510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Epoc Blood Analysis System

K061597510(k) TraditionalCleared Sep 27, 2006111-day reviewClass II21 CFR 862.1665Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Epocal, Inc. · JGS Electrode, Ion Specific, Sodium

Cleared on Sep 27, 2006 as substantially equivalent to K001387 (I-Stat Corp., 2000) after a 111-day review. Its family tree reaches back 1 generation to 2000. 7 later devices descend from it. It has been recalled 2 times.

Decided
Sep 27, 2006
Review
111 days
Class
Class II
Product code
JGS
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K001387 (2000). 7 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2020: 2 descendants cleared2021: 1 descendant cleared2023: 3 descendants cleared2026: 1 descendant clearedK001387 · I-Stat Portable Clinical Analyzer, Model 300 · I-Stat Corp. · 2000 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K001387 · I-Stat Portable Clinical Analyzer, Model 300 · I-Stat Corp. · 2000 · recalled0
  1. Root · 1 generation backK001387 I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp. · 2000 · Recalled
  2. This deviceK061597 Epoc Blood Analysis SystemEpocal, Inc. · 2006
  3. Descendant · cites it directly · 5 of its ownK191360 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  4. Descendant · 2 generations on · 3 of its ownK210958 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021
  5. Descendant · 3 generations on · 1 of its ownK223755 i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023
  6. Descendant · 4 generations on · 0 of its ownK252488 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2026
  7. Descendant · 2 generations on · 0 of its ownK230300 i-STAT CG8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2020: 22021: 12023: 32026: 1
2020high 3 (2023) · low 0 (2022)2026

Most influential descendants

Showing 5 of 7 · See all
  • K191360i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · cites it directly5descendants
  • K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021 · 2 generations on3descendants
  • K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023 · 3 generations on1descendant
  • K252488ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2026 · 4 generations on0descendants
  • K230300i-STAT CG8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K001387I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp.200013
0 (this device)K061597Epoc Blood Analysis SystemEpocal, Inc.2006—
+1K191360i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+2K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2021—
+3K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2023—
+4K252488ABL90 FLEX PLUS SystemRadiometer Medicals Aps2026—
+2K230300i-STAT CG8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202020
202110
202330
202610
FIG. 1 — Lineage of K061597, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001387I-Stat Portable Clinical Analyzer, Model 300
Summary page 3
…Device EPOC Blood Analysis System i-STAT Model 300 510(k) # To be determined K001387 Same / Item Device Predicate Differe…
I-Stat Corp.200013

Also cleared against K001387: 5 other devices →

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191360i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.20205
K200107epoc Blood Analysis SystemEpocal, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 12, 2022Class IIZ-1214-2023Under Investigation by firmpO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
Oct 16, 2008Not classifiedZ-0952-2009Otherelevated range of potassium levels

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K191360 (Abbott Point of Care, Inc., 2020) has gone furthest, with 5 descendants of its own.
  • 5 other devices also cite its predicate K001387.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—

What is it indicated for?

The EPOC Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood in the laboratory or at the point of care in hospitals, nursing homes or other clinical care institutions. The Blood Gas Electrolyte (BGE) test card panel configuration includes sensors for Sodium - Na, Potassium - K, ionized Calcium - iCa, pH, pCO2, pO2 and Hematocrit - Hct.…

Read the full statement (148 words)

The EPOC Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood in the laboratory or at the point of care in hospitals, nursing homes or other clinical care institutions.

The Blood Gas Electrolyte (BGE) test card panel configuration includes sensors for Sodium - Na, Potassium - K, ionized Calcium - iCa, pH, pCO2, pO2 and Hematocrit - Hct.

Measurement of Sodium and Potassium are used in diagnosis and treatment diseases involving electrolyte imbalance. Measurement of Ionized Calcium is used in diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. Measurement of ph pCO2, p 0 2 (blood gases) is used in the diagnosis and treatment of life-threatening acid-base disturbances. Measurement Hot distinguish normal from abnormal states of blood volume, such as anemia and erythrocytosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061597: Epoc Blood Analysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K061597

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026