GE 1.5T and 3.0T Signa HDX MR System

K052293510(k) SpecialCleared Sep 21, 200529-day reviewClass II21 CFR 892.1000Recalls 21

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Sep 21, 2005 as substantially equivalent to K041476 (Ge Medical Systems, LLC, 2004) after a 29-day review. Its family tree reaches back 4 generations to 1996. 4 of the 101 devices in its family, counting itself, have a Class I recall. 90 later devices descend from it. It has been recalled 21 times.

Decided
Sep 21, 2005
Review
29 days
Class
Class II
Product code
LNH
Recalls
21
Predicates
1

Its longest line of descent runs back 4 generations to K962061 (1996). 10 ancestors in all. 90 later devices descend from it, 7 citing it directly. 4 of the 101 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor2001: 2 other ancestors1999: 1 other ancestor1998: 1 other ancestor2005: 1 descendant cleared2012: 5 descendants cleared2008: 3 descendants cleared2009: 2 descendants cleared2007: 1 descendant cleared2011: 4 descendants cleared2010: 3 descendants cleared2013: 7 descendants cleared2020: 8 descendants cleared2014: 4 descendants cleared2015: 6 descendants cleared2018: 3 descendants cleared2021: 4 descendants cleared2022: 8 descendants cleared2025: 5 descendants cleared2023: 7 descendants cleared2016: 4 descendants cleared2017: 5 descendants cleared2019: 4 descendants cleared2024: 2 descendants cleared2026: 4 descendants clearedK962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalledK012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recallK013636 · Signa Infinity MR System with Excite Technology · GE Medical Systems · 2002 · Class I recallK041476 · GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System · Ge Medical Systems, LLC · 2004 · Class I recall1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor, 1 with a Class I recall13 generations back: 2 other ancestors4 generations back: 3 other ancestors 1 generation forward: 7 descendants2 generations forward: 12 descendants3 generations forward: 16 descendants4 generations forward: 10 descendants5 generations forward: 7 descendants6 generations forward: 7 descendants7 generations forward: 7 descendants8 generations forward: 8 descendants9 generations forward: 5 descendants10 generations forward: 2 descendants11 generations forward: 5 descendants12 generations forward: 4 descendants K041476 · GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System · Ge Medical Systems, LLC · 2004 · Class I recallK013636 · Signa Infinity MR System with Excite Technology · GE Medical Systems · 2002 · Class I recallK012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recallK962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalled0+5+10
  1. Root · 4 generations backK962061 Magnetic Resonance Diagnostic SystemGE Medical Systems · 1996 · Recalled
  2. 3 generations backK012200 1.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · Recalled (Class I)
  3. 2 generations backK013636 Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · Recalled (Class I)
  4. 1 generation backK041476 GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Recalled (Class I)
  5. This deviceK052293 GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005
  6. Descendant · cites it directly · 70 of its ownK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  7. Descendant · cites it directly · 69 of its ownK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  8. Descendant · 2 generations on · 59 of its ownK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled
  9. Descendant · 2 generations on · 52 of its ownK103327 Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · Recalled
  10. Descendant · 3 generations on · 44 of its ownK113490 Optima MR450WGe Medical Systems, LLC · 2012 · Recalled

Showing 5 of 90 descendants. See all descendants

Descendants cleared per year

2005: 12007: 12008: 32009: 22010: 32011: 42012: 52013: 72014: 42015: 62016: 42017: 52018: 32019: 42020: 82021: 42022: 82023: 72024: 22025: 52026: 4
2005high 8 (2020) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 90 · See all
  • K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · cites it directly70descendants
  • K083147Discovery MR450Ge Medical Systems, LLC · 2008 · cites it directly69descendants
  • K091536Optima MR450WGe Medical Systems, LLC · 2009 · 2 generations on59descendants
  • K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · 2 generations on52descendants
  • K113490Optima MR450WGe Medical Systems, LLC · 2012 · 3 generations on44descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K962061Magnetic Resonance Diagnostic SystemGE Medical Systems19962
−3K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems2001Class I
−2K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems2002Class I
−1K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC2004Class I
0 (this device)K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC2005—
+1K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
+1K083147Discovery MR450Ge Medical Systems, LLC200810
+2K091536Optima MR450WGe Medical Systems, LLC20098
+2K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation20116
+3K113490Optima MR450WGe Medical Systems, LLC20123

