510(k) · Cleared
Recalled · Class I · 20221.5T Infinity Twinspeed MR System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
GE Medical Systems · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Aug 3, 2001 as substantially equivalent to K980114 (GE Medical Systems, 1998) and 1 other predicate after a 21-day review. Its family tree reaches back 1 generation to 1996. 3 of the 106 devices in its family, counting itself, have a Class I recall. 103 later devices descend from it. It has been recalled 9 times, including a Class I recall.
- Decided
- Aug 3, 2001
- Review
- 21 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 9 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 1 generation to K962061 (1996). 2 ancestors in all. 103 later devices descend from it, 2 citing it directly. 3 of the 106 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK962061 Magnetic Resonance Diagnostic SystemGE Medical Systems · 1996 · Recalled
- This deviceK012200 1.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001
- Descendant · cites it directly · 100 of its ownK013636 Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · Recalled (Class I)
- Descendant · 2 generations on · 99 of its ownK041476 GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Recalled (Class I)
- Descendant · 3 generations on · 90 of its ownK052293 GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · Recalled
- Descendant · 4 generations on · 70 of its ownK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
- Descendant · 4 generations on · 69 of its ownK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
Showing 5 of 103 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 103 · See all- K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · cites it directly100descendants
- K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 2 generations on99descendants
- K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · 3 generations on90descendants
- K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · 4 generations on70descendants
- K083147Discovery MR450Ge Medical Systems, LLC · 2008 · 4 generations on69descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K962061 | Magnetic Resonance Diagnostic System | GE Medical Systems | 1996 | 2 |
| 0 (this device) | K012200 | 1.5T Infinity Twinspeed MR System | GE Medical Systems | 2001 | — |
| +1 | K013636 | Signa Infinity MR System with Excite Technology | GE Medical Systems | 2002 | Class I |
| +2 | K041476 | GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System | Ge Medical Systems, LLC | 2004 | Class I |
| +3 | K052293 | GE 1.5T and 3.0T Signa HDX MR System | Ge Medical Systems, LLC | 2005 | 21 |
| +4 | K081028 | Signa 3.0T MR750 System | Ge Medical Systems, LLC | 2008 | 11 |
| +4 | K083147 | Discovery MR450 | Ge Medical Systems, LLC | 2008 | 10 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
| 2002 | 1 | 1 |
| 2004 | 1 | 1 |
| 2005 | 3 | 0 |
| 2007 | 1 | 0 |
| 2008 | 3 | 0 |
| 2009 | 2 | 0 |
| 2010 | 3 | 0 |
| 2011 | 4 | 0 |
| 2012 | 5 | 0 |
| 2013 | 7 | 0 |
| 2014 | 6 | 0 |
| 2015 | 6 | 0 |
| 2016 | 6 | 0 |
| 2017 | 6 | 0 |
| 2018 | 3 | 0 |
| 2019 | 5 | 0 |
| 2020 | 8 | 0 |
| 2021 | 5 | 0 |
| 2022 | 8 | 0 |
| 2023 | 7 | 0 |
| 2024 | 2 | 0 |
| 2025 | 6 | 0 |
| 2026 | 4 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K980114 | Magnetic Resonance Diagnostic SystemSummary page 1…m, 21 CFR 892.1000, 90-LNH Marketed Device: The Signa 1.5T TwinSpeed System is substantially equivalent to the currently marketed Signa CVMR Magnetic Resonance System (K980114) and the Signa Horizon Cx (K962061). Device Description: The Signa 1.5T TwinSpeed incorporates the Signa Horizon Cx (K962061) a… | 1998 |
| * | K962061 | Magnetic Resonance Diagnostic SystemSummary page 1…d Device: The Signa 1.5T TwinSpeed System is substantially equivalent to the currently marketed Signa CVMR Magnetic Resonance System (K980114) and the Signa Horizon Cx (K962061). Device Description: The Signa 1.5T TwinSpeed incorporates the Signa Horizon Cx (K962061) and the Signa CVMR Magnetic Resonanc… | 1996 |
What descends from it? (103)
See all 1032 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 18, 2022 | Class II | Z-0857-2022 | Device Design | There is potential for the images to be flipped left to right. |
| Dec 23, 2021 | Class II | Z-0604-2022 | Process control | The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements. |
| Nov 15, 2018 | Class II | Z-0693-2019 | Software design | Possible incorrect software version loaded. |
| Jun 7, 2018 | Class II | Z-0112-2019 | Software change control | The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. |
| Jun 7, 2018 | Class II | Z-0110-2019 | Software change control | The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. |
| Jun 7, 2018 | Class II | Z-0111-2019 | Software change control | The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch. |
| Jul 10, 2015 | Class II | Z-2339-2015 | Nonconforming Material/Component | A gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation… |
| Mar 9, 2015 | Class II | Z-1305-2015 | Software Manufacturing/Software Deployment | GE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites. |
| Jan 7, 2015 | Class I | Z-1035-2015 | Employee error | At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be… |
The 5 latest of 9 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
- K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- 1 of the 2 ancestors has a recall of any class.
- Of its descendants, K013636 (GE Medical Systems, 2002) has gone furthest, with 100 descendants of its own.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The Signa 1.5T TwinSpeed Magnetic Resonance System is a whole body scanner designed for shorter scan times. The Signa TwinSpeed is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head or body. The images produced by the Signa TwinSpeed system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that…
Read the full statement (104 words)
The Signa 1.5T TwinSpeed Magnetic Resonance System is a whole body scanner designed for shorter scan times. The Signa TwinSpeed is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head or body. The images produced by the Signa TwinSpeed system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K012200 · printed Nov 6, 2026
Cite this page
510(k) Report. “K012200: 1.5T Infinity Twinspeed MR System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012200