510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Abl-700 Series, Model Abl 7XX (Various Models)

K002290510(k) TraditionalCleared Sep 13, 200048-day reviewClass II21 CFR 862.1120Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical A/S · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Sep 13, 2000 after a 48-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 30 later devices descend from it. It has been recalled 9 times.

Decided
Sep 13, 2000
Review
48 days
Class
Class II
Product code
CHL
Recalls
9
Predicates
0

30 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2006: 1 descendant cleared2004: 1 descendant cleared2005: 3 descendants cleared2015: 3 descendants cleared2020: 1 descendant cleared2012: 3 descendants cleared2013: 3 descendants cleared2025: 1 descendant cleared2021: 2 descendants cleared2011: 2 descendants cleared2008: 2 descendants cleared2014: 1 descendant cleared2017: 1 descendant cleared2016: 2 descendants cleared2024: 1 descendant cleared2026: 1 descendant cleared2023: 1 descendant cleared200020052010201520202025
  1. This deviceK002290 Abl-700 Series, Model Abl 7XX (Various Models)Radiometer Medical A/S · 2000
  2. Descendant · cites it directly · 29 of its ownK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  3. Descendant · 2 generations on · 18 of its ownK050869 Modification to ABL800 FlexRadiometer Medical Aps · 2005 · Recalled
  4. Descendant · cites it directly · 12 of its ownK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled
  5. Descendant · 2 generations on · 6 of its ownK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  6. Descendant · 3 generations on · 5 of its ownK122729 ABL90 FlexRadiometer Medical Aps · 2012

Showing 5 of 30 descendants. See all descendants

Descendants cleared per year

2004: 12005: 32006: 12008: 22010: 12011: 22012: 32013: 32014: 12015: 32016: 22017: 12020: 12021: 22023: 12024: 12025: 12026: 1
2004high 3 (2005) · low 0 (2007)2026

Most influential descendants

Showing 5 of 30 · See all
  • K041874ABL800 FlexRadiometer Medical Aps · 2004 · cites it directly29descendants
  • K050869Modification to ABL800 FlexRadiometer Medical Aps · 2005 · 2 generations on18descendants
  • K092686ABL90 Flex SeriesRadiometer Medical Aps · 2010 · cites it directly12descendants
  • K120197ABL90 FlexRadiometer Medical Aps · 2012 · 2 generations on6descendants
  • K122729ABL90 FlexRadiometer Medical Aps · 2012 · 3 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K002290Abl-700 Series, Model Abl 7XX (Various Models)Radiometer Medical A/S2000—
+1K041874ABL800 FlexRadiometer Medical Aps200422
+2K050869Modification to ABL800 FlexRadiometer Medical Aps20051
+1K092686ABL90 Flex SeriesRadiometer Medical Aps201010
+2K120197ABL90 FlexRadiometer Medical Aps20123
+3K122729ABL90 FlexRadiometer Medical Aps2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200530
200610
200820
201010
201120
201230
201330
201410
201530
201620
201710
202010
202120
202310
202410
202510
202610
FIG. 1 — Lineage of K002290, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.2 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (30)

See all 30

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041874ABL800 FlexRadiometer Medical Aps200429
K092686ABL90 Flex SeriesRadiometer Medical Aps201012
K051968ABL8X7 FlexRadiometer Medical Aps20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 8, 2012Class IIZ-2303-2013Process controlRADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Jul 5, 2012Class IIZ-2163-2012Software designA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are…
Jul 5, 2012Class IIZ-2162-2012Software designA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are…
Apr 16, 2009Not classifiedZ-1554-2009Process designWhen removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.
Sep 5, 2008Not classifiedZ-0457-2009Process controlThe firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these…
Jan 18, 2008Not classifiedZ-1653-2008Software designReported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and…
Nov 13, 2006Not classifiedZ-1094-2007Software designSoftware anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas…
Sep 29, 2005Not classifiedZ-1386-06OtherUnder certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and…
Jun 17, 2005Not classifiedZ-1317-06OtherSoftware defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041874 (Radiometer Medical Aps, 2004) has gone furthest, with 29 descendants of its own.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—
Cite this page

510(k) Report. “K002290: Abl-700 Series, Model Abl 7XX (Various Models).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K002290

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026