510(k) Report

510(k) · Cleared

Recalled · Class I · 2021

Fabius Mri

K072884510(k) TraditionalCleared Jul 30, 2008295-day reviewClass II21 CFR 868.5160Recalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical AG & Co. KG · BSZ Gas-Machine, Anesthesia

Cleared on Jul 30, 2008 as substantially equivalent to K042419 (Draeger Medical, Inc., 2004) and 1 other predicate after a 295-day review. Its family tree reaches back 7 generations to 1990. 3 of the 25 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times, including a Class I recall.

Decided
Jul 30, 2008
Review
295 days
Class
Class II
Product code
BSZ
Recalls
4 (1 Class I)
Predicates
2

Its longest line of descent runs back 7 generations to K901713 (1990). 24 ancestors in all, the oldest K900598 (1990). No later device cites it as a predicate yet. 3 of the 25 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2005: 1 other ancestor2003: 2 other ancestors, 1 with a Class I recall11996: 2 other ancestors1998: 2 other ancestors1995: 1 other ancestor1994: 2 other ancestors1991: 1 other ancestor2004: 1 other ancestor2002: 1 other ancestor2000: 2 other ancestors1993: 1 other ancestor1990: 1 other ancestorK901713 · Narkomed 4 Anesthesia System · Draeger Medical, Inc. · 1990K980553 · Narkomed 6000 Model Number NM6000 · Draeger Medical, Inc. · 1998K993826 · Narkomed 6000 · Draeger Medical, Inc. · 2000×4: 4 generations cleared close together, K033498 (2004), K041622 (2004), K042086 (2004), K042419 (2004) 19901995200020052010
  1. Root · 7 generations backK901713 Narkomed 4 Anesthesia SystemDraeger Medical, Inc. · 1990
  2. 6 generations backK980553 Narkomed 6000 Model Number NM6000Draeger Medical, Inc. · 1998
  3. 5 generations backK993826 Narkomed 6000Draeger Medical, Inc. · 2000
  4. 4 generations backK033498 Narkomed 6400, Model NM6400Draeger Medical, Inc. · 2004 · Recalled
  5. 3 generations backK041622 Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia…Draeger Medical, Inc. · 2004
  6. 2 generations backK042086 Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004
  7. 1 generation backK042419 Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004 · Recalled (Class I)
  8. This deviceK072884 Fabius MriDraeger Medical AG & Co. KG · 2008

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K901713Narkomed 4 Anesthesia SystemDraeger Medical, Inc.1990—
−6K980553Narkomed 6000 Model Number NM6000Draeger Medical, Inc.1998—
−5K993826Narkomed 6000Draeger Medical, Inc.2000—
−4K033498Narkomed 6400, Model NM6400Draeger Medical, Inc.20041
−3K041622Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
−2K042086Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
−1K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004Class I
0 (this device)K072884Fabius MriDraeger Medical AG & Co. KG2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199110
199310
199420
199510
199620
199820
200020
200210
200321
200410
200510
FIG. 1 — Lineage of K072884, 1990–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042419Fabius GS/Fabius Tiro Anesthesia System
Summary page 1
…sia (73 BSZ) Regulation Number: 21 CRF 868.5160 Class: II Predicate Device Identification: Fabius GS K042419 GE Datex Ohmeda Aestiva /5 MRI Anesthesia Sys…
Draeger Medical, Inc.2004Class I
*K050055GE Datex-Ohmeda Aestiva/5 Mri Anesthesia System
Summary page 1
…fication: Fabius GS K042419 GE Datex Ohmeda Aestiva /5 MRI Anesthesia System K050055 Section 3 - Page I of 3 510(k) Summary Fab…
Datex-Ohmeda, Inc.20052

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 16, 2021Class IIZ-0501-2022Process control26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate…
May 24, 2018Class IIZ-2252-2018Process change controlThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes…
Mar 1, 2018Class IZ-1481-2018Process controlThe automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
Jun 22, 2015Class IIZ-2271-2015Labeling False and Misleadingthe Fabius MRI or parts of the system were attracted by the magnetic field of the MRI. If the Fabius MRI is positioned too close to the MRI, its parts can be loosened as a result of the magnetic force of the MRI.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004 · Direct predicate · ◆ on the line of descent in FIG. 14 Class I recalls
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 2 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 18 years and passes through 2 companies; a new generation was cleared every 1.6 years on median.
  • 7 of the 24 ancestors have a recall of any class; 2 have a Class I recall.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Fabius MRI is indicated as a continuous flow anaesthesia system. The Fabius MRI may be used for spontaneous, manually assisted or automatic ventilation, delivery of gases and anaesthetic vapor, and monitoring of oxygen concentration, breathing pressure and respiratory volume.

The Fabius MRI is indicated for use in MRI scanner rooms with 1.5 and 3.0 tesla magnets at a location of 40 m tesla (400 gauss) or less.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072884: Fabius Mri.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072884

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.