510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Datex-Ohmeda 7900 Ventilator Enhancements

K023366510(k) TraditionalCleared Oct 7, 2003364-day reviewClass II21 CFR 868.5895Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 7, 2003 as substantially equivalent to K960964 (Ohmeda Medical, 1996) and 3 other predicates after a 364-day review. Its family tree reaches back 1 generation to 1994. 44 of the 139 devices in its family, counting itself, have a Class I recall. 134 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
Oct 7, 2003
Review
364 days
Class
Class II
Product code
CBK
Recalls
3 (1 Class I)
Predicates
4

Its longest line of descent runs back 1 generation to K930017 (1994). 4 ancestors in all. 134 later devices descend from it, 7 citing it directly. 44 of the 139 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 2 other ancestors1996: 1 other ancestor2004: 5 descendants cleared, 1 with a Class I recall12005: 7 descendants cleared2017: 7 descendants cleared, 4 with a Class I recall42007: 3 descendants cleared, 2 with a Class I recall22008: 8 descendants cleared, 2 with a Class I recall22010: 7 descendants cleared, 1 with a Class I recall12009: 8 descendants cleared, 2 with a Class I recall22006: 4 descendants cleared2026: 3 descendants cleared2012: 15 descendants cleared, 6 with a Class I recall62014: 5 descendants cleared, 1 with a Class I recall12011: 13 descendants cleared, 7 with a Class I recall72021: 5 descendants cleared, 2 with a Class I recall22015: 7 descendants cleared, 1 with a Class I recall12016: 7 descendants cleared, 3 with a Class I recall32019: 6 descendants cleared, 3 with a Class I recall32013: 10 descendants cleared, 4 with a Class I recall42024: 5 descendants cleared2018: 3 descendants cleared, 1 with a Class I recall12025: 2 descendants cleared2023: 1 descendant cleared, 1 with a Class I recall12020: 3 descendants cleared, 2 with a Class I recall2K930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 19941990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 7 descendants, 2 with a Class I recall22 generations forward: 8 descendants, 1 with a Class I recall13 generations forward: 14 descendants, 4 with a Class I recall44 generations forward: 27 descendants, 7 with a Class I recall75 generations forward: 28 descendants, 12 with a Class I recall126 generations forward: 19 descendants, 8 with a Class I recall87 generations forward: 15 descendants, 3 with a Class I recall38 generations forward: 10 descendants, 6 with a Class I recall69 generations forward: 5 descendants10 generations forward: 1 descendant K930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 19940+5+10
  1. Root · 1 generation backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. This deviceK023366 Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003
  3. Descendant · cites it directly · 107 of its ownK041775 GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2004
  4. Descendant · 2 generations on · 57 of its ownK051767 LTV 1000 VentilatorPulmonetic Systems, Inc. · 2005
  5. Descendant · 2 generations on · 49 of its ownK051895 GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2005
  6. Descendant · 3 generations on · 48 of its ownK060862 GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · Recalled
  7. Descendant · 3 generations on · 47 of its ownK062710 GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · Recalled

Showing 5 of 134 descendants. See all descendants

Descendants cleared per year

2004: 52005: 72006: 42007: 32008: 82009: 82010: 72011: 132012: 152013: 102014: 52015: 72016: 72017: 72018: 32019: 62020: 32021: 52023: 12024: 52025: 22026: 3
2004high 15 (2012) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 134 · See all
  • K041775GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2004 · cites it directly107descendants
  • K051767LTV 1000 VentilatorPulmonetic Systems, Inc. · 2005 · 2 generations on57descendants
  • K051895GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2005 · 2 generations on49descendants
  • K060862GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · 3 generations on48descendants
  • K062710GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · 3 generations on47descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K930017PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp.1994—
0 (this device)K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc.2003—
+1K041775GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.2004—
+2K051767LTV 1000 VentilatorPulmonetic Systems, Inc.2005—
+2K051895GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.2005—
+3K060862GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.20061
+3K062710GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.20062

