Signa Infinity MR System with Excite Technology

K013636510(k) TraditionalCleared Jan 24, 200280-day reviewClass II21 CFR 892.1000Recalls 12 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Medical Systems · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jan 24, 2002 as substantially equivalent to K012200 (GE Medical Systems, 2001) after an 80-day review. Its family tree reaches back 2 generations to 1996. 3 of the 104 devices in its family, counting itself, have a Class I recall. 100 later devices descend from it. It has been recalled 12 times, including a Class I recall.

Decided
Jan 24, 2002
Review
80 days
Class
Class II
Product code
LNH
Recalls
12 (1 Class I)
Predicates
1

Its longest line of descent runs back 2 generations to K962061 (1996). 3 ancestors in all. 100 later devices descend from it, 1 citing it directly. 3 of the 104 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor2004: 1 descendant cleared, 1 with a Class I recall12005: 3 descendants cleared2014: 6 descendants cleared2012: 5 descendants cleared2008: 3 descendants cleared2009: 2 descendants cleared2007: 1 descendant cleared2011: 4 descendants cleared2010: 3 descendants cleared2013: 7 descendants cleared2020: 8 descendants cleared2015: 6 descendants cleared2016: 5 descendants cleared2018: 3 descendants cleared2021: 5 descendants cleared2022: 8 descendants cleared2025: 6 descendants cleared2017: 6 descendants cleared2023: 7 descendants cleared2019: 5 descendants cleared2024: 2 descendants cleared2026: 4 descendants clearedReference device: K012970 · GE Signa Asset Imaging Option · GE Medical Systems · 2001K962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalledK012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recall1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 3 descendants3 generations forward: 8 descendants4 generations forward: 13 descendants5 generations forward: 17 descendants6 generations forward: 11 descendants7 generations forward: 8 descendants8 generations forward: 8 descendants9 generations forward: 7 descendants10 generations forward: 8 descendants11 generations forward: 5 descendants12 generations forward: 2 descendants13 generations forward: 5 descendants14 generations forward: 4 descendants K012200 · 1.5T Infinity Twinspeed MR System · GE Medical Systems · 2001 · Class I recallK962061 · Magnetic Resonance Diagnostic System · GE Medical Systems · 1996 · recalled0+5+10
  1. Root · 2 generations backK962061 Magnetic Resonance Diagnostic SystemGE Medical Systems · 1996 · Recalled
  2. 1 generation backK012200 1.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · Recalled (Class I)
  3. This deviceK013636 Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002
  4. Descendant · cites it directly · 99 of its ownK041476 GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Recalled (Class I)
  5. Descendant · 2 generations on · 90 of its ownK052293 GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · Recalled
  6. Descendant · 3 generations on · 70 of its ownK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  7. Descendant · 3 generations on · 69 of its ownK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  8. Descendant · 4 generations on · 59 of its ownK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled

Showing 5 of 100 descendants. See all descendants

Descendants cleared per year

2004: 12005: 32007: 12008: 32009: 22010: 32011: 42012: 52013: 72014: 62015: 62016: 52017: 62018: 32019: 52020: 82021: 52022: 82023: 72024: 22025: 62026: 4
2004high 8 (2020) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 100 · See all
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · cites it directly99descendants
  • K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC · 2005 · 2 generations on90descendants
  • K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · 3 generations on70descendants
  • K083147Discovery MR450Ge Medical Systems, LLC · 2008 · 3 generations on69descendants
  • K091536Optima MR450WGe Medical Systems, LLC · 2009 · 4 generations on59descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K962061Magnetic Resonance Diagnostic SystemGE Medical Systems19962
−1K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems2001Class I
0 (this device)K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems2002—
+1K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC2004Class I
+2K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC200521
+3K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
+3K083147Discovery MR450Ge Medical Systems, LLC200810
+4K091536Optima MR450WGe Medical Systems, LLC20098

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200411
200530
200710
200830
200920
201030
201140
201250
201370
201460
201560
201650
201760
201830
201950
202080
202150
202280
202370
202420
202560
202640
FIG. 1 — Lineage of K013636, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K0122001.5T Infinity Twinspeed MR System
Summary page 1
…Diagnostic System, 21 CFR 892.1000, 90-LNH Marketed Device: The Signa Infinity System is substantially equivalent to the currently marketed Signa TwinSpeed MR system (K012200) with the main differences being an addition of eight independent receive channels and an update of the technology used for the C…
GE Medical Systems2001Class I
†K012970GE Signa Asset Imaging OptionGE Medical Systems2001—

Also cleared against K012200: 1 other device →

What descends from it? (100)

See all 100

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR System Class IGe Medical Systems, LLC200499

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 23, 2021Class IIZ-0605-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Nov 15, 2018Class IIZ-0693-2019Software designPossible incorrect software version loaded.
Jun 7, 2018Class IIZ-0112-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0110-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Jun 7, 2018Class IIZ-0111-2019Software change controlThe latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.
Aug 31, 2015Class IIZ-0255-2016Software designWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could…
Jul 10, 2015Class IIZ-2339-2015Nonconforming Material/ComponentA gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation…
Mar 9, 2015Class IIZ-1305-2015Software Manufacturing/Software DeploymentGE Healthcare has become aware of a potential safety issue involving MRI systems due to software versions not being maintained properly at some sites.
Jan 7, 2015Class IZ-1035-2015Employee errorAt certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be…
Nov 21, 2011Class IIZ-1943-2013Software designScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system…
Nov 9, 2010Not classifiedZ-1432-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are…
May 11, 2009Not classifiedZ-0462-2011Software designSigna MR Excite" 1.5T system containing 11.1_M4_0818.a software, associated with 2D Fast Spin Echo based pulse sequences (2D FSE-XL, 2D FRFSE-XL, 2D FSE-IR, 2D T2FLAIR and 2D T1FLAIR) may experience an image flip along the phase encode…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 6 years and passes through 1 company; a new generation was cleared every 4.7 years on median.
  • 2 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K041476 (Ge Medical Systems, LLC, 2004) has gone furthest, with 99 descendants of its own.
  • 1 other device also cites its predicate K012200.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The GE Signa Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head…

Read the full statement (121 words)

The GE Signa Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013636: Signa Infinity MR System with Excite Technology.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K013636

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026