GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System

K041476510(k) SpecialCleared Jun 17, 200414-day reviewClass II21 CFR 892.1000Recalls 15 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jun 17, 2004 as substantially equivalent to K013636 (GE Medical Systems, 2002) and 1 other predicate after a 14-day review. Its family tree reaches back 3 generations to 1996. 4 of the 109 devices in its family, counting itself, have a Class I recall. 99 later devices descend from it. It has been recalled 15 times, including a Class I recall.

Decided
Jun 17, 2004
Review
14 days
Class
Class II
Product code
LNH
Recalls
15 (1 Class I)
Predicates
2

Its longest line of descent runs back 3 generations to K962061 (1996). 9 ancestors in all. 99 later devices descend from it, 3 citing it directly. 4 of the 109 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor2001: 2 other ancestors1999: 1 other ancestor1998: 1 other ancestor2005: 3 descendants cleared2014: 6 descendants cleared2012: 5 descendants cleared2008: 3 descendants cleared2009: 2 descendants cleared2007: 1 descendant cleared2011: 4 descendants cleared2010: 3 descendants cleared2013: 7 descendants cleared2020: 8 descendants cleared2015: 6 descendants cleared2016: 5 descendants cleared2018: 3 descendants cleared2021: 5 descendants cleared2022: 8 descendants cleared2025: 6 descendants cleared2017: 6 descendants cleared2023: 7 descendants cleared2019: 5 descendants cleared2024: 2 descendants cleared2026: 4 descendants clearedK962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalledK012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recallK013636 · Signa Infinity MR System with Excite Technology · GE Medical Systems · 2002 · Class I recall1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor, 1 with a Class I recall12 generations back: 2 other ancestors3 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 8 descendants3 generations forward: 13 descendants4 generations forward: 17 descendants5 generations forward: 11 descendants6 generations forward: 8 descendants7 generations forward: 8 descendants8 generations forward: 7 descendants9 generations forward: 8 descendants10 generations forward: 5 descendants11 generations forward: 2 descendants12 generations forward: 5 descendants13 generations forward: 4 descendants K013636 · Signa Infinity MR System with Excite Technology · GE Medical Systems · 2002 · Class I recallK012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recallK962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalled0+5+10
  1. Root · 3 generations backK962061 Magnetic Resonance Diagnostic SystemGE Medical Systems · 1996 · Recalled
  2. 2 generations backK012200 1.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · Recalled (Class I)
  3. 1 generation backK013636 Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · Recalled (Class I)
  4. This deviceK041476 GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004
  5. Descendant · cites it directly · 90 of its ownK052293 GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · Recalled
  6. Descendant · 2 generations on · 70 of its ownK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  7. Descendant · 2 generations on · 69 of its ownK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  8. Descendant · 3 generations on · 59 of its ownK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled
  9. Descendant · 3 generations on · 52 of its ownK103327 Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · Recalled

Showing 5 of 99 descendants. See all descendants

Descendants cleared per year

2005: 32007: 12008: 32009: 22010: 32011: 42012: 52013: 72014: 62015: 62016: 52017: 62018: 32019: 52020: 82021: 52022: 82023: 72024: 22025: 62026: 4
2005high 8 (2020) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 99 · See all
  • K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · cites it directly90descendants
  • K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · 2 generations on70descendants
  • K083147Discovery MR450Ge Medical Systems, LLC · 2008 · 2 generations on69descendants
  • K091536Optima MR450WGe Medical Systems, LLC · 2009 · 3 generations on59descendants
  • K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · 3 generations on52descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K962061Magnetic Resonance Diagnostic SystemGE Medical Systems19962
−2K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems2001Class I
−1K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems2002Class I
0 (this device)K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC2004—
+1K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC200521
+2K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
+2K083147Discovery MR450Ge Medical Systems, LLC200810
+3K091536Optima MR450WGe Medical Systems, LLC20098
+3K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation20116

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
199910
200120
200210
200311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200530
200710
200830
200920
201030
201140
201250
201370
201460
201560
201650
201760
201830
201950
202080
202150
202280
202370
202420
202560
202640
FIG. 1 — Lineage of K041476, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013636Signa Infinity MR System with Excite Technology
Summary page 1
…m, 21 CFR 892.1000, 90-LNH Marketed Device: The Signa® Excite 1.5T MR System is substantially equivalent to the currently marketed Signao 1.5T Infinity MR system (K01 3636) with the main differences being a change to the receive chain architecture that includes sixteen independent receive channels,…
GE Medical Systems2002Class I
*K030874GE 3.0T Signa Infinity Twinspeed with Excite MR System
Summary page 1
…16 channel increments. The Signa® Excite 3.OT MR System is substantially equivalent to the currently marketed Signa® 3.OT Infinity with Excite Technology MR system (K030874) with the main differences being a change to the receive chain architecture that includes sixteen independent receive channel…
General Electric Co.2003Class I

What descends from it? (99)

See all 99

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC200590
K133848Rthawk, Rthawk, or Heartvista Workstation with RthawkHeartvista, Inc.20146
K043262GE Signa Excite MR Surgical OptionGE Healthcare20050

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2022Class IIZ-0858-2022Device DesignThere is potential for the images to be flipped left to right.
Dec 23, 2021Class IIZ-0615-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Jun 7, 2018Class IIZ-0112-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0110-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0111-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Mar 30, 2018Class IIZ-2527-2018Nonconforming Material/ComponentThere is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical…
Aug 31, 2015Class IIZ-0257-2016Software designWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could…
Jul 10, 2015Class IIZ-2339-2015Nonconforming Material/ComponentA gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation…
Mar 9, 2015Class IIZ-1305-2015Software Manufacturing/Software DeploymentGE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.
Jan 7, 2015Class IZ-1035-2015Employee errorAt certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be…
Nov 21, 2011Class IIZ-1943-2013Software designScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system…
Nov 9, 2010Not classifiedZ-1429-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are…
Oct 6, 2009Not classifiedZ-0592-2010Software designGE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T, Signa HD 1.5T…
Dec 14, 2007Not classifiedZ-1891-2008Software designGE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner.
Oct 1, 2007Not classifiedZ-2211-2008Software designA software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · Direct predicate · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 2.4 years on median.
  • 5 of the 9 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K052293 (Ge Medical Systems, LLC, 2005) has gone furthest, with 90 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® Excite MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ…

Read the full statement (122 words)

The GE Signa® Excite MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® Excite MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa® Excite system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041476: GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K041476

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.