510(k) Report

510(k) · Cleared

Recalled · Class I · 2015

Inspiration Ventilator System 5I/7I

K130178510(k) TraditionalCleared Oct 29, 2013277-day reviewClass II21 CFR 868.5895Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

eVent Medical, Ltd. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 29, 2013 as substantially equivalent to K072590 (Event Medical, Inc., 2008) and 1 other predicate after a 277-day review. Its family tree reaches back 12 generations to 1991. 9 of the 107 devices in its family, counting itself, have a Class I recall. It has been recalled once, including a Class I recall.

Decided
Oct 29, 2013
Review
277 days
Class
Class II
Product code
CBK
Recalls
1 (1 Class I)
Predicates
2

Its longest line of descent runs back 12 generations to K902859 (1991). 106 ancestors in all, the oldest K791045 (1979). No later device cites it as a predicate yet. 9 of the 107 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2008: 4 other ancestors2010: 2 other ancestors2006: 4 other ancestors2005: 4 other ancestors1998: 12 other ancestors, 3 with a Class I recall32004: 8 other ancestors, 1 with a Class I recall12001: 6 other ancestors1995: 5 other ancestors1991: 5 other ancestors1990: 6 other ancestors1979: 1 other ancestor2002: 4 other ancestors1999: 7 other ancestors1996: 3 other ancestors1992: 2 other ancestors1989: 1 other ancestor1984: 2 other ancestors2003: 4 other ancestors, 1 with a Class I recall12000: 3 other ancestors, 1 with a Class I recall11997: 4 other ancestors1994: 4 other ancestors1986: 1 other ancestor1993: 1 other ancestor1988: 1 other ancestorK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK032451 · Bird Tbird Vela Ventilator · Bird Products Corp. · 2004 · recalled×2: 2 generations cleared close together, K052554 (2005), K053270 (2006) K061627 · Modification to I Vent 201 · Versamed Medical Systems, Inc. · 2006K063404 · Servo-I Ventilator System, Models 64, 87 and 800 · Maquet Critical Care AB · 2007K081842 · Engstrom Carestation, Engstrom Pro · Datex-Ohmeda, Inc. · 2008K103803 · Hamilton-G5 · Hamilton Medical AG · 2011 · Class I recall1980199020002010
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 7 other ancestors3 generations back: 13 other ancestors, 2 with a Class I recall24 generations back: 23 other ancestors5 generations back: 15 other ancestors, 2 with a Class I recall26 generations back: 5 other ancestors, 1 with a Class I recall17 generations back: 8 other ancestors, 1 with a Class I recall18 generations back: 9 other ancestors9 generations back: 10 other ancestors10 generations back: 2 other ancestors11 generations back: 1 other ancestor K103803 · Hamilton-G5 · Hamilton Medical AG · 2011 · Class I recallK081842 · Engstrom Carestation, Engstrom Pro · Datex-Ohmeda, Inc. · 2008K063404 · Servo-I Ventilator System, Models 64, 87 and 800 · Maquet Critical Care AB · 2007K061627 · Modification to I Vent 201 · Versamed Medical Systems, Inc. · 2006K053270 · IVENT 201 · Versamed Medical Systems, Inc. · 2006K052554 · Modification to: Ivent 201 Portable Ventilator · Versamed Medical Systems, Inc. · 2005K032451 · Bird Tbird Vela Ventilator · Bird Products Corp. · 2004 · recalledK013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−10−50
  1. Root · 12 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 8 generations between
  3. 3 generations backK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  4. 2 generations backK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008
  5. 1 generation backK103803 Hamilton-G5Hamilton Medical AG · 2011 · Recalled (Class I)
  6. This deviceK130178 Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013
Show all 12 generations
  1. Root · 12 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 11 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 10 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. 9 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
  5. 8 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  6. 7 generations backK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  7. 6 generations backK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  8. 5 generations backK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  9. 4 generations backK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006
  10. 3 generations backK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  11. 2 generations backK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008
  12. 1 generation backK103803 Hamilton-G5Hamilton Medical AG · 2011 · Recalled (Class I)
  13. This deviceK130178 Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K902859Servo Ventilator 300Siemens Life Support Systems19913
−11K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−10K970839Servo Ventilator 300ASiemens Elema AB19974
−9K992788Bear Cub 750 PSVBear Medical Systems, Inc.1999—
−8K013642AVEABird Products Corp.2002Class I
−7K032451Bird Tbird Vela VentilatorBird Products Corp.20041
−6K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
−5K053270IVENT 201Versamed Medical Systems, Inc.2006—
−4K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—
−3K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
−2K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc.2008—
−1K103803Hamilton-G5Hamilton Medical AG2011Class I
0 (this device)K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198420
198610
198810
198910
199060
199150
199220
199310
199440
199550
199630
199740
1998123
199970
200031
200160
200240
200341
200481
200540
200640
200840
201020
FIG. 1 — Lineage of K130178, 1991–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072590Inspiration; Ventilator System, LS Ventilator System, Infant Ventilator System and…
Summary page 1
…- Predicate Devices Manufacturer/Product 510(k) Classification eVent Medical K072590 Class 11 InspirationTM Ventilator System Continuous Ventilator…
Event Medical, Inc.2008—
*K103803Hamilton-G5
Summary page 1
…us Ventilator per 21 CFR 868.5895 Hamilton G-5 Ventilator K103803 Class 11 Continuous Ventilator ___…
Hamilton Medical AG2011Class I

Also cleared against K072590: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 13, 2015Class IZ-0230-2016Component design/selectionA component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K103803Hamilton-G5Hamilton Medical AG · 2011 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 5 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013642AVEABird Products Corp. · 2002 · 7 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 7 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 9 companies; a new generation was cleared every 1.8 years on median.
  • 32 of the 106 ancestors have a recall of any class; 8 have a Class I recall.
  • 1 other device also cites its predicate K072590.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Inspiration 7i/5i Ventilator System is intended for use with patients having body weights in the range of 0.3kg to 200kg and Tidal Volumes of 5ml to 2000ml. The Inspiration Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport. This product is intended for a wide range of patients from infant to adult and for a wide variety of clinical conditions. The intended patient population…

Read the full statement (148 words)

The Inspiration 7i/5i Ventilator System is intended for use with patients having body weights in the range of 0.3kg to 200kg and Tidal Volumes of 5ml to 2000ml. The Inspiration Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.

This product is intended for a wide range of patients from infant to adult and for a wide variety of clinical conditions.

The intended patient population includes infant through adult patients who require pressure-based or volume-based continuous respiratory support with tidal volumes as low as 5 ml and inspiratory pressures as low as 1cm H2O.

The device is intended for use in hospitals and hospital-type facilities, which provide respiratory care for patients requiring respiratory support.

The device is intended for sale by or on the order of a physician only. The device is intended for operation by trained and qualified personnel.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130178: Inspiration Ventilator System 5I/7I.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130178

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026