510(k) Report

510(k) · Cleared

Recalled · 2010

PAINPUMP1; PAINPUMP2

K042405510(k) TraditionalCleared Oct 14, 200441-day reviewClass II21 CFR 880.5725Recalls 70

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Corp. · FRN Pump, Infusion

Cleared on Oct 14, 2004 as substantially equivalent to K031249 (Stryker Instruments, 2003) and 2 other predicates after a 41-day review. Its family tree reaches back 1 generation to 2003. 3 later devices descend from it. It has been recalled 70 times.

Decided
Oct 14, 2004
Review
41 days
Class
Class II
Product code
FRN
Recalls
70
Predicates
3

Its longest line of descent runs back 1 generation to K030661 (2003). 3 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2003: 1 other ancestor2005: 2 descendants cleared2008: 1 descendant clearedK030661 · Stryker PAINPUMP2 · Stryker Instruments · 2003 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 1 descendant K030661 · Stryker PAINPUMP2 · Stryker Instruments · 2003 · recalled0
  1. Root · 1 generation backK030661 Stryker PAINPUMP2Stryker Instruments · 2003 · Recalled
  2. This deviceK042405 PAINPUMP1; PAINPUMP2Stryker Corp. · 2004
  3. Descendant · cites it directly · 1 of its ownK043466 PAINPUMP2Stryker Instruments · 2005 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK072702 Stryker Antimic CatheterStryker Instruments, Instruments Div. · 2008
  5. Descendant · cites it directly · 0 of its ownK052221 Ambit Push & Play Ambulatory Infusion PumpSorenson Medical, Inc. · 2005 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K043466PAINPUMP2Stryker Instruments · 2005 · cites it directly1descendant
  • K072702Stryker Antimic CatheterStryker Instruments, Instruments Div. · 2008 · 2 generations on0descendants
  • K052221Ambit Push & Play Ambulatory Infusion PumpSorenson Medical, Inc. · 2005 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K030661Stryker PAINPUMP2Stryker Instruments20033
0 (this device)K042405PAINPUMP1; PAINPUMP2Stryker Corp.2004—
+1K043466PAINPUMP2Stryker Instruments200572
+2K072702Stryker Antimic CatheterStryker Instruments, Instruments Div.2008—
+1K052221Ambit Push & Play Ambulatory Infusion PumpSorenson Medical, Inc.20054

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200520
200810
FIG. 1 — Lineage of K042405, 2003–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031249Modification to Stryker Painpump
Summary page 1
…Infusion Pump; Electromechanical Ambulatory Infusion Pump Classification Name: Pump, Infusion, PCA Equivalent to: K031249, Stryker PainPumpl; K030661, PainPump2, K040337; I-Flow Elastomeric Pump w…
Stryker Instruments2003—
*K030661Stryker PAINPUMP2
Summary page 1
…ctromechanical Ambulatory Infusion Pump Classification Name: Pump, Infusion, PCA Equivalent to: K031249, Stryker PainPumpl; K030661, PainPump2, K040337; I-Flow Elastomeric Pump w/ Bolus Device Description:…
Stryker Instruments20033
*K040337I-Flow Elastomeric Pump (on-Q, Painbuster, C-Bloc, Homepump)
Summary page 1
…nfusion Pump Classification Name: Pump, Infusion, PCA Equivalent to: K031249, Stryker PainPumpl; K030661, PainPump2, K040337; I-Flow Elastomeric Pump w/ Bolus Device Description: PainPump]…
I-Flow Corp.2004—

Also cleared against K031249: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K043466PAINPUMP2Stryker Instruments20051
K052221Ambit Push & Play Ambulatory Infusion PumpSorenson Medical, Inc.20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 2, 2010Not classifiedZ-2592-2010Nonconforming Material/ComponentAll of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a…
Sep 2, 2010Not classifiedZ-2593-2010Nonconforming Material/ComponentAll of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a…
Sep 2, 2010Not classifiedZ-2591-2010Nonconforming Material/ComponentAll of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a…
Sep 2, 2010Not classifiedZ-2594-2010Nonconforming Material/ComponentAll of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a…
Aug 16, 2010Not classifiedZ-0096-2011OtherThe instructions for use (IFU) has been revised for health care professionals. The indication for use has been changed to remove site specific pain management. The firm has revised the warning to not use the pump around joint spaces. A…
Jun 16, 2009Not classifiedZ-1773-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1766-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1743-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1749-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1774-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1747-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1786-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1736-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1764-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1742-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1741-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1784-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1770-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1785-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1781-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1756-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1744-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1758-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1772-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1755-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1757-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1761-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1746-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1779-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1765-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1771-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1739-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1737-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1775-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1760-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1769-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1752-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1759-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1740-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1788-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1780-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1763-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1750-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1777-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1776-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1738-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1787-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1745-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1762-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
Jun 16, 2009Not classifiedZ-1778-2009Nonconforming Material/ComponentThe catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K043466 (Stryker Instruments, 2005) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K031249.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

PainPump1 delivers controlled amounts of medication directly to the intraoperative site for pain management and/or antibiotic administration. The pump infuses the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes. PainPumpl is also intended for controlled delivery of local anesthetics in close proximity to nerves for postoperative regional anesthesia and pain…

Read the full statement (213 words)

PainPump1 delivers controlled amounts of medication directly to the intraoperative site for pain management and/or antibiotic administration. The pump infuses the medication at an hourly flow rate. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes.

PainPumpl is also intended for controlled delivery of local anesthetics in close proximity to nerves for postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural, or percutaneous.

PainPumpi is also intended to significantly decrease narcotic use and pain when used to deliver local anesthetics to surgical wound sites or close proximity to nerves when compared with narcotic only pain management.

PainPump2 delivers controlled amounts of medication directly to the intraoperative site for pain management and/or antibiotic administration. The pump infuses the medication at an hourly flow rate and provides the option for patient controlled bolus doses. Medications are infused intraoperatively and postoperatively through intramuscular or subcutaneous routes.

PainPump2 is also intended for controlled delivery of local anesthetics in close proximity to nerves for postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural, or percutaneous.

PainPump2 is also intended to significantly decrease narcotic use and pain when used to deliver local anesthetics to surgical wound sites or close proximity to nerves when compared with narcotic only pain management.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042405: PAINPUMP1; PAINPUMP2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042405

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.