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199820

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200451
200570
200640
200732
200882
200982
201071
2011137
2012156
2013104
201451
201571
201673
201774
201831
201963
202032
202152
202311
202450
202520
202630
FIG. 1 — Lineage of K023366, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

44 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K960964Ohmeda 7900 Anesthesia Ventilator
Summary page 1
…evices: 1. Ohmeda 7900 Anesthesia Ventilator - Class II - 21CFR868.5895, which has been the subject of a cleared 510(k) with FDA log number K960964 2. Ohmeda Excel 3000 Anesthesia Gas System - Class II - 21CFR868.5895, which has been the subject of…
Ohmeda Medical1996—
*K973896Ohmeda Excel 3000 Anesthesia System
Summary page 1
…64 2. Ohmeda Excel 3000 Anesthesia Gas System - Class II - 21CFR868.5895, which has been the subject of a cleared 510(k) with FDA log number K973896 3. Datex-Engstrom AS/3 Anesthesia Delivery Unit (ADU) - Class II - 21CFR868.5160, which has been the…
Ohmeda Medical19981
*K973985AS/3 Anaesthesia Delivery Unit (as/3 Adu)
Summary page 1
…3. Datex-Engstrom AS/3 Anesthesia Delivery Unit (ADU) - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with J?DA log number K973985 4. Puritan Bennett 7200 ICU Ventilator - Class IT - 21CFR868.5895, which has been the subject of a c…
Datex-Engstrom, Inc.19988
*K930017PB 7200 Series Ventilatory System W/Keyboard
Summary page 1
…973985 4. Puritan Bennett 7200 ICU Ventilator - Class IT - 21CFR868.5895, which has been the subject of a cleared 510(k) with FDA log number K930017 This version of the Datex-Ohmeda 7900 Ventilator is used in Datex-Ohmeda Aestivd5 Anesthesia Systems.…
Puritan Bennett Corp.1994—

Also cleared against K960964: 1 other device →

What descends from it? (134)

See all 134

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041775GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.2004107
K040743Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc.200438
K050626GE Datex-Ohmeda Aespire 7900 Anesthesia SystemDatex-Ohmeda, Inc.20054
K042086Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.20042
K042419Fabius GS/Fabius Tiro Anesthesia System Class IDraeger Medical, Inc.20041
K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire… Class IDatex-Ohmeda, Inc.20170
K051222AV-S Anaesthesia VentilatorPenlon Limited20050

Has anything in its family been recalled? (44 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 19, 2019Class IIZ-0115-2020Software in the Use EnvironmentCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that…
Dec 15, 2015Class IIZ-0678-2016Nonconforming Material/ComponentThe Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive…
Apr 10, 2012Class IZ-1787-2012Process controlGE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 14 Class I recalls
  • K072884Fabius MriDraeger Medical AG & Co. KG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 15 Class I recalls
  • K070899Puritan Bennett Legendair XL2Nellcor Puritan Bennett · 2007 · 3 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K122445Aespire 7900, Aespire ViewDatex-Ohmeda · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 4 generations on · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 4 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K102471AS3700 Anesthsia Delivery SystemMindray DS USA, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 43 shown.

What does this lineage say?

  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K041775 (Datex-Ohmeda, Inc., 2004) has gone furthest, with 107 descendants of its own.
  • 1 other device also cites its predicate K960964.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

This version of the Datex-Ohmeda 7900 Ventilator is used in Datex-Ohmeda AestivdS Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient…

Read the full statement (199 words)

This version of the Datex-Ohmeda 7900 Ventilator is used in Datex-Ohmeda AestivdS Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. User interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered tot he patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilatory modes for the device, include Volume Mode, Pressure Control Mode, Synchronous Jntermittent Mandatory Ventilation (optional), Pressure Support with Apnea Backup Ventilation. (optional).

This device is to be used only by trained and qualified medical professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023366: Datex-Ohmeda 7900 Ventilator Enhancements.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K023366

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026