I-Stat 1 Wireless Analyzer

K103195510(k) TraditionalCleared Feb 8, 2011102-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Point of Care, Inc. · CGA Glucose Oxidase, Glucose

Cleared on Feb 8, 2011 as substantially equivalent to K001387 (I-Stat Corp., 2000) after a 102-day review. Its family tree reaches back 1 generation to 2000. 8 later devices descend from it.

Decided
Feb 8, 2011
Review
102 days
Class
Class II
Product code
CGA
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K001387 (2000). 8 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2017: 2 descendants cleared2016: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared2023: 2 descendants cleared2026: 1 descendant clearedK001387 · I-Stat Portable Clinical Analyzer, Model 300 · I-Stat Corp. · 2000 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 1 descendant K001387 · I-Stat Portable Clinical Analyzer, Model 300 · I-Stat Corp. · 2000 · recalled0+5
  1. Root · 1 generation backK001387 I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp. · 2000 · Recalled
  2. This deviceK103195 I-Stat 1 Wireless AnalyzerAbbott Point of Care, Inc. · 2011
  3. Descendant · cites it directly · 5 of its ownK163342 i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories · 2017
  4. Descendant · 2 generations on · 4 of its ownK183680 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  5. Descendant · 3 generations on · 3 of its ownK210958 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021
  6. Descendant · 4 generations on · 1 of its ownK223755 i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023
  7. Descendant · 5 generations on · 0 of its ownK252488 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2026

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2016: 12017: 22020: 12021: 12023: 22026: 1
2016high 2 (2017) · low 0 (2018)2026

Most influential descendants

Showing 5 of 8 · See all
  • K163342i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories · 2017 · cites it directly5descendants
  • K183680i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 2 generations on4descendants
  • K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021 · 3 generations on3descendants
  • K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023 · 4 generations on1descendant
  • K252488ABL90 FLEX PLUS SystemRadiometer Medicals Aps · 2026 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K001387I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp.200013
0 (this device)K103195I-Stat 1 Wireless AnalyzerAbbott Point of Care, Inc.2011—
+1K163342i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories2017—
+2K183680i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+3K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2021—
+4K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2023—
+5K252488ABL90 FLEX PLUS SystemRadiometer Medicals Aps2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201720
202010
202110
202320
202610
FIG. 1 — Lineage of K103195, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGA device

  1. #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001387I-Stat Portable Clinical Analyzer, Model 300I-Stat Corp.200013

Also cleared against K001387: 5 other devices →

What descends from it? (8)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K163342i-STAT Hematocrit test with i-STAT Alinity SystemAbbott Laboratories20175
K163271i-Stat Alinity System with i-Stat Glucose testAbbott Point of Care, Inc.20170
K153357i-STAT Alinity System with i-STAT Sodium testAbbott Point of Care, Inc.20160

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K163342 (Abbott Laboratories, 2017) has gone furthest, with 5 descendants of its own.
  • 5 other devices also cite its predicate K001387.

How does its category compare?

All CGA devices

CGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · CGA

2in 2024

2007: 72008: 62009: 102010: 82011: 92012: 102013: 32014: 52015: 42016: 42017: 22018: 22019: 22020: 22021: 12022: 12023: 42024: 22026 (partial year): 1
2007high 10 (2009) · low 0 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGA

249in 2023

2007: 122 days median2008: 56 days median2009: 155 days median2010: 122 days median2011: 217 days median2012: 123 days median2013: 441 days median2014: 265 days median2015: 150 days median2016: 157 days median2023: 249 days median
2007high 441 (2013) · low 56 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGA

3in 2023

2007: 22008: 22009: 42010: 62011: 32013: 32014: 22015: 6, 2 Class I2016: 22019: 12023: 3
2007high 6 (2010) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · CGA

59in 2025

2007: 292008: 372009: 752010: 1722011: 4232012: 2332013: 1722014: 1222015: 1552016: 362017: 922018: 692019: 1162020: 682021: 482022: 2722023: 1,1412024: 2662025: 59
2007high 1,141 (2023) · low 29 (2007)2025

MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077122229
2008656237
200910155475
201081226172
201192173423
2012101230233
201334413172
201452652122
201541506 (2)155
20164157236
20172—092
20182—069
20192—1116
20202—068
20211—048
20221—0272
2023424931,141
20242—0266
20250—059
20261—0—

What is it indicated for?

