510(k) Report

510(k) · Cleared

Recalled · Class I · 2013

Hospira Phoenix Infusion System with Hospira Mednet Software

K041550510(k) TraditionalCleared Jun 30, 200421-day reviewClass II21 CFR 880.5725Recalls 36 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · FRN Pump, Infusion

Cleared on Jun 30, 2004 as substantially equivalent to K031185 (Abbott Laboratories, 2003) and 1 other predicate after a 21-day review. Its family tree reaches back 2 generations to 2001. 7 of the 18 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 36 times, including 2 Class I recalls.

Decided
Jun 30, 2004
Review
21 days
Class
Class II
Product code
FRN
Recalls
36 (2 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K011442 (2001). 4 ancestors in all. 13 later devices descend from it, 2 citing it directly. 7 of the 18 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2012: 2 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2015: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2025: 2 descendants clearedK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recallK031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 2 descendants, 1 with a Class I recall14 generations forward: 4 descendants, 2 with a Class I recall25 generations forward: 3 descendants K031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalledK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recall0+5
  1. Root · 2 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 1 generation backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. This deviceK041550 Hospira Phoenix Infusion System with Hospira Mednet SoftwareAbbott Laboratories · 2004
  4. Descendant · cites it directly · 10 of its ownK103344 Infusion SetsHospira, Inc. · 2010
  5. Descendant · 2 generations on · 9 of its ownK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  6. Descendant · 3 generations on · 5 of its ownK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  7. Descendant · 4 generations on · 2 of its ownK223607 Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · Recalled (Class I)
  8. Descendant · 3 generations on · 2 of its ownK153074 Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding…Covidien · 2016 · Recalled

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2010: 12012: 22015: 12016: 12017: 12019: 12021: 12022: 12023: 12025: 22026: 1
2010high 2 (2012) · low 0 (2011)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K103344Infusion SetsHospira, Inc. · 2010 · cites it directly10descendants
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for…Hospira, Inc. · 2015 · 2 generations on9descendants
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 3 generations on5descendants
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 4 generations on2descendants
  • K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets, Kangaroo…Covidien · 2016 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−1K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
0 (this device)K041550Hospira Phoenix Infusion System with Hospira Mednet SoftwareAbbott Laboratories2004—
+1K103344Infusion SetsHospira, Inc.2010—
+2K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
+3K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
+4K223607Plum Duo™ Infusion SystemIcu Medical, Inc.2023Class I
+3K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets…Covidien20165

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201221
201511
201610
201711
201911
202110
202210
202311
202520
202610
FIG. 1 — Lineage of K041550, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module
Summary page 1
…evice: Software 5. Predicate Devices: Abbott Plum A+® Infusion Pump with HPLJRS with the plug-and-play Module (K031 185) Abbott GemstarTm l.V. Infusion Pump system (K(023062) 6. Proposed Device Description:…
Abbott Laboratories20034
*K023062Gemstar Infusion Pump System, 7 Therapy, 6 Therapy & Pain Management
Summary page 1
…with HPLJRS with the plug-and-play Module (K031 185) Abbott GemstarTm l.V. Infusion Pump system (K(023062) 6. Proposed Device Description: The Abbott Phoenix Infusion System with Abbott MedNetwM Software is comprised of…
Abbott Laboratories, Inc.20021

Also cleared against K031185: 1 other device →

What descends from it? (13)

See all 13

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103344Infusion SetsHospira, Inc.201010
K110901Symbiq Infusion and Adminstration Set Class IHospira, Inc.20121

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 4, 2013Class IIZ-2056-2016Software designFrequent Proximal Occlusion Alarms (POAs) occurring on Symbiq devices with list numbers 16026 and 16027. When there are false alarms, the pump will continue to alarm even though there is no occlusion present. False POAs could result in a…
Feb 4, 2013Class IIZ-2055-2016Software designFrequent Proximal Occlusion Alarms (POAs) occurring on Symbiq devices with list numbers 16026 and 16027. When there are false alarms, the pump will continue to alarm even though there is no occlusion present. False POAs could result in a…
Aug 29, 2012Class IZ-0069-2013Device DesignThe Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a…
Aug 29, 2012Class IZ-0070-2013Device DesignThe Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a…
Mar 16, 2011Class IIZ-0362-2017Device DesignMay go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute…
Mar 16, 2011Class IIZ-0361-2017Device DesignMay go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute…
Apr 9, 2010Not classifiedZ-1895-2010Device DesignHospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation.
Apr 9, 2010Not classifiedZ-1894-2010Device DesignHospira has received numerous customer complaints of the Symbiq infusion pump's failure to detect air-in-line conditions during operation.
Mar 26, 2010Not classifiedZ-2381-2010Device DesignIf the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
Feb 22, 2010Not classifiedZ-0127-2011Device DesignHospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
Feb 22, 2010Not classifiedZ-0128-2011Device DesignHospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.
Feb 22, 2010Not classifiedZ-2391-2010Device DesignThe clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
Feb 22, 2010Not classifiedZ-2390-2010Device DesignThe clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
Jan 8, 2010Not classifiedZ-0808-2010Process controlThere is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.
Nov 24, 2009Not classifiedZ-1053-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1045-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1050-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1052-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1055-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1056-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1058-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1061-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1062-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1063-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1046-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1047-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1048-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1049-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1051-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1054-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1057-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1059-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Nov 24, 2009Not classifiedZ-1060-2010Process controlCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Mar 31, 2009Not classifiedZ-1816-2009Software designPotential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.
Mar 31, 2009Not classifiedZ-1815-2009Software designPotential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.
Dec 3, 2008Not classifiedZ-0944-2009Software designHypersensitivity to normal motor friction/drag may cause malfunction errors and alarms that are designed to stop pump operations.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K110901Symbiq Infusion and Adminstration SetHospira, Inc. · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls

What does this lineage say?

  • The line of descent spans 3 years and passes through 1 company; a new generation was cleared every 1.9 years on median.
  • 3 of the 4 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K103344 (Hospira, Inc., 2010) has gone furthest, with 10 descendants of its own.
  • 1 other device also cites its predicate K031185.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Hospira Phoenix Infusion System with Hospira MedNet Software is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural or irrigation routes of administration.

It is intended primarily for use in the hospital setting and can be used in other acute and non-acute care areas, such as, but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient/Surgical Centers, Long Term Care, Urgent Care, Transport and Physician Offices.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041550: Hospira Phoenix Infusion System with Hospira Mednet Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041550

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026