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
199910
200120
200210
200311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200710
200830
200920
201030
201140
201250
201370
201440
201560
201640
201750
201830
201940
202080
202140
202280
202370
202420
202550
202640
FIG. 1 — Lineage of K052293, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System
Summary page 1
…ic System, 21 CFR 892.1000, 90-LNH Marketed Device: The 1.5T Signa® H~x MR System is substantially equivalent to the currently marketed Signa® Excite 1.5T MR system (K041476) with the main differences being a change to the receive chain architecture that includes thirty two independent receive channel…
Ge Medical Systems, LLC2004Class I

Also cleared against K041476: 2 other devices →

What descends from it? (90)

See all 90

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200870
K083147Discovery MR450Ge Medical Systems, LLC200869
K072998Ideal Software Option for GE MR ScannersGe Medical Systems, LLC20078
K052978GE 1.5T Signa Hde MR SystemGe Medical Systems, LLC20058
K1216761.5T Signa HDXT 3.0T Signa HDXT Signa VibrantGe Medical Systems, LLC20125
K083421MR-Touch OptionGe Medical Systems, LLC20094
K081916GE Signa MR Oncology PackageGe Medical Systems, LLC20083

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 14, 2024Class IIZ-2463-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Jun 14, 2024Class IIZ-2471-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Feb 18, 2022Class IIZ-0860-2022Device DesignThere is potential for the images to be flipped left to right.
Dec 23, 2021Class IIZ-0603-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Aug 26, 2021Class IIZ-0141-2022Software designUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Jun 4, 2021Class IIZ-2065-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Dec 23, 2019Class IIZ-0877-2020Software designIt was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an…
Jun 7, 2018Class IIZ-0112-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0110-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0111-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
May 11, 2017Class IIZ-3027-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Aug 31, 2015Class IIZ-0258-2016Software designWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could…
Aug 31, 2015Class IIZ-0259-2016Software designWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could…
Jun 8, 2012Class IIZ-1207-2013Software designGE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of…
Nov 21, 2011Class IIZ-1943-2013Software designScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system…
May 10, 2011Class IIZ-0935-2013Software designAn error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports.
Nov 9, 2010Not classifiedZ-1428-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are…
Oct 6, 2009Not classifiedZ-0592-2010Software designGE Healthcare has recently become aware that the use of Confirma, Inc.'s CADstream in conjunction with the GE Healthcare's PURE (Phased Array Uniformity Enhancement) software for dynamic MRI imaging (Signa Excite 1.5T, Signa HD 1.5T…
Mar 28, 2008Not classifiedZ-2225-2008Software change controlPossible misdiagnosis or inappropriate treatment based on the image. Signal homogeneity may be affected when BRAVO applications are set with IR Prep off. The concern is that signal inhomogeneity can result in an erroneous diagnosis.
Dec 14, 2007Not classifiedZ-1890-2008Software designGE healthcare has become aware of an artifact which could affect the diagnostic capability associated with the 3.0T Propeller T2 Flair application of the 3.0T Signa HDx scanner.
Oct 1, 2007Not classifiedZ-2210-2008Software designA software issue that may result in misregistration of functional and anatomical images associated with the 3.0T BrainwaveRT fMRI Application on the 3.0T Signa HDx scanner.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 2 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 2.4 years on median.
  • 6 of the 10 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K081028 (Ge Medical Systems, LLC, 2008) has gone furthest, with 70 descendants of its own.
  • 2 other devices also cite its predicate K041476.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine…

Read the full statement (123 words)

The GE Signa® HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa® HDx MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa e HDx system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052293: GE 1.5T and 3.0T Signa HDX MR System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K052293

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026