510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

NM/CT 850, NM/CT 860

K173816510(k) TraditionalCleared Apr 17, 2018123-day reviewClass II21 CFR 892.1200Recalls 10 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · KPS System, Tomography, Computed, Emission

Cleared on Apr 17, 2018 as substantially equivalent to K093514 (Ge Medical Systems F.I. Haifa, 2009) after a 123-day review. Its family tree reaches back 8 generations to 1995. 6 of the 32 devices in its family, counting itself, have a Class I recall. It has been recalled 10 times, including a Class I recall.

Decided
Apr 17, 2018
Review
123 days
Class
Class II
Product code
KPS
Recalls
10 (1 Class I)
Predicates
1

Its longest line of descent runs back 8 generations to K952674 (1995). 31 ancestors in all, the oldest K892358 (1989). No later device cites it as a predicate yet. 6 of the 32 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2008: 1 other ancestor2005: 3 other ancestors, 1 with a Class I recall12002: 2 other ancestors, 1 with a Class I recall12004: 2 other ancestors1997: 2 other ancestors1995: 2 other ancestors, 1 with a Class I recall12003: 2 other ancestors2001: 2 other ancestors2000: 4 other ancestors1999: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestorReference device: K121019 · Optima NM/CT 640 · GE Healthcare · 2012 · Class I recallReference device: K171013 · Revolution ACT · Ge Hangwei Medical Systems Co., Ltd. · 2017Reference device: K153355 · Xeleris 4.0 Processing and Review Workstation · GE Healthcare · 2016K952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK023190 · E.Cam Computer / E.Soft Workstation · Siemens Medical Solutions USA, Inc. · 2002K063826 · Syngo MI Applications 2007A · Siemens Medical Solutions USA, Inc. · 2007K072567 · Symbia-E, Model 10275879 · Siemens Medical Solutions USA, Inc. · 2007 · recalledK082506 · Symbia, Version 4.0 · Siemens Medical Solutions USA, Inc. · 2008 · recalledK093514 · Discovery Model NM/CT 670 · Ge Medical Systems F.I. Haifa · 2009 · Class I recall19851990199520002005201020152020
  1. Root · 8 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 7 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 6 generations backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. 5 generations backK023190 E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc. · 2002
  5. 4 generations backK063826 Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc. · 2007
  6. 3 generations backK072567 Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc. · 2007 · Recalled
  7. 2 generations backK082506 Symbia, Version 4.0Siemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 1 generation backK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
  9. This deviceK173816 NM/CT 850, NM/CT 860GE Healthcare · 2018

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−7K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−6K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
−5K023190E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc.2002—
−4K063826Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc.2007—
−3K072567Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc.20072
−2K082506Symbia, Version 4.0Siemens Medical Solutions USA, Inc.20089
−1K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa2009Class I
0 (this device)K173816NM/CT 850, NM/CT 860GE Healthcare2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199110
199521
199720
199910
200040
200120
200221
200320
200420
200531
200810
FIG. 1 — Lineage of K173816, 1995–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093514Discovery Model NM/CT 670
Summary page 7
…Premarket Notification Submission Marketed Devices NM/CT 850 and NM/CT 860 are a modification to the predicate device Discovery NM/CT 670 (K093514). The modification is mainly for replacing the CT subsystem with GE’s 8 slice Revolution ACTs…
Ge Medical Systems F.I. Haifa2009Class I
†K121019Optima NM/CT 640GE Healthcare2012Class I
†K171013Revolution ACTGe Hangwei Medical Systems Co., Ltd.2017—
†K153355Xeleris 4.0 Processing and Review WorkstationGE Healthcare2016—

Also cleared against K093514: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 19, 2022Class IZ-1058-2023Process controlA mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Feb 7, 2021Class IIZ-1546-2021Device DesignGE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific…
Feb 7, 2021Class IIZ-1545-2021Device DesignGE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific…
Dec 10, 2020Class IIZ-0854-2021Process controlGE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Dec 10, 2020Class IIZ-0853-2021Process controlGE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Aug 14, 2020Class IIZ-2992-2020Use errorRotor bearing screws were found loose on detector.
Aug 14, 2020Class IIZ-2991-2020Use errorRotor bearing screws were found loose on detector.
Mar 20, 2019Class IIZ-1263-2019Device DesignDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Mar 20, 2019Class IIZ-1262-2019Device DesignDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Mar 20, 2019Class IIZ-1261-2019Device DesignDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K052434Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems · 2005 · 2 generations back · ◆ in the 2005 ancestors bar of FIG. 11 Class I recall
  • K022960GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa · 2002 · 2 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K953801VaricamElscint, Inc. · 1995 · 3 generations back · ◆ in the 1995 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 2.7 years on median.
  • 21 of the 31 ancestors have a recall of any class; 5 have a Class I recall.
  • 3 other devices also cite its predicate K093514.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The GE NM/CT 850 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in the assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors; and…

Read the full statement (601 words)

The GE NM/CT 850 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in the assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors; and planning, guiding, and monitoring therapy.

NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). Imaging modes include single photon, multi-isotope, and multi peak frame, with data stored frame/list mode. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring.

CT System: Intended specifically for attenuation correction and anatomical localization.

NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based SPECT attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration, and fusion).

The GE NM/CT 850 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may include digital processing of data and images, including display, quality check, transfer, and processing, to produce images in a variety of trans-axial and reformatted planes. The images can also be post processed to obtain additional images, imaging planes, analysis results, and uptake quantitation. The system may be used for patients of all ages.

NM/CT 850 does not support standalone CT operation.

The GE NM/CT 860 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in the assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors; and planning, guiding, and monitoring therapy.

NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT), Imaging modes include single photon, multi-isotope, and multi-peak, with data stored in frame/list mode. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring.

CT System: produces Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine and Helical acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body and vascular X-Ray Computed Tomography applications

NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based SPECT attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration, and fusion).

The GE NM/CT 860 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may include digital processing of data and images, including display, quality check, transfer, and processing, to produce images in a variety of trans-axial and reformatted planes., The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation. The system may be used for patients of all ages.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173816: NM/CT 850, NM/CT 860.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173816

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026