510(k) · Cleared
Recalled · Class II · 2012Miethke Shunt System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Aesculap, Inc. · HCA Catheter, Ventricular
Cleared on Apr 5, 2002 as substantially equivalent to K980778 (Johnson & Johnson Professionals, Inc., 1998) and 4 other predicates after a 30-day review. Its family tree reaches back 3 generations to 1995. 2 later devices descend from it. It has been recalled once.
- Decided
- Apr 5, 2002
- Review
- 30 days
- Class
- Class II
- Product code
- HCA
- Recalls
- 1
- Predicates
- 6
Its longest line of descent runs back 3 generations to K944222 (1995). 7 ancestors in all, the oldest K861377 (1986). 2 later devices descend from it, 2 citing it directly.
- Root · 3 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
- 2 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
- 1 generation backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
- This deviceK020728 Miethke Shunt SystemAesculap, Inc. · 2002
- Descendant · cites it directly · 0 of its ownK242003 XABO Ventricular Catheter, XABO Peritoneal Catheter, XABO Catheter SetAesculap, Inc. · 2024
- Descendant · cites it directly · 0 of its ownK200950 Ventricular CatheterAesculap, Inc. · 2020
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K944222 | Codman-Medos Valve Shunt System | Johnson & Johnson Professionals, Inc. | 1995 | — |
| −2 | K973774 | Codman Hakim Micro Precision Valve | Johnson & Johnson Professionals, Inc. | 1997 | — |
| −1 | K980778 | Hakim Micro Programmable Valve System | Johnson & Johnson Professionals, Inc. | 1998 | — |
| 0 (this device) | K020728 | Miethke Shunt System | Aesculap, Inc. | 2002 | — |
| +1 | K242003 | XABO Ventricular Catheter, XABO Peritoneal Catheter, XABO Catheter Set | Aesculap, Inc. | 2024 | — |
| +1 | K200950 | Ventricular Catheter | Aesculap, Inc. | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1986 | 1 | 0 |
| 1990 | 1 | 0 |
| 1994 | 1 | 0 |
| 1996 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2020 | 1 | 0 |
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HCA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (6)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K980778 | Hakim Micro Programmable Valve SystemSummary page 1…plicable ASTM and ISO standards. SUBSTANTIAL EQUIVALENCE The Aesculap Miethke Shunt System is essentially identical to the J & J Hakim Micro Programmable Valve System (K980778), Heyer-Schulte Novus (K961859), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating A… | 1998 |
| * | K961859 | Novus Standard and Pediatric ValvesSummary page 1…. SUBSTANTIAL EQUIVALENCE The Aesculap Miethke Shunt System is essentially identical to the J & J Hakim Micro Programmable Valve System (K980778), Heyer-Schulte Novus (K961859), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating Accessory (K932429) and the Cord… | 1996 |
| Kept off the family tree | K811410 | FDA’s K811410 is Medi-Trace Our OCR of the scanned summary read K811410, FDA’s number for Medi-Trace (Graphic Controls Corp.), which shares no word with the summary’s name for the device, “P/S Medtronic Bulton”, apart from generic ones like “system”. Kept off the family tree. Summary page 1…Aesculap Miethke Shunt System is essentially identical to the J & J Hakim Micro Programmable Valve System (K980778), Heyer-Schulte Novus (K961859), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating Accessory (K932429) and the Cordis NMT Hakim Standard/Pediatric… | 1981 |
| * | K902783 | PS Medical Delta CSF-Flow Control ValveSummary page 1…s essentially identical to the J & J Hakim Micro Programmable Valve System (K980778), Heyer-Schulte Novus (K961859), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating Accessory (K932429) and the Cordis NMT Hakim Standard/Pediatric Valves (K861377). === Page 2 =… | 1990 |
| * | K932429 | Cordis Gravity Compensating AccessorySummary page 1…rammable Valve System (K980778), Heyer-Schulte Novus (K961859), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating Accessory (K932429) and the Cordis NMT Hakim Standard/Pediatric Valves (K861377). SERVICES, » %, C DEPARTMENT OF HEALTH & HUMAN SER… | 1994 |
| * | K861377 | Standard/Pediatric Cordis Hydrocephalus ValvesSummary page 1…), P/S Medtronic Bulton (K811410), P/S Medtronic Delta (K902783), Cordis NMT Gravity Compensating Accessory (K932429) and the Cordis NMT Hakim Standard/Pediatric Valves (K861377). SERVICES, » %, C DEPARTMENT OF HEALTH & HUMAN SERVICES . - Public Health Service REALY) oo Ha, & i, pel R… | 1986 |
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 4, 2012 | Class II | Z-0805-2013 | Process control | The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 2.4 years on median.
- 2 of the 7 ancestors have a recall of any class.
- Of its descendants, K242003 (Aesculap, Inc., 2024) has gone furthest, with 0 descendants of its own.
- 6 other devices also cite its predicate K980778.
How does its category compare?
All HCA devicesHCA Catheter, Ventricular: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · HCA
1in 2017
2026 to Oct 25, outlined.
Median 510(k) review, days · HCA
Too few HCA 510(k)s in any year since 2007 to give a median.
Recalls per year · HCA
No recall names HCA since 2007.
Adverse event reports per year · HCA
1in 2021
MAUDE reports naming product code HCA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 8 |
| 2008 | 0 | — | 0 | 3 |
| 2009 | 0 | — | 0 | 1 |
| 2010 | 0 | — | 0 | 4 |
| 2011 | 0 | — | 0 | 19 |
| 2012 | 0 | — | 0 | 1 |
| 2013 | 1 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 2 |
| 2016 | 0 | — | 0 | 1 |
| 2017 | 1 | — | 0 | 3 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 2 |
| 2020 | 0 | — | 0 | 1 |
| 2021 | 0 | — | 0 | 1 |
| 2022 | 0 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Miethke Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K020728 · printed Nov 4, 2026
Cite this page
510(k) Report. “K020728: Miethke Shunt System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K020728