Unicel DXC 600I Synchron Access Clinical System

K060256510(k) SpecialCleared Feb 27, 200626-day reviewClass II21 CFR 862.1345Recalls 24

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · CFR Hexokinase, Glucose

Cleared on Feb 27, 2006 as substantially equivalent to K023049 (Beckman Coulter, Inc., 2002) after a 26-day review. Its family tree reaches back 2 generations to 2001. 1 later device descends from it. It has been recalled 24 times.

Decided
Feb 27, 2006
Review
26 days
Class
Class II
Product code
CFR
Recalls
24
Predicates
1

Its longest line of descent runs back 2 generations to K011213 (2001). 1 later device descends from it, 1 citing it directly.

  1. Root · 2 generations backK011213 Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc. · 2001 · Recalled
  2. 1 generation backK023049 Synchron LX I 725 Clinical SystemBeckman Coulter, Inc. · 2002 · Recalled
  3. This deviceK060256 Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc. · 2006
  4. Descendant · cites it directly · 0 of its ownK103842 Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC…Beckman Coulter, Inc. · 2011 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800 Synchron System…Beckman Coulter, Inc. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K011213Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc.200111
−1K023049Synchron LX I 725 Clinical SystemBeckman Coulter, Inc.200218
0 (this device)K060256Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc.2006—
+1K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800…Beckman Coulter, Inc.20113

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
FIG. 1 — Lineage of K060256, 2001–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CFR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CFR device

  1. #1 K253490 Glucose HK Gen.3; ONLINE DAT Methadone II…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023049Synchron LX I 725 Clinical System
Summary page 1
…se [862.2160] 4.0 Legally Marketed Device The UniCel DxC 600i System claims substantial equivalence to the SYNCHRON LXi 725 System (Docket Number K023049) currently in commercial distribution. 5.0 Device Description The UniCel DxC 600i System combines the UniCel DxC 60…
Beckman Coulter, Inc.200218
†K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc.200434

Also cleared against K023049: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800 Synchron System…Beckman Coulter, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 26, 2023Class IIZ-1257-2023Component change controlAn increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a…
Nov 6, 2017Class IIIZ-0310-2018Device DesignPotential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Oct 6, 2014Class IIZ-0176-2015Vendor change controlBeckman Coulter is recalling the Small Form Factor Console PC (SFF PC) (B23083) for use with the DxC 600i analyzer because it may experience a "MFC Exception" error during normal operation of the analyzer system.
Apr 22, 2014Class IIZ-2272-2016Nonconforming Material/ComponentPossible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems.
Feb 12, 2013Class IIZ-1035-2013Device DesignBeckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley…
Nov 1, 2012Class IIZ-0612-2013Device DesignIt has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at…
Sep 3, 2012Class IIZ-0123-2013Software change controlBeckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned…
Jul 6, 2012Class IIZ-0059-2013Nonconforming Material/ComponentThe recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause…
Nov 16, 2010Not classifiedZ-1972-2011PendingThe hose clamps in the DxC (with serial numbers less than 3412, labeled behind the lower right side instrument door) hydropneumatic system are sharp and pose a possible risk of injury. An operator could be cut by the hose clamps, which…
Oct 5, 2010Not classifiedZ-1936-2011PendingThe recall was initiated because Beckman Coulter confirmed Access waste bottles that ruptured and expelled liquid waste, exposing the operator to the system liquid waste. Over time, waste bottles can degrade, become brittle and…
Aug 13, 2010Not classifiedZ-0607-2011PendingThis recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated as a result of…
Aug 10, 2010Not classifiedZ-1619-2011OtherThe recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be…
Apr 15, 2010Not classifiedZ-2388-2010OtherUniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
Feb 25, 2010Not classifiedZ-2590-2010OtherBeckman Coulter is initiating a recall on their UniCel DxC Clinical Systems (880i, 860i,680i and 660i Systems with serial number <2760 that have software version 4.9.01 installed and have the VME Motorola ICS board) because they may…
Jan 18, 2010Not classifiedZ-2442-2010OtherThe recall was initiated due to " A loose Aliquot Probe tube fitting in the Closed Tube Aliquotter (CTA), may result in erroneous Access immunoassay results or Synchron chemistry results". This may result in dilution of the sample with…
Jan 4, 2010Not classifiedZ-0863-2010Device DesignExcessive Buildup of protein, bacteria and sample tube additives in the ISE flow cell may cause erroneous NA (sodium) results.
Oct 28, 2009Not classifiedZ-0856-2010Software designThis recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and…
Jul 23, 2009Not classifiedZ-2633-2011PendingThe recall was initiated because the Wash Concentrate Tube Assembly 142 may fail to deliver the appropriate amount of wash concentrate on some SYNCHRON DxC Systems due to an accumulation of crystals. The impact of this issue would…
May 7, 2009Not classifiedZ-2771-2011Software designThe recall was initiated after Beckman Coulter confirmed reports of inaccurate Access Intrinsic Factor Ab test results displayed on, or printed from, the UniCel Dxl portion of an integrated workstation listed above. The problem is limited…
Oct 9, 2008Not classifiedZ-0914-2010OtherThe recall was initiated after Beckman Coulter confirmed reports of a new premature failure of the upper aspirate peristaltic pump in the systems listed above. Premature failure of the pump may negatively affect precision. The failure is…
Sep 11, 2008Not classifiedZ-1375-2012Software designThe recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Apr 3, 2008Not classifiedZ-2386-2008Nonconforming Material/ComponentThe field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively affect precision.
Oct 17, 2007Not classifiedZ-1183-2008Process controlElectrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Apr 5, 2007Not classifiedZ-1098-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 3.4 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K103842 (Beckman Coulter, Inc., 2011) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K023049.

