Nellcor Puritan Bennett, 840 Ventilator System (Model 840)

K970460510(k) TraditionalCleared Apr 30, 1998448-day reviewClass II21 CFR 868.5895Recalls 8 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Puritan Bennett Corp. · CBK Ventilator, Continuous, Facility Use

Cleared on Apr 30, 1998 as substantially equivalent to K902506 (Puritan Bennett Corp., 1991) and 3 other predicates after a 448-day review. Its family tree reaches back 2 generations to 1991. 49 of the 181 devices in its family, counting itself, have a Class I recall. 174 later devices descend from it. It has been recalled 8 times, including a Class I recall.

Decided
Apr 30, 1998
Review
448 days
Class
Class II
Product code
CBK
Recalls
8 (1 Class I)
Predicates
4

Its longest line of descent runs back 2 generations to K902859 (1991). 6 ancestors in all, the oldest K902506 (1991). 174 later devices descend from it, 7 citing it directly. 49 of the 181 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1992: 1 other ancestor1991: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2008: 12 descendants cleared, 2 with a Class I recall22002: 4 descendants cleared, 2 with a Class I recall21998: 2 descendants cleared, 1 with a Class I recall12006: 9 descendants cleared, 1 with a Class I recall12007: 6 descendants cleared, 1 with a Class I recall12012: 15 descendants cleared, 5 with a Class I recall52004: 6 descendants cleared, 1 with a Class I recall12003: 1 descendant cleared2005: 4 descendants cleared2011: 14 descendants cleared, 7 with a Class I recall72014: 6 descendants cleared, 1 with a Class I recall12010: 8 descendants cleared, 2 with a Class I recall22013: 10 descendants cleared, 4 with a Class I recall42009: 6 descendants cleared, 1 with a Class I recall12025: 6 descendants cleared2021: 7 descendants cleared, 3 with a Class I recall32015: 8 descendants cleared, 1 with a Class I recall12017: 11 descendants cleared, 5 with a Class I recall52016: 8 descendants cleared, 3 with a Class I recall32023: 5 descendants cleared, 1 with a Class I recall12019: 8 descendants cleared, 3 with a Class I recall32024: 5 descendants cleared2018: 3 descendants cleared, 1 with a Class I recall12026: 5 descendants cleared2020: 4 descendants cleared, 3 with a Class I recall32022: 1 descendant clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 19961990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors 1 generation forward: 7 descendants, 3 with a Class I recall32 generations forward: 13 descendants, 2 with a Class I recall23 generations forward: 22 descendants, 2 with a Class I recall24 generations forward: 34 descendants, 10 with a Class I recall105 generations forward: 34 descendants, 14 with a Class I recall146 generations forward: 28 descendants, 8 with a Class I recall87 generations forward: 17 descendants, 5 with a Class I recall58 generations forward: 8 descendants, 2 with a Class I recall29 generations forward: 5 descendants10 generations forward: 2 descendants, 1 with a Class I recall111 generations forward: 3 descendants, 1 with a Class I recall112 generations forward: 1 descendant K961687 · EVITA 4 · Drager, Inc. · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled0+5+10
  1. Root · 2 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 1 generation backK961687 EVITA 4Drager, Inc. · 1996
  3. This deviceK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998
  4. Descendant · cites it directly · 133 of its ownK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  5. Descendant · 2 generations on · 114 of its ownK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  6. Descendant · 3 generations on · 112 of its ownK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  7. Descendant · 3 generations on · 87 of its ownK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  8. Descendant · 4 generations on · 86 of its ownK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006

Showing 5 of 174 descendants. See all descendants

Descendants cleared per year

1998: 22002: 42003: 12004: 62005: 42006: 92007: 62008: 122009: 62010: 82011: 142012: 152013: 102014: 62015: 82016: 82017: 112018: 32019: 82020: 42021: 72022: 12023: 52024: 52025: 62026: 5
1998high 15 (2012) · low 0 (1999)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 174 · See all
  • K013642AVEABird Products Corp. · 2002 · cites it directly133descendants
  • K032451Bird Tbird Vela VentilatorBird Products Corp. · 2004 · 2 generations on114descendants
  • K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005 · 3 generations on112descendants
  • K053270IVENT 201Versamed Medical Systems, Inc. · 2006 · 3 generations on87descendants
  • K061627Modification to I Vent 201Versamed Medical Systems, Inc. · 2006 · 4 generations on86descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K902859Servo Ventilator 300Siemens Life Support Systems19913
−1K961687EVITA 4Drager, Inc.1996—
0 (this device)K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998—
+1K013642AVEABird Products Corp.2002Class I
+2K032451Bird Tbird Vela VentilatorBird Products Corp.20041
+3K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
+3K053270IVENT 201Versamed Medical Systems, Inc.2006—
+4K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199210
199410
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199821
200242
200310
200461
200540
200691
200761
2008122
200961
201082
2011147
2012155
2013104
201461
201581
201683
2017115
201831
201983
202043
202173
202210
202351
202450
202560
202650
FIG. 1 — Lineage of K970460, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

