Inspiration Ventilator System

K051550510(k) SpecialCleared Jul 12, 200529-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

eVent Medical, Ltd. · CBK Ventilator, Continuous, Facility Use

Cleared on Jul 12, 2005 as substantially equivalent to K030341 (eVent Medical, Ltd., 2004) and 1 other predicate after a 29-day review. Its family tree reaches back 6 generations to 1991. 3 of the 26 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once.

Decided
Jul 12, 2005
Review
29 days
Class
Class II
Product code
CBK
Recalls
1
Predicates
2

Its longest line of descent runs back 6 generations to K902859 (1991). 22 ancestors in all, the oldest K864394 (1986). 3 later devices descend from it, 1 citing it directly. 3 of the 26 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 3 other ancestors2002: 1 other ancestor2000: 2 other ancestors, 1 with a Class I recall11999: 1 other ancestor1998: 1 other ancestor1996: 1 other ancestor1997: 2 other ancestors1991: 1 other ancestor1992: 1 other ancestor1986: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2008: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant cleared, 1 with a Class I recall1K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984379 · Nellcor Puritan Bennett, 760 Ventilator System · Puritan Bennett Corp. · 1998 · recalledK021112 · Inspiration Ventilator System · eVent Medical, Ltd. · 2002K030341 · Modification to Inspiration Ventilator System · eVent Medical, Ltd. · 20041990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 7 other ancestors, 1 with a Class I recall14 generations back: 3 other ancestors5 generations back: 2 other ancestors6 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants, 1 with a Class I recall1 K030341 · Modification to Inspiration Ventilator System · eVent Medical, Ltd. · 2004K021112 · Inspiration Ventilator System · eVent Medical, Ltd. · 2002K984379 · Nellcor Puritan Bennett, 760 Ventilator System · Puritan Bennett Corp. · 1998 · recalledK970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK961687 · EVITA 4 · Drager, Inc. · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−50
  1. Root · 6 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 5 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 4 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 3 generations backK984379 Nellcor Puritan Bennett, 760 Ventilator SystemPuritan Bennett Corp. · 1998 · Recalled
  5. 2 generations backK021112 Inspiration Ventilator SystemeVent Medical, Ltd. · 2002
  6. 1 generation backK030341 Modification to Inspiration Ventilator SystemeVent Medical, Ltd. · 2004
  7. This deviceK051550 Inspiration Ventilator SystemeVent Medical, Ltd. · 2005
  8. Descendant · cites it directly · 2 of its ownK072590 Inspiration; Ventilator System, LS Ventilator System, Infant…Event Medical, Inc. · 2008
  9. Descendant · 2 generations on · 0 of its ownK130178 Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · Recalled (Class I)
  10. Descendant · 2 generations on · 0 of its ownK113743 Evolution 3E VentilatorEvent Medical, Inc. · 2012

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System and…Event Medical, Inc. · 2008 · cites it directly2descendants
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 2 generations on0descendants
  • K113743Evolution 3E VentilatorEvent Medical, Inc. · 2012 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K902859Servo Ventilator 300Siemens Life Support Systems19913
−5K961687EVITA 4Drager, Inc.1996—
−4K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−3K984379Nellcor Puritan Bennett, 760 Ventilator SystemPuritan Bennett Corp.19984
−2K021112Inspiration Ventilator SystemeVent Medical, Ltd.2002—
−1K030341Modification to Inspiration Ventilator SystemeVent Medical, Ltd.2004—
0 (this device)K051550Inspiration Ventilator SystemeVent Medical, Ltd.2005—
+1K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System…Event Medical, Inc.2008—
+2K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd.2013Class I
+2K113743Evolution 3E VentilatorEvent Medical, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199110
199210
199410
199510
199610
199720
199810
199910
200021
200130
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201210
201311
FIG. 1 — Lineage of K051550, 1991–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030341Modification to Inspiration Ventilator System
Summary page 1
…ollowing two predicate medical devices: Predicate Device 510(k) Number Classification eVent Medical Ltd.. K030341 Class LI, Inspiration T M Ventilator Continuous Ventilator…
eVent Medical, Ltd.2004—
*K010925Servoi Ventilator System, Model 64 87 800 E407E
Summary page 1
…Continuous Ventilator per 21 CFR 868.5895 Siemens - Elma AB. K010925 Class II, Servoi Ventilator Continuous Ventilator…
Siemens Elema AB20011

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System and…Event Medical, Inc.20082

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 27, 2010Not classifiedZ-1596-2011Component design/selectionComplaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 7 of the 22 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K072590 (Event Medical, Inc., 2008) has gone furthest, with 2 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The InspirationTM Ventilator System with Smart Positive Airway Pressure (SPAP) and Volume Targeted Ventilation (VTV) options is indicated for use with a wide range of patients from infant through adult, requiring respiratory support for a wide range of clinical conditions in hospital, hospital-type facilities and intra-hospital transport.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051550: Inspiration Ventilator System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051550

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026