The i-STAT 1 Wireless Analyzer is used by trained medical professionals for running a variety of clinical chemistry tests and test panels contained in i-STAT test cartridges. These tests include hematocrit, glucose, blood urea nitrogen, sodium, potassium, chloride, ionized calcium, blood gases (oxygen, carbon dioxide and pH), creatinine, lactate, activated clotting time, prothrombin time, bicarbonate/carbon dioxide, troponin, creatine phosphokinase, and…

Read the full statement (608 words)

The i-STAT 1 Wireless Analyzer is used by trained medical professionals for running a variety of clinical chemistry tests and test panels contained in i-STAT test cartridges. These tests include hematocrit, glucose, blood urea nitrogen, sodium, potassium, chloride, ionized calcium, blood gases (oxygen, carbon dioxide and pH), creatinine, lactate, activated clotting time, prothrombin time, bicarbonate/carbon dioxide, troponin, creatine phosphokinase, and beta natriuretic peptide.

* Sodium measurements are used for monitoring electrolyte imbalances.

* Potassium measurements are used for diagnosis and monitoring of diseases and clinical conditions that manifest high and low Potassium levels.

* Chloride measurements are primarily used in the diagnosis, monitoring and treatment of electrolyte and metabolic disorders including but not limited to cystic fibrosis, diabetic acidosis and hydration disorders.

* Glucose measurements are used in the diagnosis, monitoring and treatment of carbohydrate metabolism disorders including but not limited to diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemnia, and of pancreatic islet cell carcinoma.

* Hematocrit measurements can aid in the determination and monitoring of normal or abnormal total red cell volume status including but not limited to conditions such anemia and erythrocytosis and blood loss related to trauma and surgery.

* Blood urea nitrogen measurements are used for the diagnosis, monitoring and treatment of certain renal and metabolic diseases.

* Ionized calcium measurements are used in the diagnosis, monitoring and treatment of conditions including but not limited to parathyroid disease, a variety of bone diseases, chronic renal disease and tetany and disturbances related to surgical and intensive care.

* pH-, pCO2 and p02 measurements are used in the diagnosis, monitoring and treatment of respiratory disturbances and metabolic and respiratory based acid-base disturbances.

* Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

* The i-STAT lactate test is useful for (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia.

* The i-STAT Kaolin Activated Clotting Time (ACT) test is an in vitro diagnostic test used to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery.

* The i-STAT PT, a prothrombin time test, is useful in monitory patients receiving oral anticoagulation therapy such as Coumnadin or warfarin.

* Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

* The i-STAT Cardiac Troponin I (cTnI) test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in whole blood or plasma. Measurements of cardiac troponin I are used in the diagnosis and treatment of myocardial infarction and as an aid in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality.

* The i-STAT CKMB test is an in vitro diagnostic test for the quantitative measurement of creatinine kinase MB mass in whole blood or plasma samples. CK-MB; measurements can be used as an aid in the diagnosis and treatment of myocardial infarction (MI).

* The i-STAT BNP test is an in vitro diagnostic test for the quantitative measurement of B-Type Natriuretic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure.

* The i-STAT Celite ACT test is useful for monitoring patients receiving heparin for treatment of pulmonary embolism or venous thrombosis, and for monitoring anticoagulation therapy in patients undergoing medical procedures such as catheterization, cardiac surgery, surgery, organ transplant and dialysis

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103195: I-Stat 1 Wireless Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103195

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.