How does its category compare?

All CFR devices

CFR Hexokinase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CFR

1in 2025

2007: 12008: 52009: 32011: 32012: 42013: 12014: 12016: 22017: 12019: 12022: 12025: 12026 (partial year): 1
2007high 5 (2008) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CFR

115in 2012

2008: 182 days median2009: 101 days median2011: 316 days median2012: 115 days median
2007high 316 (2011) · low 101 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CFR

1in 2016

2008: 72009: 12012: 12014: 12016: 1
2007high 7 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CFR

16in 2025

2007: 12008: 22009: 72010: 112011: 432012: 492013: 362014: 22015: 32016: 12017: 132018: 82019: 102020: 42021: 52022: 32023: 292024: 152025: 16
2007high 49 (2012) · low 1 (2007)2025

MAUDE reports naming product code CFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008518272
2009310117
20100—011
20113316043
20124115149
20131—036
20141—12
20150—03
20162—11
20171—013
20180—08
20191—010
20200—04
20210—05
20221—03
20230—029
20240—015
20251—016
20261—0—

What is it indicated for?

The UniCel® DxC 600i System combines the UniCel® DxC 600 analyzer and the Access® 2 analyzer into a single instrument presentation. Samples are loaded from a single point of entry through a Closed Tube Aliquoter (CTA) unit. The CTA functions as a sample processing manager by aliquotting and routing samples to the Access 2 and/or UniCel DxC 600 analyzer according to programming requirements. The UniCel DxC 600 Synchron® Clinical System is a fully…

Read the full statement (176 words)

The UniCel® DxC 600i System combines the UniCel® DxC 600 analyzer and the Access® 2 analyzer into a single instrument presentation. Samples are loaded from a single point of entry through a Closed Tube Aliquoter (CTA) unit. The CTA functions as a sample processing manager by aliquotting and routing samples to the Access 2 and/or UniCel DxC 600 analyzer according to programming requirements.

The UniCel DxC 600 Synchron® Clinical System is a fully automated, computer-controlled clinical chemistry analyzer intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid, (sample type is chemistry dependent).

The Access 2 Immunoassay Analyzer is a microcomputer controlled, random access instrument. The analyzer performs enzyme immunoassays utilizing paramagnetic in particle solid phase and chemiluminescent detection. The Access 2 Analyzer is intended for the in vitro determination of a variety of analytes of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid, (sample type is chemistry dependent).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060256: Unicel DXC 600I Synchron Access Clinical System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K060256

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026