510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)

K052007510(k) TraditionalCleared Dec 16, 2005144-day reviewClass II21 CFR 862.1345Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Horiba Abx · CFR Hexokinase, Glucose

Cleared on Dec 16, 2005 as substantially equivalent to K920402 (Roche Diagnostic Systems, Inc., 1992) and 4 other predicates after a 144-day review. Its family tree reaches back 3 generations to 1994. 15 later devices descend from it. It has been recalled 6 times.

Decided
Dec 16, 2005
Review
144 days
Class
Class II
Product code
CFR
Recalls
6
Predicates
5

Its longest line of descent runs back 3 generations to K933930 (1994). 12 ancestors in all, the oldest K801297 (1980). 15 later devices descend from it, 9 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1996: 1 other ancestor1992: 1 other ancestor1980: 1 other ancestor1999: 1 other ancestor1993: 1 other ancestor1991: 3 other ancestors2010: 1 descendant cleared2012: 1 descendant cleared2011: 5 descendants cleared2007: 2 descendants cleared2006: 2 descendants cleared2013: 1 descendant cleared2021: 1 descendant cleared2019: 1 descendant cleared2015: 1 descendant clearedK933930 · Puritan-Bennett Companion 550 Liquid Oxygen Portable Unit · Puritan Bennett Corp. · 1994K993220 · Puritan-Bennett Helios Portable Liquid Oxygen System · Puritan Bennett Corp. · 1999K041122 · Precision Liquid Oxygen System · Precision Medical, Inc. · 2004 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 5 other ancestors 1 generation forward: 9 descendants2 generations forward: 6 descendants K041122 · Precision Liquid Oxygen System · Precision Medical, Inc. · 2004 · recalledK993220 · Puritan-Bennett Helios Portable Liquid Oxygen System · Puritan Bennett Corp. · 1999K933930 · Puritan-Bennett Companion 550 Liquid Oxygen Portable Unit · Puritan Bennett Corp. · 19940
  1. Root · 3 generations backK933930 Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp. · 1994
  2. 2 generations backK993220 Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp. · 1999
  3. 1 generation backK041122 Precision Liquid Oxygen SystemPrecision Medical, Inc. · 2004 · Recalled
  4. This deviceK052007 Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)Horiba Abx · 2005
  5. Descendant · cites it directly · 3 of its ownK060854 Abx Pentra 400: LipoproteinsHoriba Abx · 2006 · Recalled
  6. Descendant · cites it directly · 2 of its ownK123171 Abx Pentra Calcium as CP, Abx Pentra Urine Control L/H, Abx Pentra…Horiba Abx Sas · 2013
  7. Descendant · cites it directly · 1 of its ownK110530 Abx Pentra Creatinine 120 CP; Abx Pentra Multical; Abx Pentra N…Horiba Abx Sas · 2012 · Recalled
  8. Descendant · cites it directly · 1 of its ownK110137 Abx Pentra Enzymatic Creatinine CP, Abx Pentra Multical, Abx Pentra…Horiba Abx Sas · 2011 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK193649 Yumizen C1200 Creatinine PAPHoriba Abx Sas · 2021

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2006: 22007: 22010: 12011: 52012: 12013: 12015: 12019: 12021: 1
2006high 5 (2011) · low 0 (2008)2026

Most influential descendants

Showing 5 of 15 · See all
  • K060854Abx Pentra 400: LipoproteinsHoriba Abx · 2006 · cites it directly3descendants
  • K123171Abx Pentra Calcium as CP, Abx Pentra Urine Control L/H, Abx Pentra Multical, Abx Pentra…Horiba Abx Sas · 2013 · cites it directly2descendants
  • K110530Abx Pentra Creatinine 120 CP; Abx Pentra Multical; Abx Pentra N Control; Abx Pentra P…Horiba Abx Sas · 2012 · cites it directly1descendant
  • K110137Abx Pentra Enzymatic Creatinine CP, Abx Pentra Multical, Abx Pentra N Control, Abx…Horiba Abx Sas · 2011 · cites it directly1descendant
  • K193649Yumizen C1200 Creatinine PAPHoriba Abx Sas · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K933930Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp.1994—
−2K993220Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp.1999—
−1K041122Precision Liquid Oxygen SystemPrecision Medical, Inc.20041
0 (this device)K052007Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)Horiba Abx2005—
+1K060854Abx Pentra 400: LipoproteinsHoriba Abx20062
+1K123171Abx Pentra Calcium as CP, Abx Pentra Urine Control L/H, Abx Pentra Multical…Horiba Abx Sas2013—
+1K110530Abx Pentra Creatinine 120 CP; Abx Pentra Multical; Abx Pentra N Control; Abx…Horiba Abx Sas20123
+1K110137Abx Pentra Enzymatic Creatinine CP, Abx Pentra Multical, Abx Pentra N Control…Horiba Abx Sas20112
+2K193649Yumizen C1200 Creatinine PAPHoriba Abx Sas2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
199130
199210
199310
199610
199910
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200620
200720
201010
201150
201210
201310
201510
201910
202110
FIG. 1 — Lineage of K052007, 1994–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CFR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CFR device

