510(k) Report

510(k) · Cleared

Cleared

Airlife Hepa Filter

K011132510(k) SpecialCleared Jun 8, 200156-day reviewClass II21 CFR 868.5260

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Allegiance Healthcare Corp. · CAH Filter, Bacterial, Breathing-Circuit

Cleared on Jun 8, 2001 after a 56-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 8 of the 41 devices in its family, counting itself, have a Class I recall. 40 later devices descend from it.

Decided
Jun 8, 2001
Review
56 days
Class
Class II
Product code
CAH
Recalls
None
Predicates
0

40 later devices descend from it, 1 citing it directly. 8 of the 41 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2002: 3 descendants cleared2014: 1 descendant cleared2004: 2 descendants cleared2008: 3 descendants cleared, 2 with a Class I recall22005: 3 descendants cleared2003: 2 descendants cleared2011: 3 descendants cleared2006: 1 descendant cleared2010: 1 descendant cleared2009: 1 descendant cleared2022: 2 descendants cleared, 1 with a Class I recall12012: 5 descendants cleared, 2 with a Class I recall22026: 2 descendants cleared2013: 2 descendants cleared, 1 with a Class I recall12016: 2 descendants cleared2019: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared2025: 2 descendants cleared2021: 1 descendant cleared, 1 with a Class I recall12024: 2 descendants cleared200020052010201520202025
  1. This deviceK011132 Airlife Hepa FilterAllegiance Healthcare Corp. · 2001
  2. Descendant · cites it directly · 39 of its ownK013089 Ems Hepa Filter and Hme CombinationsEngineered Medical Systems · 2002
  3. Descendant · 2 generations on · 29 of its ownK033008 Air Safety Hepa and Non-Hepa FiltersAir Safety, Ltd. · 2004
  4. Descendant · 3 generations on · 22 of its ownK060705 Carina HomeDraeger Medical B.V. · 2006 · Recalled
  5. Descendant · 4 generations on · 20 of its ownK103167 Stellar 150Resmed Germany, Inc. · 2011
  6. Descendant · 2 generations on · 11 of its ownK021265 Datex-Ohmeda Hmef 500Datex-Ohmeda · 2002

Showing 5 of 40 descendants. See all descendants

Descendants cleared per year

2002: 32003: 22004: 22005: 32006: 12008: 32009: 12010: 12011: 32012: 52013: 22014: 12016: 22017: 12019: 12021: 12022: 22024: 22025: 22026: 2
2002high 5 (2012) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 40 · See all
  • K013089Ems Hepa Filter and Hme CombinationsEngineered Medical Systems · 2002 · cites it directly39descendants
  • K033008Air Safety Hepa and Non-Hepa FiltersAir Safety, Ltd. · 2004 · 2 generations on29descendants
  • K060705Carina HomeDraeger Medical B.V. · 2006 · 3 generations on22descendants
  • K103167Stellar 150Resmed Germany, Inc. · 2011 · 4 generations on20descendants
  • K021265Datex-Ohmeda Hmef 500Datex-Ohmeda · 2002 · 2 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K011132Airlife Hepa FilterAllegiance Healthcare Corp.2001—
+1K013089Ems Hepa Filter and Hme CombinationsEngineered Medical Systems2002—
+2K033008Air Safety Hepa and Non-Hepa FiltersAir Safety, Ltd.2004—
+3K060705Carina HomeDraeger Medical B.V.20061
+4K103167Stellar 150Resmed Germany, Inc.2011—
+2K021265Datex-Ohmeda Hmef 500Datex-Ohmeda2002—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200230
200320
200420
200530
200610
200832
200910
201010
201130
201252
201321
201410
201620
201710
201911
202111
202221
202420
202520
202620
FIG. 1 — Lineage of K011132, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAH device

  1. #1 K251583 Ambu® Virobac II® Exhalation Filterscore 11.6 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (40)

See all 40

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K013089Ems Hepa Filter and Hme CombinationsEngineered Medical Systems200239

Has anything in its family been recalled? (8 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072002Filter Carestar 30; Filter Safestar 55; Filter/Hme Twinstar 55Draeger Medical AG & Co. KG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K221836Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · 2022 · 3 generations on · ◆ in the 2022 descendants bar of FIG. 14 Class I recalls
  • K072885Carina System, Model 8419350; Carina Basic Device, Model 5704110Draeger Medical AG & Co. KG · 2008 · 4 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K113053Bipap a 30 Ventilatory Support SystemRespironics, Inc. · 2012 · 5 generations on · ◆ in the 2012 descendants bar of FIG. 15 Class I recalls
  • K121623Bipap a 40 Ventilatory Support SystemRespironics, Inc. · 2012 · 6 generations on · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls
  • K122715Stellar 150Resmed Germany, Inc. · 2013 · 6 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 7 generations on · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 8 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K013089 (Engineered Medical Systems, 2002) has gone furthest, with 39 descendants of its own.

How does its category compare?

All CAH devices

CAH Filter, Bacterial, Breathing-Circuit: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CAH

2in 2024

2007: 22008: 12009: 42010: 52011: 32012: 32013: 12014: 42015: 22017: 12019: 22020: 22021: 22022: 32023: 32024: 22026 (partial year): 2
2007high 5 (2010) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAH

237in 2023

2009: 111 days median2010: 216 days median2011: 281 days median2012: 212 days median2014: 250 days median2022: 204 days median2023: 237 days median
2007high 281 (2011) · low 111 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CAH

1in 2025

2010: 22013: 12014: 12017: 12022: 6, 6 Class I2025: 1, 1 Class I
2007high 6 (2022) · low 0 (2007)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · CAH

77in 2025

2007: 322008: 272009: 182010: 402011: 92012: 162013: 472014: 982015: 512016: 342017: 292018: 332019: 302020: 332021: 562022: 672023: 402024: 512025: 77
2007high 98 (2014) · low 9 (2011)2025

MAUDE reports naming product code CAH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—032
20081—027
20094111018
20105216240
2011328109
20123212016
20131—147
20144250198
20152—051
20160—034
20171—129
20180—033
20192—030
20202—033
20212—056
202232046 (6)67
20233237040
20242—051
20250—1 (1)77
20262—0—
Cite this page

510(k) Report. “K011132: Airlife Hepa Filter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K011132

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026