Impax Volume Viewing 3.0

K133135510(k) TraditionalCleared Mar 7, 2014154-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Agfa Healthcare N.V. · LLZ System, Image Processing, Radiological

Cleared on Mar 7, 2014 as substantially equivalent to K111638 (Agfa Healthcare N.V., 2011) and 4 other predicates after a 154-day review. Its family tree reaches back 6 generations to 1991. 3 later devices descend from it.

Decided
Mar 7, 2014
Review
154 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
5

Its longest line of descent runs back 6 generations to K913770 (1991). 20 ancestors in all, the oldest K913475 (1991). 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor2008: 1 other ancestor2007: 1 other ancestor2002: 2 other ancestors2005: 1 other ancestor1998: 3 other ancestors1996: 1 other ancestor1994: 2 other ancestors1992: 1 other ancestor1991: 1 other ancestor2015: 1 descendant cleared2025: 1 descendant cleared2016: 1 descendant clearedReference device: K043441 · Mirage, Release 5.5 · Segami Corporation · 2005K913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK043415 · Centricity Pacs System · Ge Medical Systems Information Technologies · 2005 · recalledK053458 · WEB1000 · Agfa Corp. · 2005 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 3 other ancestors3 generations back: 3 other ancestors4 generations back: 3 other ancestors6 generations back: 1 other ancestor 1 generation forward: 3 descendants K053458 · WEB1000 · Agfa Corp. · 2005 · recalledK043415 · Centricity Pacs System · Ge Medical Systems Information Technologies · 2005 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991−50
  1. Root · 6 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 5 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 4 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 3 generations backK033400 Seno AdvantageGe Medical Systems W-400 · 2003 · Recalled
  5. 2 generations backK043415 Centricity Pacs SystemGe Medical Systems Information Technologies · 2005 · Recalled
  6. 1 generation backK053458 WEB1000Agfa Corp. · 2005 · Recalled
  7. This deviceK133135 Impax Volume Viewing 3.0Agfa Healthcare N.V. · 2014
  8. Descendant · cites it directly · 1 of its ownK142316 IMPAX AgilityAgfa Healthcare N.V. · 2015
  9. Descendant · cites it directly · 0 of its ownK252199 AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025
  10. Descendant · cites it directly · 0 of its ownK161061 IMPAX Volume Viewing 4.0Agfa Healthcare N.V. · 2016

Most influential descendants

Showing 3 of 3 · See all
  • K142316IMPAX AgilityAgfa Healthcare N.V. · 2015 · cites it directly1descendant
  • K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025 · cites it directly0descendants
  • K161061IMPAX Volume Viewing 4.0Agfa Healthcare N.V. · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−5K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−4K020483Advantage Workstation 4.1GE Medical Systems20022
−3K033400Seno AdvantageGe Medical Systems W-40020036
−2K043415Centricity Pacs SystemGe Medical Systems Information Technologies200523
−1K053458WEB1000Agfa Corp.20052
0 (this device)K133135Impax Volume Viewing 3.0Agfa Healthcare N.V.2014—
+1K142316IMPAX AgilityAgfa Healthcare N.V.2015—
+1K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V.2025—
+1K161061IMPAX Volume Viewing 4.0Agfa Healthcare N.V.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199210
199420
199610
199830
200220
200510
200710
200810
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
202510
FIG. 1 — Lineage of K133135, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111638Impax Volume Viewing
Summary page 2
…000 NEW DEVICE PREDICATES PREDICATE K0222921K1116381K053458 K070831 ___________________ K08001 3 3D Volume rendering…
Agfa Healthcare N.V.2011—
*K080013Registration and Fusion
Summary page 1
…ng, which is a picture archiving and communications system. It is substantially equivalent to Agfa's IMPAX Volume Viewing 2.0 (KI 11638), Agfa's Registration and Fusion (K080013), Agfa's Web 1000 (K053458) and Voxar's 3D Enterprise with ColonMetrix and PET/CT Perfusion (K07083 1). B. DEVICE DESCRIPTI…
Agfa Healthcare Corporation2008—
*K053458WEB1000
Summary page 1
…iving and communications system. It is substantially equivalent to Agfa's IMPAX Volume Viewing 2.0 (KI 11638), Agfa's Registration and Fusion (K080013), Agfa's Web 1000 (K053458) and Voxar's 3D Enterprise with ColonMetrix and PET/CT Perfusion (K07083 1). B. DEVICE DESCRIPTION The new device is simila…
Agfa Corp.20052
*K070831Voxar 3D Enterprise with Colonmetrix and Pet/CT Perfusion
Summary page 1
…IMPAX Volume Viewing 2.0 (KI 11638), Agfa's Registration and Fusion (K080013), Agfa's Web 1000 (K053458) and Voxar's 3D Enterprise with ColonMetrix and PET/CT Perfusion (K07083 1). B. DEVICE DESCRIPTION The new device is similar to the predicate devices. All are PACS system accessories that allow the…
Barcoview Mis Edinburgh2007—
*K022292Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for Cardiology…
Summary page 1
…TIAL EQUIVALENCE SUMMARY Agfa's IMPAX Volume Viewing has an Indications for Use statement largely similar to the statements for the predicate devices (KI111638, K053458, K022292, K080013 and K07083 1). Intended uses are the same. As software accessories to PACS systems, the predicate and new devices have t…
Agfa Corp.20029
†K043441Mirage, Release 5.5Segami Corporation2005—

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142316IMPAX AgilityAgfa Healthcare N.V.20151
K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V.20250
K161061IMPAX Volume Viewing 4.0Agfa Healthcare N.V.20160

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 4.3 years on median.
  • 6 of the 20 ancestors have a recall of any class.
  • Of its descendants, K142316 (Agfa Healthcare N.V., 2015) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

IMPAX Volume Viewing software is a visualization package for PACS workstations. It is intended to support radiographer, medical imaging technician, radiologist and referring physician in the reading, analysis and diagnosis of DICOM compliant volumetric medical datasets. The software is intended as a general purpose digital medical image processing tool, with optional functionality to facilitate visualization and measurement of vessel features. Other…

Read the full statement (158 words)

IMPAX Volume Viewing software is a visualization package for PACS workstations. It is intended to support radiographer, medical imaging technician, radiologist and referring physician in the reading, analysis and diagnosis of DICOM compliant volumetric medical datasets. The software is intended as a general purpose digital medical image processing tool, with optional functionality to facilitate visualization and measurement of vessel features.

Other optional functionality is intended for the registration of anatomical (CT) on functional volumetric image data (MR) to facilitate the comparison of various lesions. Volume and distance measurements are intended for evaluation and quantification of tumor measurements, and other analysis and evaluation of both hard and soft tissues. The software also supports interactive segmentation of a region of interest (ROI).

Web-browser access is available for review purposes. Images accessed through a web-browser (via a mobile device or by other means) should not be used to create a diagnosis, treatment plan, or other decision that may affect patient care.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133135: Impax Volume Viewing 3.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133135

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026