PS Medical Delta CSF-Flow Control Valve

K902783510(k) TraditionalCleared Aug 8, 199043-day reviewClass II21 CFR 882.5550Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pudenz-Schulte Medical Research Corp. · JXG Shunt, Central Nervous System And Components

Cleared on Aug 8, 1990 after a 43-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 40 devices in its family, counting itself, has a Class I recall. 39 later devices descend from it. It has been recalled 5 times.

Decided
Aug 8, 1990
Review
43 days
Class
Class II
Product code
JXG
Recalls
5
Predicates
0

39 later devices descend from it, 4 citing it directly. 1 of the 40 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2004: 3 descendants cleared2000: 1 descendant cleared2002: 3 descendants cleared2005: 2 descendants cleared2006: 3 descendants cleared2020: 1 descendant cleared2024: 2 descendants cleared2017: 2 descendants cleared2012: 3 descendants cleared2009: 2 descendants cleared2008: 2 descendants cleared2007: 1 descendant cleared2011: 2 descendants cleared2016: 1 descendant cleared, 1 with a Class I recall12022: 1 descendant cleared2015: 2 descendants cleared2010: 2 descendants cleared2013: 1 descendant cleared2018: 2 descendants cleared2021: 1 descendant cleared2025: 1 descendant cleared2019: 1 descendant cleared1990200020102020
  1. This deviceK902783 PS Medical Delta CSF-Flow Control ValvePudenz-Schulte Medical Research Corp. · 1990
  2. Descendant · cites it directly · 25 of its ownK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
  3. Descendant · 2 generations on · 23 of its ownK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
  4. Descendant · 2 generations on · 19 of its ownK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
  5. Descendant · cites it directly · 9 of its ownK992173 Siphonguard CSF Control DeviceJohnson & Johnson Professionals, Inc. · 2000
  6. Descendant · 2 generations on · 8 of its ownK020667 Codman Hakim Shunt SystemsCodman & Shurtleff, Inc. · 2002

Showing 5 of 39 descendants. See all descendants

Descendants cleared per year

2000: 12002: 32004: 32005: 22006: 32007: 12008: 22009: 22010: 22011: 22012: 32013: 12015: 22016: 12017: 22018: 22019: 12020: 12021: 12022: 12024: 22025: 1
2000high 3 (2002) · low 0 (2001)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 39 · See all
  • K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002 · cites it directly25descendants
  • K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without BioglideMedtronic Neurosurgery · 2004 · 2 generations on23descendants
  • K042465PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · 2 generations on19descendants
  • K992173Siphonguard CSF Control DeviceJohnson & Johnson Professionals, Inc. · 2000 · cites it directly9descendants
  • K020667Codman Hakim Shunt SystemsCodman & Shurtleff, Inc. · 2002 · 2 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K902783PS Medical Delta CSF-Flow Control ValvePudenz-Schulte Medical Research Corp.1990—
+1K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical2002—
+2K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without…Medtronic Neurosurgery20042
+2K042465PS Medical Strata II ValveMedtronic Neurosurgery20044
+1K992173Siphonguard CSF Control DeviceJohnson & Johnson Professionals, Inc.2000—
+2K020667Codman Hakim Shunt SystemsCodman & Shurtleff, Inc.2002—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200230
200430
200520
200630
200710
200820
200920
201020
201120
201230
201310
201520
201611
201720
201820
201910
202010
202110
202210
202420
202510
FIG. 1 — Lineage of K902783, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (39)

See all 39

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical200225
K992173Siphonguard CSF Control DeviceJohnson & Johnson Professionals, Inc.20009
K020728Miethke Shunt SystemAesculap, Inc.20022
K042198PS Medical Neurosurgical ImplantsMedtronic Neurosurgery20040

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 11, 2021Class IIZ-1115-2021Process controlPotential for variation in radiopaque marking visibility under radiographic imaging for adjustable and fixed-pressure valves that are used in the management of hydrocephalus
Dec 16, 2020Class IIZ-0871-2021Device DesignNeonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of…
Dec 16, 2020Class IIZ-0872-2021Device DesignNeonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of…
Dec 16, 2020Class IIZ-0873-2021Device DesignNeonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of…
Nov 18, 2013Class IIZ-0398-2014OtherMedtronic Neurosurgery is recalling the CSF-Unitized Shunts, Delta Shunt Assembly Kit, and the Strata II Shunt Assembly Kit because they were found to have label discrepancies.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152700StrataMR Valves and ShuntsMedtronic Neurosurgery · 2016 · 3 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K012052 (Medtronic PS Medical, 2002) has gone furthest, with 25 descendants of its own.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—
Cite this page

510(k) Report. “K902783: PS Medical Delta CSF-Flow Control Valve.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K902783

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.