Fabius GS/Fabius Tiro Anesthesia System

K042419510(k) TraditionalCleared Oct 8, 200431-day reviewClass II21 CFR 868.5160Recalls 11 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Oct 8, 2004 as substantially equivalent to K042086 (Draeger Medical, Inc., 2004) and 2 other predicates after a 31-day review. Its family tree reaches back 6 generations to 1990. 3 of the 24 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 11 times, including 4 Class I recalls.

Decided
Oct 8, 2004
Review
31 days
Class
Class II
Product code
BSZ
Recalls
11 (4 Class I)
Predicates
3

Its longest line of descent runs back 6 generations to K901713 (1990). 22 ancestors in all, the oldest K900598 (1990). 1 later device descends from it, 1 citing it directly. 3 of the 24 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors, 1 with a Class I recall11996: 2 other ancestors1998: 2 other ancestors1995: 1 other ancestor1994: 2 other ancestors1991: 1 other ancestor2004: 1 other ancestor2002: 1 other ancestor2000: 2 other ancestors1993: 1 other ancestor1990: 1 other ancestor2008: 1 descendant cleared, 1 with a Class I recall1K901713 · Narkomed 4 Anesthesia System · Draeger Medical, Inc. · 1990K980553 · Narkomed 6000 Model Number NM6000 · Draeger Medical, Inc. · 1998K993826 · Narkomed 6000 · Draeger Medical, Inc. · 2000×3: 3 generations cleared close together, K033498 (2004), K041622 (2004), K042086 (2004) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 7 other ancestors3 generations back: 2 other ancestors4 generations back: 4 other ancestors5 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall1 K042086 · Fabius GS Anesthesia System; Fabius Tiro Anesthesia System · Draeger Medical, Inc. · 2004K041622 · Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia System · Draeger Medical, Inc. · 2004K033498 · Narkomed 6400, Model NM6400 · Draeger Medical, Inc. · 2004 · recalledK993826 · Narkomed 6000 · Draeger Medical, Inc. · 2000K980553 · Narkomed 6000 Model Number NM6000 · Draeger Medical, Inc. · 1998K901713 · Narkomed 4 Anesthesia System · Draeger Medical, Inc. · 1990−50
  1. Root · 6 generations backK901713 Narkomed 4 Anesthesia SystemDraeger Medical, Inc. · 1990
  2. 5 generations backK980553 Narkomed 6000 Model Number NM6000Draeger Medical, Inc. · 1998
  3. 4 generations backK993826 Narkomed 6000Draeger Medical, Inc. · 2000
  4. 3 generations backK033498 Narkomed 6400, Model NM6400Draeger Medical, Inc. · 2004 · Recalled
  5. 2 generations backK041622 Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia…Draeger Medical, Inc. · 2004
  6. 1 generation backK042086 Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004
  7. This deviceK042419 Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004
  8. Descendant · cites it directly · 0 of its ownK072884 Fabius MriDraeger Medical AG & Co. KG · 2008 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K072884Fabius MriDraeger Medical AG & Co. KG · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K901713Narkomed 4 Anesthesia SystemDraeger Medical, Inc.1990—
−5K980553Narkomed 6000 Model Number NM6000Draeger Medical, Inc.1998—
−4K993826Narkomed 6000Draeger Medical, Inc.2000—
−3K033498Narkomed 6400, Model NM6400Draeger Medical, Inc.20041
−2K041622Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
−1K042086Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
0 (this device)K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
+1K072884Fabius MriDraeger Medical AG & Co. KG2008Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199110
199310
199420
199510
199620
199820
200020
200210
200321
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200811
FIG. 1 — Lineage of K042419, 1990–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042086Fabius GS Anesthesia System; Fabius Tiro Anesthesia System
Summary page 1
…lvania 18969 USA Establishment Registration No.: 2517967 Devices to which substantial equivalence is claimed: Fabius GS/Fabius Tiro Anesthesia System - K042086 Evita 4 Continuous Ventilator - K961687 7900 Ventilator -…
Draeger Medical, Inc.2004—
*K961687EVITA 4
Summary page 1
…ich substantial equivalence is claimed: Fabius GS/Fabius Tiro Anesthesia System - K042086 Evita 4 Continuous Ventilator - K961687 7900 Ventilator - K023366 Device Description: The Fabius GS/Fabius Tiro is a continuous flow…
Drager, Inc.1996—
*K023366Datex-Ohmeda 7900 Ventilator Enhancements
Summary page 1
…Fabius Tiro Anesthesia System - K042086 Evita 4 Continuous Ventilator - K961687 7900 Ventilator - K023366 Device Description: The Fabius GS/Fabius Tiro is a continuous flow gas anesthesia system. Intended Use: The Fabius GS/Fabius Tiro…
Datex-Ohmeda, Inc.2003Class I

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072884Fabius Mri Class IDraeger Medical AG & Co. KG20080

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2018Class IIZ-2251-2018Process change controlThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes…
May 24, 2018Class IIZ-2250-2018Process change controlThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes…
Mar 1, 2018Class IZ-1484-2018Process controlThe automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Mar 1, 2018Class IZ-1482-2018Process controlThe automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Mar 1, 2018Class IZ-1483-2018Process controlThe automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Aug 14, 2013Class IZ-0008-2014Device DesignDuring final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit housing was not…
Aug 26, 2010Not classifiedZ-2479-2010Otherpotential failures to Electri-Cord Manufacturing AC power cord
Aug 26, 2010Not classifiedZ-2478-2010Otherpotential failures to Electri-Cord Manufacturing AC power cord
Jul 7, 2008Not classifiedZ-2338-2008Process controlThe parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.
Feb 5, 2008Not classifiedZ-1315-2008Nonconforming Material/ComponentDetachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.
Oct 3, 2006Not classifiedZ-0048-2007OtherOne of four casters may break loose from chassis

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Direct predicate · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K072884Fabius MriDraeger Medical AG & Co. KG · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 14 years and passes through 1 company; a new generation was cleared every 1.6 years on median.
  • 5 of the 22 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K072884 (Draeger Medical AG & Co. KG, 2008) has gone furthest, with 0 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Fabius GS/Fabius Tiro is indicated as a continuous flow anesthesia system. The Fabius GS/Fabius Tiro can be used for spontaneous, manually assisted, automatic, pressure support, or synchronized mandatory intermittent ventilation, delivery of gases and anesthetic vapor, and monitoring oxygen concentration, breathing pressure and respiratory volume of patients during anesthesia. Federal law restricts this device to sale by or on the order of a physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042419: Fabius GS/Fabius Tiro Anesthesia System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042419

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026