510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Agfa Heartlab Cardiovascular

K050228510(k) SpecialCleared Feb 18, 200517-day reviewClass II21 CFR 892.2050Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Heartlab, Inc. · LLZ System, Image Processing, Radiological

Cleared on Feb 18, 2005 as substantially equivalent to K022292 (Agfa Corp., 2002) and 1 other predicate after a 17-day review. Its family tree reaches back 4 generations to 1991. 9 later devices descend from it. It has been recalled 22 times.

Decided
Feb 18, 2005
Review
17 days
Class
Class II
Product code
LLZ
Recalls
22
Predicates
2

Its longest line of descent runs back 4 generations to K913475 (1991). 11 ancestors in all. 9 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1998: 1 other ancestor1997: 1 other ancestor1996: 2 other ancestors1994: 1 other ancestor1991: 1 other ancestor2012: 2 descendants cleared2005: 1 descendant cleared2011: 1 descendant cleared2019: 1 descendant cleared2018: 1 descendant cleared2024: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant clearedK913475 · Magnetic Resonance Diagnostic Device · GE Medical Systems · 1991K960265 · Advantage Windows with Functool Option · GE Medical Systems · 1996 · recalledK971965 · Philips Easyvision · Philips Medical Systems, Inc. · 1998K022292 · Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for Cardiology Clinical Review Satation,Impax Ortho · Agfa Corp. · 2002 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant K022292 · Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for Cardiology Clinical Review Satation,Impax Ortho · Agfa Corp. · 2002 · recalledK971965 · Philips Easyvision · Philips Medical Systems, Inc. · 1998K960265 · Advantage Windows with Functool Option · GE Medical Systems · 1996 · recalledK913475 · Magnetic Resonance Diagnostic Device · GE Medical Systems · 19910+5
  1. Root · 4 generations backK913475 Magnetic Resonance Diagnostic DeviceGE Medical Systems · 1991
  2. 3 generations backK960265 Advantage Windows with Functool OptionGE Medical Systems · 1996 · Recalled
  3. 2 generations backK971965 Philips EasyvisionPhilips Medical Systems, Inc. · 1998
  4. 1 generation backK022292 Impax Diagnostic Display Station,Impax Clinical Review Station,Impax…Agfa Corp. · 2002 · Recalled
  5. This deviceK050228 Agfa Heartlab CardiovascularHeartlab, Inc. · 2005
  6. Descendant · cites it directly · 7 of its ownK053133 Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc. · 2005 · Recalled
  7. Descendant · 2 generations on · 6 of its ownK111319 Ibis Explorer and Markup SoftwareMedqia · 2011
  8. Descendant · 3 generations on · 5 of its ownK122205 EXINIExini Diagnostics AB · 2012
  9. Descendant · 4 generations on · 2 of its ownK191262 aBSIExini Diagnostics AB · 2019
  10. Descendant · 5 generations on · 1 of its ownK211655 aPROMISEExini Diagnostics AB · 2021

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2005: 12011: 12012: 22018: 12019: 12021: 12022: 12024: 1
2005high 2 (2012) · low 0 (2006)2026

Most influential descendants

Showing 5 of 9 · See all
  • K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc. · 2005 · cites it directly7descendants
  • K111319Ibis Explorer and Markup SoftwareMedqia · 2011 · 2 generations on6descendants
  • K122205EXINIExini Diagnostics AB · 2012 · 3 generations on5descendants
  • K191262aBSIExini Diagnostics AB · 2019 · 4 generations on2descendants
  • K211655aPROMISEExini Diagnostics AB · 2021 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K913475Magnetic Resonance Diagnostic DeviceGE Medical Systems1991—
−3K960265Advantage Windows with Functool OptionGE Medical Systems19961
−2K971965Philips EasyvisionPhilips Medical Systems, Inc.1998—
−1K022292Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for…Agfa Corp.20029
0 (this device)K050228Agfa Heartlab CardiovascularHeartlab, Inc.2005—
+1K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc.20052
+2K111319Ibis Explorer and Markup SoftwareMedqia2011—
+3K122205EXINIExini Diagnostics AB2012—
+4K191262aBSIExini Diagnostics AB2019—
+5K211655aPROMISEExini Diagnostics AB2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
199620
199710
199810
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
201110
201220
201810
201910
202110
202210
202410
FIG. 1 — Lineage of K050228, 1991–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022292Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for Cardiology…
Summary page 1
…mparable and substantially equivalent to the following predicate devices currently in commercial distribution: Model Manufacturer Impax (k022292) Agfa Corp. Inturis Suite (k9942 10) Philips Medical Systems O…
Agfa Corp.20029
*K994210Inturis Suite
Summary page 1
…edicate devices currently in commercial distribution: Model Manufacturer Impax (k022292) Agfa Corp. Inturis Suite (k9942 10) Philips Medical Systems ONE CROSSWIND ROAD · WESTERLY, RI - 02891…
Philips Medical Systems, Inc.2000—

