191360 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
- Descendant · 2 generations on · 5 of its ownK163342 i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories · 2017
- Descendant · 3 generations on · 4 of its ownK183680 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
Showing 5 of 18 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 18 · See all- K103195I-Stat 1 Wireless AnalyzerAbbott Point of Care, Inc. · 2011 · cites it directly8descendants
- K061597Epoc Blood Analysis SystemEpocal, Inc. · 2006 · cites it directly7descendants
- K191360i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 2 generations on5descendants
- K163342i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories · 2017 · 2 generations on5descendants
- K183680i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K001387 | I-Stat Portable Clinical Analyzer, Model 300 | I-Stat Corp. | 2000 | — |
| +1 | K103195 | I-Stat 1 Wireless Analyzer | Abbott Point of Care, Inc. | 2011 | — |
| +1 | K061597 | Epoc Blood Analysis System | Epocal, Inc. | 2006 | 2 |
| +2 | K191360 | i-STAT CHEM8+ cartridge with the i-STAT 1 System | Abbott Point of Care, Inc. | 2020 | — |
| +2 | K163342 | i-STAT Hematocrit test with i-STAT Alinity System | Abbott Laboratories | 2017 | — |
| +3 | K183680 | i-STAT CHEM8+ cartridge with the i-STAT 1 System | Abbott Point of Care, Inc. | 2020 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 2 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2020 | 4 | 0 |
| 2021 | 1 | 0 |
| 2023 | 3 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in CGA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CGA device
- #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (18)
See all 186 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K103195 | I-Stat 1 Wireless Analyzer | 2011 | 8 |
| K061597 | Epoc Blood Analysis System | 2006 | 7 |
| K093297 | Epoc Bgem, Epoc Reader, Epoc Host, Epoc Care-Fill Capillary Tube | 2010 | 2 |
| K113726 | Epoc Chloride Test and Epoc Creatinine Test | 2012 | 1 |
| K092849 | Blood Collection Tube Samples for Use with Point of Care Blood Analyzer | 2010 | 1 |
| K090109 | Blood Gas, Electrolyte and Metabolite Test Card | 2009 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 8, 2011 | Class II | Z-1973-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 8, 2011 | Class II | Z-1975-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 8, 2011 | Class II | Z-1974-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 8, 2011 | Class II | Z-1976-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 8, 2011 | Class II | Z-1978-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 8, 2011 | Class II | Z-1977-2012 | Device Design | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension. |
| Sep 20, 2010 | Not classified | Z-0598-2011 | Other | i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. |
| Sep 20, 2010 | Not classified | Z-0597-2011 | Other | i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component. |
| Aug 20, 2010 | Not classified | Z-2462-2010 | Device Design | Reliability issues associated with broken connection pins within the downloader that mate with the pins from the analyzer. |
| Nov 6, 2009 | Not classified | Z-1133-2010 | Component design/selection | Some Martel Printers used with the i-STAT 1 system have the potential of being damaged due to the use or rechargeable battery packs that do not contain a fuse. |
| Sep 29, 2009 | Not classified | Z-0865-2010 | Software design | Abbott Point of Care has determined that it is possible under certain circumstances that some test records could have their status update messages mis-matched with another record, resulting in the incorrect patient name, order number of… |
| Sep 9, 2009 | Not classified | Z-0812-2010 | Labeling mix-ups | The storage temperature on the label on the side of the i-Stat Portable Clinical Analyzer shipping box does not match product specification. |
| Apr 22, 2009 | Not classified | Z-2328-2009 | Nonconforming Material/Component | It appears that the rechargeable battery pack is overheating with or without the presence of smoke. |
The 5 latest of 13 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K103195 (Abbott Point of Care, Inc., 2011) has gone furthest, with 8 descendants of its own.
How does its category compare?
All CGA devicesCGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · CGA
2in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · CGA
249in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CGA
3in 2023
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · CGA
59in 2025
MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 122 | 2 | 29 |
| 2008 | 6 | 56 | 2 | 37 |
| 2009 | 10 | 155 | 4 | 75 |
| 2010 | 8 | 122 | 6 | 172 |
| 2011 | 9 | 217 | 3 | 423 |
| 2012 | 10 | 123 | 0 | 233 |
| 2013 | 3 | 441 | 3 | 172 |
| 2014 | 5 | 265 | 2 | 122 |
| 2015 | 4 | 150 | 6 (2) | 155 |
| 2016 | 4 | 157 | 2 | 36 |
| 2017 | 2 | — | 0 | 92 |
| 2018 | 2 | — | 0 | 69 |
| 2019 | 2 | — | 1 | 116 |
| 2020 | 2 | — | 0 | 68 |
| 2021 | 1 | — | 0 | 48 |
| 2022 | 1 | — | 0 | 272 |
| 2023 | 4 | 249 | 3 | 1,141 |
| 2024 | 2 | — | 0 | 266 |
| 2025 | 0 | — | 0 | 59 |
| 2026 | 1 | — | 0 | — |
https://510kreport.com/device/K001387 · printed Nov 6, 2026
Cite this page
510(k) Report. “K001387: I-Stat Portable Clinical Analyzer, Model 300.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K001387
Report an extraction error