510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-Fcrec Module

K043276510(k) TraditionalCleared Dec 23, 200427-day reviewClass II21 CFR 870.1025Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Dec 23, 2004 as substantially equivalent to K022740 (Datex-Ohmeda, 2002) and 3 other predicates after a 27-day review. Its family tree reaches back 2 generations to 2000. 2 of the 18 devices in its family, counting itself, have a Class I recall. 11 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Dec 23, 2004
Review
27 days
Class
Class II
Product code
MHX
Recalls
1 (1 Class I)
Predicates
4

Its longest line of descent runs back 2 generations to K002158 (2000). 6 ancestors in all, the oldest K933285 (1994). 11 later devices descend from it, 1 citing it directly. 2 of the 18 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1994: 2 other ancestors2000: 1 other ancestor2012: 1 descendant cleared2013: 2 descendants cleared2014: 1 descendant cleared, 1 with a Class I recall12015: 2 descendants cleared2020: 1 descendant cleared2017: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant clearedK002158 · S/5 Compact Critical Care Monitor with S-00C03 or S-00C04 Software · Datex-Ohmeda, Inc. · 2000K022740 · Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU02 & L-CICU02A Software · Datex-Ohmeda · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant, 1 with a Class I recall14 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant7 generations forward: 2 descendants K022740 · Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU02 & L-CICU02A Software · Datex-Ohmeda · 2002 · recalledK002158 · S/5 Compact Critical Care Monitor with S-00C03 or S-00C04 Software · Datex-Ohmeda, Inc. · 20000+5
  1. Root · 2 generations backK002158 S/5 Compact Critical Care Monitor with S-00C03 or S-00C04 SoftwareDatex-Ohmeda, Inc. · 2000
  2. 1 generation backK022740 Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU02 &…Datex-Ohmeda · 2002 · Recalled
  3. This deviceK043276 Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and…Datex-Ohmeda · 2004
  4. Descendant · cites it directly · 10 of its ownK120598 Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · Recalled
  5. Descendant · 2 generations on · 8 of its ownK130584 Monitor B40Ge Medical Systems China Co., Ltd. · 2013 · Recalled
  6. Descendant · 3 generations on · 7 of its ownK133576 Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · Recalled (Class I)
  7. Descendant · 4 generations on · 5 of its ownK151063 Monitor B40Ge Medical Systems China Co., Ltd. · 2015 · Recalled
  8. Descendant · 5 generations on · 4 of its ownK171580 Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd. · 2017 · Recalled

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2012: 12013: 22014: 12015: 22017: 12020: 12022: 12024: 12025: 1
2012high 2 (2013) · low 0 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 11 · See all
  • K120598Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · cites it directly10descendants
  • K130584Monitor B40Ge Medical Systems China Co., Ltd. · 2013 · 2 generations on8descendants
  • K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · 3 generations on7descendants
  • K151063Monitor B40Ge Medical Systems China Co., Ltd. · 2015 · 4 generations on5descendants
  • K171580Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd. · 2017 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K002158S/5 Compact Critical Care Monitor with S-00C03 or S-00C04 SoftwareDatex-Ohmeda, Inc.2000—
−1K022740Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU02 & L-CICU02A…Datex-Ohmeda20022
0 (this device)K043276Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-Fcrec ModuleDatex-Ohmeda2004—
+1K120598Procare Monitor B40Ge Medical Systems China Co., Ltd.20122
+2K130584Monitor B40Ge Medical Systems China Co., Ltd.20132
+3K133576Monitor B40Ge Medical Systems China Co., Ltd.2014Class I
+4K151063Monitor B40Ge Medical Systems China Co., Ltd.20151
+5K171580Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd.20171

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
200010
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201320
201411
201520
201710
202010
202210
202410
202510
FIG. 1 — Lineage of K043276, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022740Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU02 & L-CICU02A Software
Summary page 2
…nd L- FICU04A software is substantially equivalent to the predicate Datex-Ohmeda S/5 TM Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740). The N-F(C)(REC) Extension module used with the FM Monitor allows measuring C02 (module versions N-FC and N-FCREC) and pr…
Datex-Ohmeda20022
*K023454Datex-Ohmeda S/5 Single-Width Airway Module M-Minic and Accessories
Summary page 2
…recorder chart (module versions N-FREC and N-FCREC). The C02 measurement of the N-FC(REC) is based on the C02 measurement of the predicate M-miniC module (K023454). The recorder option is based on the predicate M-REC module, originally cleared in the Datex AS/3 AM Monitor (K933…
Datex-Ohmeda2003—
*K933285Datex as/3 TM Anaesthesia Monitor
Summary page 2
…ent of the predicate M-miniC module (K023454). The recorder option is based on the predicate M-REC module, originally cleared in the Datex AS/3 AM Monitor (K933285) and the Datex AS/3 Compact Monitor (K933156) submissions. DEVICE DESCRIPTION as required by 807.92(a)(4) The S/5TM FM…
Datex Division Instrumentarium Corp.1994—
*K933156Datex as/3 Patient Monitor
Summary page 2
…. The recorder option is based on the predicate M-REC module, originally cleared in the Datex AS/3 AM Monitor (K933285) and the Datex AS/3 Compact Monitor (K933156) submissions. DEVICE DESCRIPTION as required by 807.92(a)(4) The S/5TM FM Monitor with L-FICU04(A) software is a patie…
Datex Division Instrumentarium Corp.1994—

Also cleared against K022740: 1 other device →

What descends from it? (11)

See all 11

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120598Procare Monitor B40Ge Medical Systems China Co., Ltd.201210

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 11, 2014Class IZ-2068-2014Device DesignFailure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 4 years and passes through 2 companies; a new generation was cleared every 2.2 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K120598 (Ge Medical Systems China Co., Ltd., 2012) has gone furthest, with 10 descendants of its own.
  • 1 other device also cites its predicate K022740.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The S/5 FM with L-FICU04 or L-FICU04A software is indicated for monitoring

of hemodynamics (including arrhythmia and ST-segment analysis) and

respiratory status of all hospital patients.

Extension module N-FCREC (option N-FCREC or N-FC) is indicated for

monitoring C02 and respiration rate of all hospital patients. C02

measurements are indicated for patients weighing over 5kg (11 lbs).

The S/5 FM Monitor and N-F(C)(REC) Extension Module are indicated for use

by qualified medical personnel only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043276: Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-Fcrec Module.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K043276

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026