Also cleared against K022292: 11 other devices →

What descends from it? (9)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc.20057
K120305Thinkingnet, Thinkingpacs/Ris - Proprietary Name Fro Thinkingnet, Thinkingarchive (or…Thinking Systems Corp.20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 22, 2016Class IIZ-2652-2017Software designA customer experienced when using IMPAX CV Reporting software, specifically, when building a NIV report, the NIV Cardio report was showing incorrect findings for Aneurysm.
Sep 1, 2015Class IIZ-0283-2016Software designCustomers have experienced IMPAX CV software, specifically, Demographics Manager (DM), not producing correct pdf reports after the reports, at the time of physician signature, were correct.
Apr 30, 2014Class IIZ-1662-2014Software designFailure to effectively install and test a hotfix to correct the known software defect within the IMPAX CV 7.8 SU3-OCR Service.
Mar 15, 2013Class IIZ-1070-2013Software designBaseline pulmonary capillary wedge (PCW) pressure values from IMPAX CV Reporting Cardiac Catheterization module were not correctly transferring into the hospital's electronic health record (EHR) system.
Mar 15, 2013Class IIZ-1069-2013Software designSoftware design error in IMPAX CV 7.8 SU3
Aug 9, 2012Class IIZ-2252-2012Process designLoss of patient data can occur under certain circumstances due to misconfiguration of DICOMstore configured with Agfa's Medical Image Storage Devices - Media Purge Daemon (MPD) and Cardiovascular Purge Service (CPS).
Jul 27, 2012Class IIZ-2184-2012Software change controlContent entered into the "Conclusions" free text box on the Report Writer screen was not represented on the final printed report.
Jul 10, 2012Class IIZ-2069-2012Error in labelingWhen users selected "Left stenosis" in the "Graft Duplex Conclusion" section in IMPAX CV Reporting Non-Invasive(NIV) module, "Right stenosis" appeared in the sentence display within the printed representation of the report.
Jun 6, 2011Not classifiedZ-0211-2012Software designImproper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable.
Jun 6, 2011Not classifiedZ-0212-2012PendingIncorrect entries in mapping files could lead to inaccurate measurement display in adult and pediatric echo reports when using the IMPAX CV Echo Measurement Import via Optical Character Recognition (OCR).
Jun 6, 2011Not classifiedZ-0272-2012Software designPotential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it.
Mar 21, 2011Not classifiedZ-2552-2011OtherCustomer error with the angiography (XA: X-Ray Angiography) measurement calibration functionality within the Cardiovascular Review Station resulted in improper calibration.
Feb 10, 2011Not classifiedZ-1676-2011Device DesignA discrepancy in the validation testing which resulted in the product not performing as intended.
Jan 27, 2011Not classifiedZ-1434-2011OtherEchocardiography volume measurements performed within CRS were not reflected correctly in associated Results Management (RM) reports.
Jun 11, 2010Not classifiedZ-2112-2010OtherAgfa Service technician had incorrectly modified a report template at one site resulting in erroneous anatomic segment locators.
Apr 15, 2009Not classifiedZ-1435-2011OtherUser could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.
Dec 24, 2008Not classifiedZ-0945-2009Software designPolling procedure may fail when multiple merge/split is carried out.
Feb 22, 2008Not classifiedZ-1349-2008Software designReporting Differences: The Outbound Report software produces an Outbound Report (text version) which does not contain all the clinical content elements present in the PDF version of the report, also generated on the Nuclear Cardiology…
Jul 27, 2007Not classifiedZ-1242-2008Software designIncorrect Results: Erroneous echocardiographic measurement values due to mathematical formula being mis-configured.
Jun 21, 2007Not classifiedZ-1241-2008Software designMIsidentification: The electrocardiogram (ECG) data of one patient is misidentified as the data of another patient.
May 14, 2007Not classifiedZ-1243-2008Software change controlJava version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
May 14, 2007Not classifiedZ-1244-2008Software change controlJava version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 4.6 years on median.
  • 2 of the 11 ancestors have a recall of any class.
  • Of its descendants, K053133 (Siemens Medical Solutions USA, Inc., 2005) has gone furthest, with 7 descendants of its own.
  • 11 other devices also cite its predicate K022292.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Ascentia TM (formally EncompassTM) is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. AscentiaTM is personal software comprised of modular software programs that run on standard "off-the-shelf" computers and servers running the Windows 2000/2003/XP operating system. AscentiaTM is image data storage and display software that accepts DICOM (Digital Imaging and Communications in…

Read the full statement (167 words)

Ascentia TM (formally EncompassTM) is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. AscentiaTM is personal software comprised of modular software programs that run on standard "off-the-shelf" computers and servers running the Windows 2000/2003/XP operating system. AscentiaTM is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) data from laboratories, which support DICOM standard imaging transfer. The system provides the capability to; consolidate images generated by equipment from multiple OEM vendors, view images, enter clinical findings while viewing the associated images, perform digital subtraction, create graphical representation of coronary arteries, perform quantitative measurements on both cath and ultrasound images, perform quantitative analysis on cath images, generate and review patient reports with additional measurement and report writer capabilities and provides accessible digital image archive. AscentiaTM is a scalable network system designed to service customers ranging in size from small departments (with 2 or 3 users) to large hospital networks (with tens of users).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050228: Agfa Heartlab Cardiovascular.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050228

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026