  1. #1 K253490 Glucose HK Gen.3; ONLINE DAT Methadone II…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K920402Roche Cobas Mira Plus
Summary page 3
…vice developed by HoribaABX. It has been demonstrated that the ABX PENTRA 400 can be considered substantially equivalent to the predicate device Roche Cobas Mira Plus (K920402). HORIBA ABX, FRANCE Section 6-3 The…
Roche Diagnostic Systems, Inc.1992—
*K963627(Modified Version) Integra Ion Selective Electrode (Ise) Module
Summary page 4
…, FRANCE Section 6-3 The optional ISE module has been similarly demonstrated as being substantially equivalent to K963627; the Roche Ion Selective Electrode. Whilst the ABX PENTRA Glucose HK CP, and ABX PENTRA Glucose PAP CP have been demonstrated sub…
Roche Diagnostic Systems, Inc.1996—
*K801297Cobas Reagent for Glucose Rapid
Summary page 4
…lent to K963627; the Roche Ion Selective Electrode. Whilst the ABX PENTRA Glucose HK CP, and ABX PENTRA Glucose PAP CP have been demonstrated substantially equivalent to K801297; the Cobas Reagent for Glucose Rapid (K801297). ABX PENTRA Multical, ABX PENTRA N Control and ABX PENTRA P Control…
Roche Analytical Instruments, Inc.1980—
*K033501Modification to Calibrator for Automatic Systems (Cfas)
Summary page 4
…apid (K801297). ABX PENTRA Multical, ABX PENTRA N Control and ABX PENTRA P Control have been respectively demonstrated as being substantially equivalent to K033501, Roche Calibrator for Automated Systems and K0411227, Roche Precinorm U and Roche Precipath U. Description: The ABX PENTRA 400…
Roche Diagnostics Corp.2003—
*K041122Precision Liquid Oxygen System
Summary page 4
…ntrol and ABX PENTRA P Control have been respectively demonstrated as being substantially equivalent to K033501, Roche Calibrator for Automated Systems and K0411227, Roche Precinorm U and Roche Precipath U. Description: The ABX PENTRA 400 is a benchtop clinical chemistry analyzer using two…
Precision Medical, Inc.20041

Also cleared against K920402: 1 other device →

What descends from it? (15)

See all 15

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060854Abx Pentra 400: LipoproteinsHoriba Abx20063
K123171Abx Pentra Calcium as CP, Abx Pentra Urine Control L/H, Abx Pentra Multical, Abx Pentra…Horiba Abx Sas20132
K110530Abx Pentra Creatinine 120 CP; Abx Pentra Multical; Abx Pentra N Control; Abx Pentra P…Horiba Abx Sas20121
K110137Abx Pentra Enzymatic Creatinine CP, Abx Pentra Multical, Abx Pentra N Control, Abx…Horiba Abx Sas20111
K103788Pentra C200, Ise Module and Abx Pentra Glucose HK CP ReagentHoriba Abx Sas20110
K102647Elitech Clinical Systems Selectra Prom/Elical 2/Elitrol I and Elitrol II/Ast/Got 4+1…Elitechgroup Vital Scientific BV20110
K100525Elitech Clinical Systems Glucose Pap SL Elitech Clinical Systems Elical 2, Elitech…Seppim S.A.S.20100
K072115Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and P Controls, and…Horiba Abx20070
K062180Abx Pentra Amylase CP; Multical; N Control; P Control; Clean-Chem CP; Clean-Chem 99 CPHoriba Abx20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2017Class IIZ-0494-2018Process controlThe ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without…
Nov 21, 2017Class IIZ-0493-2018Process controlThe ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without…
Jan 11, 2016Class IIZ-0740-2016Under Investigation by firmN-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with…
Nov 13, 2014Class IIZ-0902-2015Package design/selectionHORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead…
Nov 21, 2013Class IIZ-0904-2014Under Investigation by firmHoriba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium.
Mar 20, 2013Class IIZ-1141-2013Software designHORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 5.0 years on median.
  • 2 of the 12 ancestors have a recall of any class.
  • Of its descendants, K060854 (Horiba Abx, 2006) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K920402.

How does its category compare?

All CFR devices

CFR Hexokinase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CFR

1in 2025

2007: 12008: 52009: 32011: 32012: 42013: 12014: 12016: 22017: 12019: 12022: 12025: 12026 (partial year): 1
2007high 5 (2008) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CFR

115in 2012

2008: 182 days median2009: 101 days median2011: 316 days median2012: 115 days median
2007high 316 (2011) · low 101 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CFR

1in 2016

2008: 72009: 12012: 12014: 12016: 1
2007high 7 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CFR

16in 2025

2007: 12008: 22009: 72010: 112011: 432012: 492013: 362014: 22015: 32016: 12017: 132018: 82019: 102020: 42021: 52022: 32023: 292024: 152025: 16
2007high 49 (2012) · low 1 (2007)2025

MAUDE reports naming product code CFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008518272
2009310117
20100—011
20113316043
20124115149
20131—036
20141—12
20150—03
20162—11
20171—013
20180—08
20191—010
20200—04
20210—05
20221—03
20230—029
20240—015
20251—016
20261—0—

What is it indicated for?

The ABX PENTRA 400 is a discrete photometric benchtop chemistry analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The device is intended for use in conjunction with certain materials to measure a variety of analytes. ABX PENTRA Glucose HK CP, Glucose PAP CP reagents with associated calibrators and controls are for…

Read the full statement (245 words)

The ABX PENTRA 400 is a discrete photometric benchtop chemistry analyzer for clinical use.

The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity.

The device is intended for use in conjunction with certain materials to measure a variety of analytes.

ABX PENTRA Glucose HK CP, Glucose PAP CP reagents with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase and glucose oxidase methods by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The option of an I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Chloride, and Potassium by potentiometry using ion selective electrode with associated reference solution, calibrators and controls. Measurement of Sodium, Chloride, and Potassium are used in diagnosis and treatment diseases involving electrolyte imbalance.

The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers as specified in the enclosed annex.

The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed annex.

The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed annex.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052007: Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K052007

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026