Magnetom Trio System

K013586510(k) TraditionalCleared Dec 28, 200159-day reviewClass II21 CFR 892.1000Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Systems, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Dec 28, 2001 after a 59-day review. The FDA summary names its predicate without a number. 22 of the 437 devices in its family, counting itself, have a Class I recall. 436 later devices descend from it. It has been recalled 4 times.

Decided
Dec 28, 2001
Review
59 days
Class
Class II
Product code
LNH
Recalls
4
Predicates
0

436 later devices descend from it, 3 citing it directly. 22 of the 437 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2002: 1 descendant cleared2006: 5 descendants cleared2005: 3 descendants cleared2011: 14 descendants cleared2010: 6 descendants cleared, 1 with a Class I recall12004: 1 descendant cleared2007: 4 descendants cleared, 1 with a Class I recall12013: 23 descendants cleared, 2 with a Class I recall22024: 31 descendants cleared2023: 42 descendants cleared, 3 with a Class I recall32015: 18 descendants cleared, 1 with a Class I recall12012: 13 descendants cleared, 2 with a Class I recall22008: 8 descendants cleared2009: 5 descendants cleared2014: 16 descendants cleared2018: 26 descendants cleared, 2 with a Class I recall22021: 18 descendants cleared, 1 with a Class I recall12020: 39 descendants cleared, 3 with a Class I recall32016: 15 descendants cleared, 1 with a Class I recall12022: 37 descendants cleared, 2 with a Class I recall22025: 32 descendants cleared, 1 with a Class I recall12019: 29 descendants cleared, 1 with a Class I recall12017: 23 descendants cleared, 1 with a Class I recall12026: 27 descendants cleared200020052010201520202025
  1. This deviceK013586 Magnetom Trio SystemSiemens Medical Systems, Inc. · 2001
  2. Descendant · cites it directly · 428 of its ownK050200 Magnetom Trio a Tim SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
  3. Descendant · 2 generations on · 312 of its ownK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  4. Descendant · 2 generations on · 308 of its ownK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  5. Descendant · 2 generations on · 300 of its ownK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  6. Descendant · 3 generations on · 229 of its ownK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008

Showing 5 of 436 descendants. See all descendants

Descendants cleared per year

2002: 12004: 12005: 32006: 52007: 42008: 82009: 52010: 62011: 142012: 132013: 232014: 162015: 182016: 152017: 232018: 262019: 292020: 392021: 182022: 372023: 422024: 312025: 322026: 27
2002high 42 (2023) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 436 · See all
  • K050200Magnetom Trio a Tim SystemSiemens Medical Solutions USA, Inc. · 2005 · cites it directly428descendants
  • K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005 · 2 generations on312descendants
  • K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006 · 2 generations on308descendants
  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 2 generations on300descendants
  • K082427Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008 · 3 generations on229descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K013586Magnetom Trio SystemSiemens Medical Systems, Inc.2001—
+1K050200Magnetom Trio a Tim SystemSiemens Medical Solutions USA, Inc.20053
+2K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
+2K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
+2K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
+3K082427Syngo MR B17Siemens Medical Solutions USA, Inc.2008—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200410
200530
200650
200741
200880
200950
201061
2011140
2012132
2013232
2014160
2015181
2016151
2017231
2018262
2019291
2020393
2021181
2022372
2023423
2024310
2025321
2026270
FIG. 1 — Lineage of K013586, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

22 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—MAGNETOM 3.0 T Allegra Dedicated Head only System
Named without a submission number
—
*—GE’s Body Imaging Coil for 3.0T Signa VH /1 system
Named without a submission number
—
*—MAGNETOM 1.5 T Sonata System
Named without a submission number
—
*—3.0 T Dedicated Head Only Allegra System
Named without a submission number
—

What descends from it? (436)

See all 436

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050200Magnetom Trio a Tim SystemSiemens Medical Solutions USA, Inc.2005428
K052423Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc.2006199
K020991Syngo MR 2002BSiemens Medical Solutions USA, Inc.20026

Has anything in its family been recalled? (22 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 3, 2011Not classifiedZ-2021-2011OtherFirm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. The potential exists for a cable connection to become overheated, leading to the emission…
Jul 28, 2004Not classifiedZ-1403-04OtherAn input error can be made with Magnetom MRI software. Software allows users to manually change Transmit Reference Voltage in certain windows
Jul 28, 2004Not classifiedZ-1404-04OtherAn input error can be made with Magnetom MRI software. Software allows users to manually change Transmit Reference Voltage in certain windows
Dec 18, 2003Not classifiedZ-0521-04Othertissue roll can become magnetic

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · 3 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K130885Magnetom Avanto-Fit, Magnetom Skyra-FitSiemens Medical Solutions USA, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K121160Magnetom SpectraSiemens Medical Solutions USA, Inc. · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
  • K181613MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11DSiemens Medical Solutions USA, Inc. · 2018 · 5 generations on · ◆ in the 2018 descendants bar of FIG. 12 Class I recalls
  • K132119Magnetom Prisma, Magnetom Prisma FitSiemens Medical Solutions USA, Inc. · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K153343MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medi Cal Solutions, Inc. · 2016 · 6 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K151579MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medical Solutions USA, Inc. · 2015 · 6 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K200213Biograph mMR with syngo MR E11P-AP01 system softwareSiemens Medical Solutions USA, Inc. · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K170396MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K173592MAGNETOM Aera, MAGNETOM Skyra/Skyrafit, MAGNETOM Prisma/Prismafit, MAGNETOM AvantofitSiemens Medical Solutions USA, Inc. · 2018 · 7 generations on · ◆ in the 2018 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 22 shown.

What does this lineage say?

  • Of its descendants, K050200 (Siemens Medical Solutions USA, Inc., 2005) has gone furthest, with 428 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The MAGNETOM Trio system is a whole body scanner designed to support higher resolution imaging and shorter scan times. The Trio system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head and whole body. The images produced by the Trio system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin echo time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013586: Magnetom Trio System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K013586

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026