Evolution 3E Ventilator

K113743510(k) TraditionalCleared Dec 18, 2012364-day reviewClass II21 CFR 868.5895

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Event Medical, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Dec 18, 2012 as substantially equivalent to K072590 (Event Medical, Inc., 2008) after a 364-day review. Its family tree reaches back 8 generations to 1991. 2 of the 32 devices in its family, counting itself, have a Class I recall.

Decided
Dec 18, 2012
Review
364 days
Class
Class II
Product code
CBK
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K902859 (1991). 31 ancestors in all, the oldest K833982 (1984). No later device cites it as a predicate yet. 2 of the 32 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2006: 1 other ancestor2004: 1 other ancestor2001: 4 other ancestors2002: 1 other ancestor1999: 2 other ancestors1992: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestor1984: 2 other ancestors2000: 2 other ancestors, 1 with a Class I recall11998: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor1997: 2 other ancestors1986: 1 other ancestorK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984379 · Nellcor Puritan Bennett, 760 Ventilator System · Puritan Bennett Corp. · 1998 · recalledK021112 · Inspiration Ventilator System · eVent Medical, Ltd. · 2002K030341 · Modification to Inspiration Ventilator System · eVent Medical, Ltd. · 2004K051550 · Inspiration Ventilator System · eVent Medical, Ltd. · 2005 · recalledK072590 · Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System and Infant LS Ventilator System · Event Medical, Inc. · 20081980198519901995200020052010
  1. Root · 8 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 7 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 6 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 5 generations backK984379 Nellcor Puritan Bennett, 760 Ventilator SystemPuritan Bennett Corp. · 1998 · Recalled
  5. 4 generations backK021112 Inspiration Ventilator SystemeVent Medical, Ltd. · 2002
  6. 3 generations backK030341 Modification to Inspiration Ventilator SystemeVent Medical, Ltd. · 2004
  7. 2 generations backK051550 Inspiration Ventilator SystemeVent Medical, Ltd. · 2005 · Recalled
  8. 1 generation backK072590 Inspiration; Ventilator System, LS Ventilator System, Infant…Event Medical, Inc. · 2008
  9. This deviceK113743 Evolution 3E VentilatorEvent Medical, Inc. · 2012

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K902859Servo Ventilator 300Siemens Life Support Systems19913
−7K961687EVITA 4Drager, Inc.1996—
−6K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−5K984379Nellcor Puritan Bennett, 760 Ventilator SystemPuritan Bennett Corp.19984
−4K021112Inspiration Ventilator SystemeVent Medical, Ltd.2002—
−3K030341Modification to Inspiration Ventilator SystemeVent Medical, Ltd.2004—
−2K051550Inspiration Ventilator SystemeVent Medical, Ltd.20051
−1K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System…Event Medical, Inc.2008—
0 (this device)K113743Evolution 3E VentilatorEvent Medical, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198420
198610
198910
199110
199210
199410
199510
199610
199720
199810
199920
200021
200140
200210
200410
200610
FIG. 1 — Lineage of K113743, 1991–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System and…
Summary page 1
…evices are: Table I1- Predicate Devices Manufacturer/Product 510(k) Classification eVent Medical K072590 Class 11 lnspirationTM Ventilator System Continuous Ventilator…
Event Medical, Inc.2008—

Also cleared against K072590: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 2.5 years on median.
  • 10 of the 31 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K072590.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The eVolution 3e ventilator is intended for and designed to provide continuous and or intermittent mechanical ventilation to patients requiring ventilatory support through invasive or non-invasive interfaces and is suitable for use in the ICU, sub acute, long-term acute care, rehabilitation, and emergency room venues. This product is intended for a wide range of patients from pediatric to adult having body weights in the range of 5 kg (11 Ilb) or more…

Read the full statement (123 words)

The eVolution 3e ventilator is intended for and designed to provide continuous and or intermittent mechanical ventilation to patients requiring ventilatory support through invasive or non-invasive interfaces and is suitable for use in the ICU, sub acute, long-term acute care, rehabilitation, and emergency room venues.

This product is intended for a wide range of patients from pediatric to adult having body weights in the range of 5 kg (11 Ilb) or more who require pressure-based or volume-based continuous respiratory support, as prescribed by an attending physician.

The eVolution 3e ventilator system is a class Ilb medical device to-be used by trained and qualified healthcare professionals in hospitals or healthcare facilities and intended for sale by or on the order of a physician only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113743: Evolution 3E Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113743

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026