510(k) · Cleared
Recalled · 2012Hospira Gemstar Infusion Pump System, Hospira Gemstar Connect Software
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Hospira, Inc. · FRN Pump, Infusion
Cleared on Nov 17, 2004 as substantially equivalent to K023062 (Abbott Laboratories, Inc., 2002) and 1 other predicate after a 19-day review. Its family tree reaches back 1 generation to 2000. It has been recalled 2 times.
- Decided
- Nov 17, 2004
- Review
- 19 days
- Class
- Class II
- Product code
- FRN
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 1 generation to K000821 (2000). 2 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK000821 Abbott Gemstar I.V. Infusion Pump - 7 Therapy Configuration, 6…Abbott Laboratories · 2000
- This deviceK042980 Hospira Gemstar Infusion Pump System, Hospira Gemstar Connect SoftwareHospira, Inc. · 2004
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K000821 | Abbott Gemstar I.V. Infusion Pump - 7 Therapy Configuration, 6 Therapy… | Abbott Laboratories | 2000 | — |
| 0 (this device) | K042980 | Hospira Gemstar Infusion Pump System, Hospira Gemstar Connect Software | Hospira, Inc. | 2004 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRN device
- #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K023062 | Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain ManagementSummary page 2…on: Infusion pumps cleared for commercial distribution and determined to be appropriate for use as predicates are sunmmnarized in the following table. [K023062'~ ~ Abbott GemSta? Infusion PumpSytm03022 IK000821 [Gemt ~.V..rV. Infusion Pump… | 2002 |
| * | K000821 | Abbott Gemstar I.V. Infusion Pump - 7 Therapy Configuration, 6 Therapy Configuration, 1…Summary page 2…ned to be appropriate for use as predicates are sunmmnarized in the following table. [K023062'~ ~ Abbott GemSta? Infusion PumpSytm03022 IK000821 [Gemt ~.V..rV. Infusion Pump [05/24/200 11. Statement of Substantial Equivalence: The H… | 2000 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 28, 2012 | Not classified | Z-1463-2012 | Nonconforming Material/Component | Hospira has received customer reports of bolus delivery failures when using the GemStar bolus cord. Possible failures may be a result of cracked or damaged bolus cords which may result in an open or short circuit of the Bolus Cord. |
| Aug 1, 2011 | Not classified | Z-3107-2011 | Device Design | Hospira has received reports of sparking, smoking, charring and electrical shock when using the GemStar Docking Station due to fluid ingress into the docking station. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- 1 of the 2 ancestors has a recall of any class.
- 3 other devices also cite its predicate K023062.
How does its category compare?
All FRN devicesFRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FRN
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRN
172in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRN
7in 2025
Class I (37 since 2007) Class II/III not classified
Adverse event reports per year · FRN
52,792in 2025
MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 102 | 15 | 1,086 |
| 2008 | 8 | 89 | 24 | 8,261 |
| 2009 | 9 | 66 | 102 | 1,952 |
| 2010 | 3 | 93 | 37 | 6,133 |
| 2011 | 3 | 103 | 69 | 41,367 |
| 2012 | 4 | 217 | 42 (2) | 15,482 |
| 2013 | 6 | 312 | 22 (3) | 40,082 |
| 2014 | 6 | 172 | 4 (3) | 94,180 |
| 2015 | 3 | 232 | 5 | 26,742 |
| 2016 | 3 | 301 | 7 (1) | 16,560 |
| 2017 | 3 | 305 | 13 | 12,060 |
| 2018 | 3 | 267 | 1 | 16,591 |
| 2019 | 2 | — | 18 (1) | 15,850 |
| 2020 | 1 | — | 16 (5) | 94,186 |
| 2021 | 2 | — | 3 (2) | 199,342 |
| 2022 | 6 | 268 | 2 | 888,020 |
| 2023 | 7 | 265 | 16 (10) | 165,838 |
| 2024 | 5 | 103 | 9 (4) | 57,730 |
| 2025 | 8 | 172 | 7 (4) | 52,792 |
| 2026 | 2 | — | 11 (2) | — |
What is it indicated for?
The Hospira GemStar® Infusion Pump System with Hospira GemStar Connect" Software is intended for use in intravenous arterial, subcutaneous, short term epidural infusion and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products.
The indications for use include hospital, ambulatory, and home care environments. The pump must be used with sterile, dedicated, GemStar® administration sets.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K042980 · printed Nov 4, 2026
Cite this page
510(k) Report. “K042980: Hospira Gemstar Infusion Pump System, Hospira Gemstar Connect Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K042980
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026