49 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K9025067200AE Microprocessor Ventilator W/Enhanced Hard.
Summary page 1
…medical devices: Predicate Device 510(k) Number Classification Puritan-Bennett Corp. 7200 Series Ventilator Bear Medical Systems Inc. K912619 Bear 1000 Ventilator K902506/B (7200ae) Continuous, Critical Care Ventilator ' Continuous, Critical Care Ventilator Siemens Corp. K902859 Continuous, Criti…
Puritan Bennett Corp.1991—
*K912619Bear 7 Ventilator
Summary page 1
…ce to the following predicate medical devices: Predicate Device 510(k) Number Classification Puritan-Bennett Corp. 7200 Series Ventilator Bear Medical Systems Inc. K912619 Bear 1000 Ventilator K902506/B (7200ae) Continuous, Critical Care Ventilator ' Continuous, Critical Care Ventilator Siemens…
Bear Medical Systems, Inc.1992—
*K902859Servo Ventilator 300
Summary page 1
…Bear Medical Systems Inc. K912619 Bear 1000 Ventilator K902506/B (7200ae) Continuous, Critical Care Ventilator ' Continuous, Critical Care Ventilator Siemens Corp. K902859 Continuous, Critical Servo 300 Ventilator Care Ventilator Dréager K961687 Continuous, Critical Evita 4 Ventilator Care Ventilat…
Siemens Life Support Systems19913
*K961687EVITA 4
Summary page 1
…e) Continuous, Critical Care Ventilator ' Continuous, Critical Care Ventilator Siemens Corp. K902859 Continuous, Critical Servo 300 Ventilator Care Ventilator Dréager K961687 Continuous, Critical Evita 4 Ventilator Care Ventilator 1. DEVICE DESCRIPTION: The device is a critical care v…
Drager, Inc.1996—

Also cleared against K902506: 6 other devices →

What descends from it? (174)

See all 174

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K013642AVEA Class IBird Products Corp.2002133
K984535Puritan-Bennett 840 Ventilatory System with Bilevel Option Class IPuritan Bennett Corp.199855
K062093Avea VentilatorViasys Respiratory Care, Inc.200621
K984379Nellcor Puritan Bennett, 760 Ventilator SystemPuritan Bennett Corp.19988
K023350Esprit Ventilator with Respiratory Mechanics Class IRespironics California, Inc.20022
K072926Helontix VentLinde Gas Therapeutics20080
K063650Ventilator with Respiratory Mechanics, Model 840Puritan Bennett Corp.20070

Has anything in its family been recalled? (49 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 4, 2014Class IIZ-1809-2014Component design/selectionIn the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss of display and thus…
Jun 3, 2014Class IIZ-1844-2014Nonconforming Material/ComponentCovidien is conducting a medical device field correction for specific Puritan Bennett 840 Ventilators due to customer reports of loss of graphical user interface (GUI) display information while the ventilator continues to provide breath…
Dec 16, 2013Class IZ-0611-2014Software designCovidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical…
Jan 25, 2013Class IIZ-0822-2014Labeling Change ControlCovidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators…
Jan 25, 2013Class IIZ-0821-2014Labeling Change ControlCovidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators…
Mar 21, 2008Not classifiedZ-1692-2008Process controlWiring in battery back-up power supply may short and cause thermal damage to ventilator.
Mar 12, 2007Not classifiedZ-0760-2007OtherIn rare instances, the safety valve may remain open after an occlusion is resolved. If the safety valve does not reset, the ventilator will detect a circuit disconnect condition. When this happens, the ventilator will annunciate a…
Apr 20, 2005Not classifiedZ-1085-05OtherManufacturing defect at contract manufacturer, employee error, poor training, improper inspection.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K013642AVEABird Products Corp. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 113 Class I recalls
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Cites it directly · ◆ in the 1998 descendants bar of FIG. 11 Class I recall
  • K023350Esprit Ventilator with Respiratory MechanicsRespironics California, Inc. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
  • K034040Esprit Ventilator, Model V1000Respironics, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 3 generations on · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp. · 2006 · 4 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K093261MOVESThornhill Research, Inc. · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls

Nearest 12 of 48 shown.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 5.6 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K013642 (Bird Products Corp., 2002) has gone furthest, with 133 descendants of its own.
  • 6 other devices also cite its predicate K902506.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The 840 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from infant to adult and for a wide variety of clinical conditions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970460: Nellcor Puritan Bennett, 840 Ventilator System (Model 840).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K970460

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.