510(k) Report

510(k) · Cleared

Recalled · 2010

Heartstream/Heartstart FR2 Aed Ecg Cable, Models M3860A, M3861A, M3840A, M3841A

K013425510(k) TraditionalCleared Jan 14, 200290-day reviewClass III21 CFR 870.5310Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jan 14, 2002 after a 90-day review. The FDA summary names its predicate without a number. 2 of the 41 devices in its family, counting itself, have a Class I recall. 40 later devices descend from it. It has been recalled 10 times.

Decided
Jan 14, 2002
Review
90 days
Class
Class III
Product code
MKJ
Recalls
10
Predicates
0

40 later devices descend from it, 9 citing it directly. 2 of the 41 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2008: 5 descendants cleared2005: 2 descendants cleared2013: 2 descendants cleared2012: 8 descendants cleared, 1 with a Class I recall12006: 3 descendants cleared, 1 with a Class I recall12011: 4 descendants cleared2010: 2 descendants cleared2014: 6 descendants cleared2015: 4 descendants cleared2018: 1 descendant cleared2021: 1 descendant cleared2024: 1 descendant cleared200020052010201520202025
  1. This deviceK013425 Heartstream/Heartstart FR2 Aed Ecg Cable, Models M3860A, M3861A…Philips Medical Systems · 2002
  2. Descendant · cites it directly · 23 of its ownK041892 Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005
  3. Descendant · 2 generations on · 21 of its ownK060559 Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · Recalled
  4. Descendant · 3 generations on · 11 of its ownK100654 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010
  5. Descendant · 3 generations on · 9 of its ownK110742 Onestep Adult Multi-Function ElectrodeZOLL Medical Corporation · 2011 · Recalled
  6. Descendant · 4 generations on · 9 of its ownK102468 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010

Showing 5 of 40 descendants. See all descendants

Descendants cleared per year

2004: 12005: 22006: 32008: 52010: 22011: 42012: 82013: 22014: 62015: 42018: 12021: 12024: 1
2004high 8 (2012) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 40 · See all
  • K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005 · cites it directly23descendants
  • K060559Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · 2 generations on21descendants
  • K100654Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010 · 3 generations on11descendants
  • K110742Onestep Adult Multi-Function ElectrodeZOLL Medical Corporation · 2011 · 3 generations on9descendants
  • K102468Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010 · 4 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K013425Heartstream/Heartstart FR2 Aed Ecg Cable, Models M3860A, M3861A, M3840A, M3841APhilips Medical Systems2002—
+1K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation2005—
+2K060559Zoll R Series DefibrillatorZOLL Medical Corporation20061
+3K100654Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters2010—
+3K110742Onestep Adult Multi-Function ElectrodeZOLL Medical Corporation20111
+4K102468Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200520
200631
200850
201020
201140
201281
201320
201460
201540
201810
202110
202410
FIG. 1 — Lineage of K013425, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Heartstream FR2 AED
Named without a submission number
—
*—Laerdal Heartstart 3000
Named without a submission number
—
*—Heartstream XLT Defibrillator/Monitor
Named without a submission number
—
*—Active Corporation ActiveECG device
Named without a submission number
—

What descends from it? (40)

See all 40

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation200523
K052057Lifepak 500 Automated External DefibrillatorMedtronic Emergency Response Systems, Inc.20064
K040637Powerheart Aed G3 ProCardiac Science, Inc.20043
K042404Lifepak, Model 1000Medtronic Emergency Response Systems, Inc.20051
K130684Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20130
K121853Ddu-2400/2450 Semiautomatic External Defibrillator and AccessoriesDefibtech, LLC20120
K120350Extended Life BatteryAed Battery Exchange, LLC20120
K073716Life+Cel or Lifecel BatteryAmco International Manufacturing & Design, Inc.20080
K062970I-PADCu Medical Systems, Inc.20080

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 22, 2010Not classifiedZ-0487-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Sep 22, 2010Not classifiedZ-0483-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Sep 22, 2010Not classifiedZ-0485-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Sep 22, 2010Not classifiedZ-0486-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Sep 22, 2010Not classifiedZ-0488-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
Sep 22, 2010Not classifiedZ-0484-2011Nonconforming Material/ComponentA flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
May 28, 2010Not classifiedZ-0470-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0472-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0473-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0471-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K052465Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models Sde 251, SPP 351Heartsine Technologies, Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K041892 (ZOLL Medical Corporation, 2005) has gone furthest, with 23 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The FR2 is intended to be used with disposable Heartstream defibrillation pads applied to a person who is experiencing the symptoms of sudden cardiac arrest (SCA): unresponsiveness and absence of breathing. When the patient is under 8 years or weighs less than 55 pounds (25 kg), the FR2 should be used with attenuated pediatric defibrillation pads. The FR2 is intended for use by personnel who have been trained in its operation. The user should be…

Read the full statement (161 words)

The FR2 is intended to be used with disposable Heartstream defibrillation pads applied to a person who is experiencing the symptoms of sudden cardiac arrest (SCA): unresponsiveness and absence of breathing.

When the patient is under 8 years or weighs less than 55 pounds (25 kg), the FR2 should be used with attenuated pediatric defibrillation pads.

The FR2 is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support (BLS), advanced life support (ALS) or other physician-authorized emergency medical response.

At the discretion of emergency care personnel, the M3860A FR2 with ECG display enabled can also be used with the FR2 ECG Cable to display the rhythm of a responsive or breathing patient, regardless of age. The FR2 and ECG Cable system provides a non-diagnostic display for attended patient monitoring. While connected to the FR2 ECG Cable, the FR2 evaluates the patient’s ECG and disables it’s shock capability.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013425: Heartstream/Heartstart FR2 Aed Ecg Cable, Models M3860A, M3861A, M3840A, M3841A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K013425

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026