Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model 13086; Pain Management, Model 13088

K060806510(k) SpecialCleared Apr 21, 200628-day reviewClass II21 CFR 880.5725Recalls 25 (7 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · FRN Pump, Infusion

Cleared on Apr 21, 2006 as substantially equivalent to K051079 (Hospicare, Inc., 2005) and 1 other predicate after a 28-day review. Its family tree reaches back 2 generations to 2002. 7 of the 15 devices in its family, counting itself, have a Class I recall. 12 later devices descend from it. It has been recalled 25 times, including 7 Class I recalls.

Decided
Apr 21, 2006
Review
28 days
Class
Class II
Product code
FRN
Recalls
25 (7 Class I)
Predicates
3

Its longest line of descent runs back 2 generations to K023062 (2002). 12 later devices descend from it, 2 citing it directly. 7 of the 15 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2011: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2025: 2 descendants clearedK023062 · Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain Management · Abbott Laboratories, Inc. · 2002 · recalledK051079 · Hospira Gemstar Infusion Pump System with Hospira Gemstar Spring Assist Mechanism Lockbox-7 Therapy · Hospicare, Inc. · 2005 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants, 2 with a Class I recall22 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 2 descendants, 1 with a Class I recall14 generations forward: 4 descendants, 2 with a Class I recall25 generations forward: 3 descendants K051079 · Hospira Gemstar Infusion Pump System with Hospira Gemstar Spring Assist Mechanism Lockbox-7 Therapy · Hospicare, Inc. · 2005 · recalledK023062 · Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain Management · Abbott Laboratories, Inc. · 2002 · recalled0+5
  1. Root · 2 generations backK023062 Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain ManagementAbbott Laboratories, Inc. · 2002 · Recalled
  2. 1 generation backK051079 Hospira Gemstar Infusion Pump System with Hospira Gemstar Spring…Hospicare, Inc. · 2005 · Recalled
  3. This deviceK060806 Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6…Hospira, Inc. · 2006
  4. Descendant · cites it directly · 10 of its ownK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  5. Descendant · 2 generations on · 9 of its ownK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  6. Descendant · 3 generations on · 5 of its ownK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  7. Descendant · 4 generations on · 2 of its ownK223607 Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · Recalled (Class I)
  8. Descendant · 3 generations on · 2 of its ownK153074 Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding…Covidien · 2016 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12015: 12016: 12017: 12019: 12021: 12022: 12023: 12025: 22026: 1
2008high 2 (2025) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · cites it directly10descendants
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for…Hospira, Inc. · 2015 · 2 generations on9descendants
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 3 generations on5descendants
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 4 generations on2descendants
  • K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets, Kangaroo…Covidien · 2016 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K023062Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain ManagementAbbott Laboratories, Inc.20021
−1K051079Hospira Gemstar Infusion Pump System with Hospira Gemstar Spring Assist…Hospicare, Inc.20051
0 (this device)K060806Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model…Hospira, Inc.2006—
+1K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
+2K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
+3K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
+4K223607Plum Duo™ Infusion SystemIcu Medical, Inc.2023Class I
+3K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets…Covidien20165

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200811
201111
201511
201610
201711
201911
202110
202210
202311
202520
202610
FIG. 1 — Lineage of K060806, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K051079Hospira Gemstar Infusion Pump System with Hospira Gemstar Spring Assist Mechanism…
Summary page 2
…nformation: Devices cleared for commercial distribution and determined to be appropriate for use as predicates are summarized in the following table. KO051079 Hospira GemStar Infusion Pumlp System with Hospira 05/19/2005 GemStarSpring AssistMchanism Lo…
Hospicare, Inc.20051
Kept off the family treeK041739FDA’s K041739 is Edwards E360 Surgical Ablation Device
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Edwards Lifesciences, LLC2004—
*K023062Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain Management
Summary page 2
…ssistMchanism Lockbox KHospira GemStarOR Infusion Pump System with Hospira 11/17/2004 GemStar ConnectrM Soflware K023062 Abbott GemStar® Infusion Pump System 09/30/2002 11. Comparison to Legally Marketed Device(s)…
Abbott Laboratories, Inc.20021

Also cleared against K051079: 1 other device →

What descends from it? (12)

See all 12

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe Tubing Class IHospira, Inc.201110
K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite Software Class IHospira, Inc.20080

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2014Class IZ-1512-2014Device DesignThere are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List…
Feb 3, 2014Class IZ-0095-2015Process controlThe GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
Feb 3, 2014Class IZ-0096-2015Process controlThe GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
Jan 31, 2014Class IIZ-2031-2014Component design/selectionThe connection between the beeper subassembly and the pump may fail. The GemStar infusion pump will identify this failure during the "self-test" while powering up which will result in a Beeper Error ("code 10/001/000"). This Beeper Error…
Mar 21, 2013Class IIZ-1116-2013Nonconforming Material/ComponentDuring infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
Mar 21, 2013Class IIZ-1117-2013Nonconforming Material/ComponentDuring infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
Mar 19, 2013Class IIZ-0421-2015OtherSome patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via…
Mar 19, 2013Class IIZ-0422-2015OtherSome patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via…
Mar 18, 2013Class IZ-1159-2013Device DesignBattery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
Mar 18, 2013Class IZ-1160-2013Device DesignBattery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
Mar 18, 2013Class IZ-1169-2013Device DesignPump shutting off during use without warning.
Mar 15, 2013Class IZ-0349-2014Device DesignThe proximal and distal pressure sensor calibration can drift resulting in the pump failing the Proximal or Distal Occlusion Operational Test, as described in the GemStar Technical Service Manual, or reporting one of the following errors…
Mar 28, 2012Not classifiedZ-1463-2012Nonconforming Material/ComponentHospira has received customer reports of bolus delivery failures when using the GemStar bolus cord. Possible failures may be a result of cracked or damaged bolus cords which may result in an open or short circuit of the Bolus Cord.
Aug 12, 2010Not classifiedZ-2633-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Aug 12, 2010Not classifiedZ-2632-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication..
Aug 12, 2010Not classifiedZ-2629-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Aug 12, 2010Not classifiedZ-2628-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Aug 12, 2010Not classifiedZ-2630-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Aug 12, 2010Not classifiedZ-2627-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Aug 12, 2010Not classifiedZ-2631-2010Nonconforming Material/ComponentUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.
Nov 24, 2009Not classifiedZ-1337-2010Component design/selectionReports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Nov 24, 2009Not classifiedZ-1338-2010Component design/selectionReports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Nov 24, 2009Not classifiedZ-1339-2010Component design/selectionReports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Aug 11, 2009Not classifiedZ-0014-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Aug 11, 2009Not classifiedZ-0015-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 14 Class I recalls
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K103224 (Hospira, Inc., 2011) has gone furthest, with 10 descendants of its own.
  • 1 other device also cites its predicate K051079.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Hospira GemStar® Infusion Pump System is intended for use in intravenous

arterial, subcutaneous, short term epidural infusion and parenteral administration

of general I.V. fluids, medications, nutritional fluids, and blood/blood products.

The indications for use include hospital, ambulatory, and home care

environments. The pump must be used with sterile, dedicated, GemStar®

administration sets.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060806: Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model 13086; Pain Management, Model 13088.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K060806

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026