510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Modification to Coulter LH 750 Hematology Analyzer, Model 6605632

K022161510(k) SpecialCleared Jul 29, 200226-day reviewClass II21 CFR 864.5220Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on Jul 29, 2002 as substantially equivalent to K011342 (Beckman Coulter, Inc., 2001) after a 26-day review. Its family tree reaches back 1 generation to 2001. 2 later devices descend from it. It has been recalled 2 times.

Decided
Jul 29, 2002
Review
26 days
Class
Class II
Product code
GKZ
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K011342 (2001). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2004: 1 descendant clearedK011342 · Coulter LH 750 Hematology Analyzer, Model 6605632 · Beckman Coulter, Inc. · 2001 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K011342 · Coulter LH 750 Hematology Analyzer, Model 6605632 · Beckman Coulter, Inc. · 2001 · recalled0
  1. Root · 1 generation backK011342 Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc. · 2001 · Recalled
  2. This deviceK022161 Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc. · 2002
  3. Descendant · cites it directly · 1 of its ownK032013 Coulter Act 5DIFF AL Hematology AnalyzerBeckman Coulter, Inc. · 2003
  4. Descendant · 2 generations on · 0 of its ownK042173 Coulter a T 5DIFF Cap Pierce (CP)Beckman Coulter, Inc. · 2004

Most influential descendants

Showing 2 of 2 · See all
  • K032013Coulter Act 5DIFF AL Hematology AnalyzerBeckman Coulter, Inc. · 2003 · cites it directly1descendant
  • K042173Coulter a T 5DIFF Cap Pierce (CP)Beckman Coulter, Inc. · 2004 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K011342Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.200115
0 (this device)K022161Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.2002—
+1K032013Coulter Act 5DIFF AL Hematology AnalyzerBeckman Coulter, Inc.2003—
+2K042173Coulter a T 5DIFF Cap Pierce (CP)Beckman Coulter, Inc.2004—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
FIG. 1 — Lineage of K022161, 2001–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011342Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.200115

Also cleared against K011342: 7 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032013Coulter Act 5DIFF AL Hematology AnalyzerBeckman Coulter, Inc.20031

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 26, 2017Class IIZ-3113-2017Labeling Change ControlBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Nov 8, 2002Not classifiedZ-0415-03OtherSoftware vote out logic changed. Erroneous white blood cell counts.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K032013 (Beckman Coulter, Inc., 2003) has gone furthest, with 1 descendant of its own.
  • 7 other devices also cite its predicate K011342.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The COULTER@LH 750 Hematology Analyzer is a quantitative, automated hematology Analyze; and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER@LH 750 Hematology Analyzer also provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs).

Future commercialization will add ISOTONB 4 diluent /Lyse S@4 Lytic reagent to the indications for use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K022161: Modification to Coulter LH 750 Hematology Analyzer, Model 6605632.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022161

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026