510(k) Report

510(k) · Cleared

Cleared

Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide

K220361510(k) SpecialCleared Mar 10, 202230-day reviewClass II21 CFR 876.1500

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · NZA Accessories, Germicide, Cleaning, For Endoscopes

Cleared on Mar 10, 2022 as substantially equivalent to K203223 (Steris, 2021) after a 30-day review. Its family tree reaches back 4 generations to 2004.

Decided
Mar 10, 2022
Review
30 days
Class
Class II
Product code
NZA
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K040049 (2004). No later device cites it as a predicate yet.

  1. Root · 4 generations backK040049 The Reliance Endoscope Processing SystemSTERIS Corporation · 2004 · Recalled
  2. 3 generations backK102244 Reliance Endoscope Processing SystemSTERIS Corporation · 2010 · Recalled
  3. 2 generations backK200989 Reliance Endoscope Processing SystemSTERIS Corporation · 2020
  4. 1 generation backK203223 Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSteris · 2021
  5. This deviceK220361 Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSTERIS Corporation · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K040049The Reliance Endoscope Processing SystemSTERIS Corporation20042
−3K102244Reliance Endoscope Processing SystemSTERIS Corporation20101
−2K200989Reliance Endoscope Processing SystemSTERIS Corporation2020—
−1K203223Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSteris2021—
0 (this device)K220361Reliance EPS Endoscope Processing System and Reliance DG Dry GermicideSTERIS Corporation2022—
FIG. 1 — Lineage of K220361, 2004–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NZA device

  1. #1 K230560 enspire 300 Series Automated Endoscope…score 9.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K203223Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide
Summary page 5
…Cleaning, For Endoscopes Classification Number: 21 CFR 876.1500 Product Code: NZA 2. Predicate Device Reliance Endoscope Processing System, K203223 3. Description of Device The Reliance Endoscope Processing System is a high level disinfection system that can was…
Steris2021—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 2 companies; a new generation was cleared every 6.0 years on median.
  • 2 of the 4 ancestors have a recall of any class.

How does its category compare?

All NZA devices

NZA Accessories, Germicide, Cleaning, For Endoscopes: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · NZA

2in 2023

2010: 12011: 12013: 12020: 12021: 12022: 12023: 2
2007high 2 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NZA

Too few NZA 510(k)s in any year since 2007 to give a median.

Recalls per year · NZA

No recall names NZA since 2007.

Adverse event reports per year · NZA

2in 2022

2010: 12011: 12012: 142013: 112014: 62015: 102016: 212017: 92018: 82019: 22020: 12021: 42022: 2
2007high 21 (2016) · low 0 (2007)2025

MAUDE reports naming product code NZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—01
20111—01
20120—014
20131—011
20140—06
20150—010
20160—021
20170—09
20180—08
20190—02
20201—01
20211—04
20221—02
20232—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as bronchoscopes, GI flexible endoscopes including duodenoscopes, and their accessories. High level disinfection is achieved within the 50-57°C HLD phase of the endoscope processing cycle (4-minute generation sequence followed by a 6-minute exposure sequence).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K220361: Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K220361

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026