510(k) · Cleared
Recalled · Class I · 2019Forte Gamma Camera System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Adac Laboratories · KPS System, Tomography, Computed, Emission
Cleared on Oct 22, 2003 as substantially equivalent to K982911 (Adac Laboratories, 1998) after a 14-day review. Its family tree reaches back 1 generation to 1998. 4 of the 5 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 4 times, including 3 Class I recalls.
- Decided
- Oct 22, 2003
- Review
- 14 days
- Class
- Class II
- Product code
- KPS
- Recalls
- 4 (3 Class I)
- Predicates
- 1
Its longest line of descent runs back 1 generation to K982911 (1998). 3 later devices descend from it, 1 citing it directly. 4 of the 5 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK982911 FORTEAdac Laboratories · 1998 · Recalled (Class I)
- This deviceK033254 Forte Gamma Camera SystemAdac Laboratories · 2003
- Descendant · cites it directly · 2 of its ownK062298 Apollo Gamma Camera SystemAdac Laboratories · 2006 · Recalled (Class I)
- Descendant · 2 generations on · 1 of its ownK080927 Brightview VCT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2008 · Recalled (Class I)
- Descendant · 3 generations on · 0 of its ownK083649 Cardius 3 X-ActDigirad Corp. · 2009
Class I recalls in this family
Most influential descendants
Showing 3 of 3 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K982911 | FORTE | Adac Laboratories | 1998 | Class I |
| 0 (this device) | K033254 | Forte Gamma Camera System | Adac Laboratories | 2003 | — |
| +1 | K062298 | Apollo Gamma Camera System | Adac Laboratories | 2006 | Class I |
| +2 | K080927 | Brightview VCT Imaging System | Philips Medical Systems (Cleveland), Inc. | 2008 | Class I |
| +3 | K083649 | Cardius 3 X-Act | Digirad Corp. | 2009 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 1 |
| 2008 | 1 | 1 |
| 2009 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPS device
- #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.7 · 17 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K982911 | FORTESummary page 1…90 KPS C. Date prepared: September 18,2003 D. Predicate Device: ForteTMGamma Camera System (K982911) E. Intended Use: The Forte Gamma Camera System is intended to produce images depicting the anatomical distributions of s… | 1998 |
What descends from it? (3)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K062298 | Apollo Gamma Camera System Class I | 2006 | 2 |
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 19, 2019 | Class I | Z-0214-2020 | Nonconforming Material/Component | An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit. |
| Sep 19, 2019 | Class I | Z-0217-2020 | Nonconforming Material/Component | An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit. |
| Sep 19, 2019 | Class I | Z-0215-2020 | Nonconforming Material/Component | An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit. |
| Mar 6, 2015 | Class II | Z-1777-2015 | Nonconforming Material/Component | The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K982911FORTEAdac Laboratories · 1998 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
- K062298Apollo Gamma Camera SystemAdac Laboratories · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
- K080927Brightview VCT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
How does its category compare?
All KPS devicesKPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KPS
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPS
65in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPS
16in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · KPS
36in 2025
MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 38 | 22 | 38 |
| 2008 | 17 | 25 | 16 | 11 |
| 2009 | 9 | 42 | 14 | 15 |
| 2010 | 3 | 91 | 4 | 44 |
| 2011 | 7 | 28 | 34 | 56 |
| 2012 | 6 | 94 | 13 | 34 |
| 2013 | 11 | 84 | 13 | 240 |
| 2014 | 11 | 92 | 18 | 33 |
| 2015 | 3 | 112 | 5 | 24 |
| 2016 | 11 | 65 | 16 | 10 |
| 2017 | 11 | 104 | 11 | 31 |
| 2018 | 13 | 59 | 12 (3) | 26 |
| 2019 | 3 | 60 | 5 (5) | 105 |
| 2020 | 12 | 72 | 8 | 22 |
| 2021 | 15 | 66 | 4 | 10 |
| 2022 | 5 | 60 | 8 (6) | 15 |
| 2023 | 12 | 63 | 6 (3) | 22 |
| 2024 | 11 | 59 | 15 | 23 |
| 2025 | 12 | 65 | 16 | 36 |
| 2026 | 6 | 67 | 6 | — |
What is it indicated for?
The ForteTMGamma Camera System is intended to produce images depicting the
anatomical distributions of single photon and positron emitting radioisotopes within the
human body for interpretation by medical personnel.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K033254 · printed Nov 3, 2026
Cite this page
510(k) Report. “K033254: Forte Gamma Camera System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033254
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026