510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Impax Next Generation

K111945510(k) TraditionalCleared Aug 5, 201128-day reviewClass II21 CFR 892.2050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Agfa Healthcare N.V. · LLZ System, Image Processing, Radiological

Cleared on Aug 5, 2011 as substantially equivalent to K022292 (Agfa Corp., 2002) and 2 other predicates after a 28-day review. Its family tree reaches back 6 generations to 1991. 2 later devices descend from it. It has been recalled 2 times.

Decided
Aug 5, 2011
Review
28 days
Class
Class II
Product code
LLZ
Recalls
2
Predicates
3

Its longest line of descent runs back 6 generations to K913770 (1991). 31 ancestors in all, the oldest K904519 (1991). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2002: 4 other ancestors2005: 2 other ancestors1998: 4 other ancestors2001: 3 other ancestors2000: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1994: 3 other ancestors1993: 2 other ancestors1992: 1 other ancestor1991: 2 other ancestors2015: 1 descendant cleared2025: 1 descendant clearedK913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK040555 · Impax Client Embrace · Agfa Corp. · 2004 · recalledK081976 · Impax MA3000 Diagnostic Pacs Workstation · Agfa Healthcare Corporation · 2008 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 5 other ancestors3 generations back: 9 other ancestors4 generations back: 7 other ancestors5 generations back: 1 other ancestor6 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K081976 · Impax MA3000 Diagnostic Pacs Workstation · Agfa Healthcare Corporation · 2008 · recalledK040555 · Impax Client Embrace · Agfa Corp. · 2004 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991−50
  1. Root · 6 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 5 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 4 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 3 generations backK033400 Seno AdvantageGe Medical Systems W-400 · 2003 · Recalled
  5. 2 generations backK040555 Impax Client EmbraceAgfa Corp. · 2004 · Recalled
  6. 1 generation backK081976 Impax MA3000 Diagnostic Pacs WorkstationAgfa Healthcare Corporation · 2008 · Recalled
  7. This deviceK111945 Impax Next GenerationAgfa Healthcare N.V. · 2011
  8. Descendant · cites it directly · 1 of its ownK142316 IMPAX AgilityAgfa Healthcare N.V. · 2015
  9. Descendant · 2 generations on · 0 of its ownK252199 AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025

Most influential descendants

Showing 2 of 2 · See all
  • K142316IMPAX AgilityAgfa Healthcare N.V. · 2015 · cites it directly1descendant
  • K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−5K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−4K020483Advantage Workstation 4.1GE Medical Systems20022
−3K033400Seno AdvantageGe Medical Systems W-40020036
−2K040555Impax Client EmbraceAgfa Corp.20043
−1K081976Impax MA3000 Diagnostic Pacs WorkstationAgfa Healthcare Corporation20081
0 (this device)K111945Impax Next GenerationAgfa Healthcare N.V.2011—
+1K142316IMPAX AgilityAgfa Healthcare N.V.2015—
+2K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199120
199210
199320
199430
199610
199710
199840
200010
200130
200240
200520
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
202510
FIG. 1 — Lineage of K111945, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022292Impax Diagnostic Display Station,Impax Clinical Review Station,Impax for Cardiology…
Summary page 1
…ext Generation, which is a picture archiving and communications system. It is substantially equivalent to Agfa's IMPAX workstations with MPR, Digital Subtraction and 3D (K022292), and for orthopedic use to its Agfa Orthopedic Software for IMPAX Workstations (K07 1972), and for mammography to Agfa's IMPAX M…
Agfa Corp.20029
*K071972Orthopedic Software for Impax Workstations
Summary page 1
…ly equivalent to Agfa's IMPAX workstations with MPR, Digital Subtraction and 3D (K022292), and for orthopedic use to its Agfa Orthopedic Software for IMPAX Workstations (K07 1972), and for mammography to Agfa's IMPAX MA3000 (KOS81976). B. DEVICE DESCRIPTION The new device is largely similar to the pre…
Agfa Healthcare Corp.2007—
*K081976Impax MA3000 Diagnostic Pacs Workstation
Summary page 2
…L EQUIVALENCE SUMMARY Agfa's IMPAX Next Generation has an Indications for Use statement largely similar to the statements for the predicate devices (K022292, K071972 and K08 1976). The statements have been combined and simplified. Intended uses are the same. The devices have the same technological character…
Agfa Healthcare Corporation20081

Also cleared against K022292: 11 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142316IMPAX AgilityAgfa Healthcare N.V.20151

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 3, 2018Class IIZ-0101-2019Software designAdding an additional comparison study to the Clinical Side Bar in the Enterprise Imaging Diagnostics Desktop, can result in a change of the sort order of the studies in the Clinical Side Bar. The thumbnail images of the studies are no…
Jan 8, 2014Class IIZ-0919-2014Unknown/Undetermined by firmPatient name displayed (printed) on the Patient Report was the wrong patient name.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 6 companies; a new generation was cleared every 4.2 years on median.
  • 10 of the 31 ancestors have a recall of any class.
  • Of its descendants, K142316 (Agfa Healthcare N.V., 2015) has gone furthest, with 1 descendant of its own.
  • 11 other devices also cite its predicate K022292.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Agfa's IMPAX Next Generation isa Picture Archiving and Communications System (PACS). It provides an interface for the acquisition, display, digital processing, annotation, review, printing, storage and distribution of multimodality medical images, reports and demographic information for diagnostic purposes within the system and across computer networks. IMPAX Next Generation isintended to be used by trained healthcare professionals including, but not…

Read the full statement (193 words)

Agfa's IMPAX Next Generation isa Picture Archiving and Communications System (PACS). It provides an interface for the acquisition, display, digital processing, annotation, review, printing, storage and distribution of multimodality medical images, reports and demographic information for diagnostic purposes within the system and across computer networks. IMPAX Next Generation isintended to be used by trained healthcare professionals including, but not limited to physicians, radiologists, orthopaedlic surgeons, cardiologists, mammographers, technologists, and clinicians for diagnosis and treatment planning using DICOM compliant medical images and other healthcare data.

MPR, MIP and 3D rendering functionality allows the user to view image data from perspectives different from that in which it was acquired. Other digital image processing functionality such as multi-scale window leveling and image registration can be used to enhance image viewing.

As acomprehensive imaging suite, IMPAX Next Generation integrates with servers, archives, Radiology Information Systems (RIS), Hospital Information Systems (HIS), reporting and 3rd party applications for customer specific workflows.

Lossy compressed mammography images and digitized film images should not be used for primary image interpretation. Uncompressed or non-lossy compressed "for presentation" images may be used for diagnosis or screening on monitors that are FDA-cleared for mammographic use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111945: Impax Next Generation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K111945

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.