510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Brightview VCT Imaging System

K080927510(k) TraditionalCleared Apr 11, 20089-day reviewClass II21 CFR 892.1200Recalls 30 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · KPS System, Tomography, Computed, Emission

Cleared on Apr 11, 2008 as substantially equivalent to K062298 (Adac Laboratories, 2006) and 1 other predicate after a 9-day review. Its family tree reaches back 4 generations to 1995. 6 of the 9 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 30 times, including 2 Class I recalls.

Decided
Apr 11, 2008
Review
9 days
Class
Class II
Product code
KPS
Recalls
30 (2 Class I)
Predicates
2

Its longest line of descent runs back 4 generations to K952674 (1995). 7 ancestors in all. 1 later device descends from it, 1 citing it directly. 6 of the 9 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall11998: 1 other ancestor, 1 with a Class I recall12009: 1 descendant clearedK952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK052434 · Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera · GE Medical Systems · 2005 · Class I recall1995200020052010201520202025
  1. Root · 4 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 3 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 2 generations backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. 1 generation backK052434 Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems · 2005 · Recalled (Class I)
  5. This deviceK080927 Brightview VCT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2008
  6. Descendant · cites it directly · 0 of its ownK083649 Cardius 3 X-ActDigirad Corp. · 2009

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K083649Cardius 3 X-ActDigirad Corp. · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−3K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−2K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
−1K052434Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems2005Class I
0 (this device)K080927Brightview VCT Imaging SystemPhilips Medical Systems (Cleveland), Inc.2008—
+1K083649Cardius 3 X-ActDigirad Corp.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199811
200311
200611

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K080927, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062298Apollo Gamma Camera System
Summary page 1
…duct Code: 90 KPS C. Date prepared: March 19, 2008 D. Predicate Devices: BrightViewTM Gamma Camera System (K062298) GE 4 Option for Dual-Head Variable Angle Gamma…
Adac Laboratories2006Class I
*K052434Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera
Summary page 1
…TM Gamma Camera System (K062298) GE 4 Option for Dual-Head Variable Angle Gamma Camera (K052434) E. Intended Use: BrightViewTm VCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates…
GE Medical Systems2005Class I

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083649Cardius 3 X-ActDigirad Corp.20090

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 15, 2023Class IZ-0930-2024Component design/selectionDetector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an…
Dec 15, 2023Class IZ-0931-2024Component design/selectionDetector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an…
Nov 30, 2023Class IIZ-0679-2024Device DesignWhile using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture…
Nov 30, 2023Class IIZ-0678-2024Device DesignWhile using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture…
Jun 10, 2019Class IIZ-2011-2019Under Investigation by firmAn issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Jun 10, 2019Class IIZ-2010-2019Under Investigation by firmAn issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Jun 10, 2019Class IIZ-2009-2019Under Investigation by firmAn issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.
Feb 24, 2017Class IIZ-1752-2017Nonconforming Material/ComponentThe brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Feb 24, 2017Class IIIZ-1753-2017Nonconforming Material/ComponentThe brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Feb 24, 2017Class IIZ-1750-2017Nonconforming Material/ComponentThe brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Feb 24, 2017Class IIZ-1751-2017Nonconforming Material/ComponentThe brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Jan 31, 2017Class IIZ-1481-2017Software designFour issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar…
Jan 31, 2017Class IIZ-1479-2017Software designFour issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar…
Jan 31, 2017Class IIZ-1480-2017Software designFour issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar…
Dec 23, 2016Class IIZ-1187-2017Under Investigation by firmThe intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.
Dec 23, 2016Class IIZ-1186-2017Under Investigation by firmThe intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.
May 25, 2016Class IIZ-2736-2016Software designThe firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
May 25, 2016Class IIZ-2737-2016Software designThe firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
Dec 17, 2015Class IIZ-1022-2016Software designThe firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of…
Mar 6, 2015Class IIZ-1792-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Nov 12, 2014Class IIZ-1011-2015Software designUnintended detector and gantry movement due to software issues.
Nov 4, 2014Class IIZ-0450-2015Software designSoftware issues
Jun 23, 2014Class IIZ-1982-2014Device DesignFlat Panel Display (FPD) failed to remain securely locked in the deployed position.
Dec 19, 2013Class IIZ-1074-2014Software designThree problems were reported to the firm: When positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly causing an inappropriate rotation orbit calculation. When the…
Oct 23, 2013Class IIZ-0203-2015Radiation Control for Health and Safety ActIt was discovered the CT portion of the scan was interrupted ( stopping the CT exposure at the time of interruption) and did not complete successfully.
Sep 3, 2013Class IIZ-0089-2014Device DesignPhilips received reports from the field that the Flat Panel Detector (FPD) does not always remain securely latched in the stowed position.
Jun 15, 2011Not classifiedZ-2948-2011Software designNon-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.
Jun 15, 2011Not classifiedZ-2949-2011Software designNon-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.
Mar 28, 2011Not classifiedZ-2155-2011Device DesignThe current Headholder to support Bright View XCT brain imaging allowed for imaging the brain but doe allow for imaging of the associated shoulder/neck area of the body.
Mar 28, 2011Not classifiedZ-2156-2011Device DesignThe current Headholder to support Bright View XCT brain imaging allowed for imaging the brain but doe allow for imaging of the associated shoulder/neck area of the body.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062298Apollo Gamma Camera SystemAdac Laboratories · 2006 · Direct predicate · ◆ in the 2006 ancestors bar of FIG. 11 Class I recall
  • K052434Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems · 2005 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K033254Forte Gamma Camera SystemAdac Laboratories · 2003 · 2 generations back · ◆ in the 2003 ancestors bar of FIG. 13 Class I recalls
  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K982911FORTEAdac Laboratories · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.7 years on median.
  • 6 of the 7 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K083649 (Digirad Corp., 2009) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

BrightView VCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView VCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and the VCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data. The BrightView VCT Imaging System should only be used by trained healthcare professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080927: Brightview VCT Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080927

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.