The Reliance Endoscope Processing System

K040049510(k) TraditionalCleared Dec 17, 2004340-day reviewClass II21 CFR 876.1500Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · NZA Accessories, Germicide, Cleaning, For Endoscopes

Cleared on Dec 17, 2004 after a 340-day review. The FDA summary names its predicate without a number. 8 later devices descend from it. It has been recalled 2 times.

Decided
Dec 17, 2004
Review
340 days
Class
Class II
Product code
NZA
Recalls
2
Predicates
0

8 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2011: 1 descendant cleared2013: 1 descendant cleared2020: 1 descendant cleared2023: 2 descendants cleared2021: 1 descendant cleared2022: 1 descendant cleared200020052010201520202025
  1. This deviceK040049 The Reliance Endoscope Processing SystemSTERIS Corporation · 2004
  2. Descendant · cites it directly · 7 of its ownK102244 Reliance Endoscope Processing SystemSTERIS Corporation · 2010 · Recalled
  3. Descendant · 2 generations on · 2 of its ownK200989 Reliance Endoscope Processing SystemSTERIS Corporation · 2020
  4. Descendant · 2 generations on · 2 of its ownK123768 Reliance Advance Endoscope Processing SystemSTERIS Corporation · 2013
  5. Descendant · 3 generations on · 1 of its ownK230560 enspire 300 Series Automated Endoscope Reprocessor SystemSTERIS Corporation · 2023
  6. Descendant · 3 generations on · 1 of its ownK203223 Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSteris · 2021

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12013: 12020: 12021: 12022: 12023: 2
2010high 2 (2023) · low 0 (2012)2026

Most influential descendants

Showing 5 of 8 · See all
  • K102244Reliance Endoscope Processing SystemSTERIS Corporation · 2010 · cites it directly7descendants
  • K200989Reliance Endoscope Processing SystemSTERIS Corporation · 2020 · 2 generations on2descendants
  • K123768Reliance Advance Endoscope Processing SystemSTERIS Corporation · 2013 · 2 generations on2descendants
  • K230560enspire 300 Series Automated Endoscope Reprocessor SystemSTERIS Corporation · 2023 · 3 generations on1descendant
  • K203223Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSteris · 2021 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K040049The Reliance Endoscope Processing SystemSTERIS Corporation2004—
+1K102244Reliance Endoscope Processing SystemSTERIS Corporation20101
+2K200989Reliance Endoscope Processing SystemSTERIS Corporation2020—
+2K123768Reliance Advance Endoscope Processing SystemSTERIS Corporation2013—
+3K230560enspire 300 Series Automated Endoscope Reprocessor SystemSTERIS Corporation2023—
+3K203223Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSteris2021—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201310
202010
202110
202210
202320
FIG. 1 — Lineage of K040049, 2004–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NZA device

  1. #1 K230560 enspire 300 Series Automated Endoscope…score 9.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Custom Ultrasonic System 83 Plus
Named without a submission number
—
*—STERIS 20 Sterilant Concentrate
Named without a submission number
—
*—Sterilox Liquid Chemical HLD System
Named without a submission number
—
†K921559STERIS PROCESS Indicator Chemical Monitor—

What descends from it? (8)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102244Reliance Endoscope Processing SystemSTERIS Corporation20107

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 29, 2013Class IIZ-2263-2013Use errorThe air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG)…
Jul 2, 2010Not classifiedZ-0823-2011OtherReliance EPS had internal water leaks and interrupted cycles when air leaks triggered false Fault 38 alarm that DG/HLD cup has not been replaced.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K102244 (STERIS Corporation, 2010) has gone furthest, with 7 descendants of its own.

How does its category compare?

All NZA devices

NZA Accessories, Germicide, Cleaning, For Endoscopes: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NZA

2in 2023

2010: 12011: 12013: 12020: 12021: 12022: 12023: 2
2007high 2 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NZA

Too few NZA 510(k)s in any year since 2007 to give a median.

Recalls per year · NZA

No recall names NZA since 2007.

Adverse event reports per year · NZA

2in 2022

2010: 12011: 12012: 142013: 112014: 62015: 102016: 212017: 92018: 82019: 22020: 12021: 42022: 2
2007high 21 (2016) · low 0 (2007)2025

MAUDE reports naming product code NZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—01
20111—01
20120—014
20131—011
20140—06
20150—010
20160—021
20170—09
20180—08
20190—02
20201—01
20211—04
20221—02
20232—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as Gl flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 -57°C HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6 minute exposure sequence).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K040049: The Reliance Endoscope Processing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K040049

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026