510(k) Report

510(k) · Cleared

Recalled · 2006

Precision Liquid Oxygen System

K041122510(k) TraditionalCleared Oct 4, 2004158-day reviewClass II21 CFR 868.5655Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Precision Medical, Inc. · BYJ Unit, Liquid-Oxygen, Portable

Cleared on Oct 4, 2004 as substantially equivalent to K993220 (Puritan Bennett Corp., 1999) after a 158-day review. Its family tree reaches back 2 generations to 1994. 19 of the 68 devices in its family, counting itself, have a Class I recall. 65 later devices descend from it. It has been recalled once.

Decided
Oct 4, 2004
Review
158 days
Class
Class II
Product code
BYJ
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K933930 (1994). 65 later devices descend from it, 3 citing it directly. 19 of the 68 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2005: 1 descendant cleared2006: 3 descendants cleared2011: 11 descendants cleared, 4 with a Class I recall42010: 2 descendants cleared2012: 6 descendants cleared, 2 with a Class I recall22007: 2 descendants cleared2013: 7 descendants cleared, 3 with a Class I recall32008: 3 descendants cleared, 1 with a Class I recall12009: 1 descendant cleared2021: 4 descendants cleared, 1 with a Class I recall12019: 4 descendants cleared, 1 with a Class I recall12015: 6 descendants cleared, 1 with a Class I recall12017: 4 descendants cleared, 2 with a Class I recall22014: 1 descendant cleared2016: 2 descendants cleared, 2 with a Class I recall22020: 2 descendants cleared, 1 with a Class I recall12018: 1 descendant cleared2024: 2 descendants cleared2026: 2 descendants cleared2023: 1 descendant cleared, 1 with a Class I recall1K933930 · Puritan-Bennett Companion 550 Liquid Oxygen Portable Unit · Puritan Bennett Corp. · 1994K993220 · Puritan-Bennett Helios Portable Liquid Oxygen System · Puritan Bennett Corp. · 19991990200020102020
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 12 descendants, 1 with a Class I recall13 generations forward: 10 descendants, 1 with a Class I recall14 generations forward: 7 descendants, 6 with a Class I recall65 generations forward: 13 descendants, 4 with a Class I recall46 generations forward: 11 descendants, 5 with a Class I recall57 generations forward: 5 descendants, 1 with a Class I recall18 generations forward: 3 descendants, 1 with a Class I recall19 generations forward: 1 descendant K993220 · Puritan-Bennett Helios Portable Liquid Oxygen System · Puritan Bennett Corp. · 1999K933930 · Puritan-Bennett Companion 550 Liquid Oxygen Portable Unit · Puritan Bennett Corp. · 19940+5
  1. Root · 2 generations backK933930 Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp. · 1994
  2. 1 generation backK993220 Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp. · 1999
  3. This deviceK041122 Precision Liquid Oxygen SystemPrecision Medical, Inc. · 2004
  4. Descendant · cites it directly · 47 of its ownK062710 GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · Recalled
  5. Descendant · 2 generations on · 37 of its ownK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008
  6. Descendant · 3 generations on · 21 of its ownK093886 Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C · 2010
  7. Descendant · 2 generations on · 21 of its ownK080181 Modification to Galileo GoldHamilton Medical AG · 2008
  8. Descendant · 2 generations on · 16 of its ownK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)

Showing 5 of 65 descendants. See all descendants

Descendants cleared per year

2005: 12006: 32007: 22008: 32009: 12010: 22011: 112012: 62013: 72014: 12015: 62016: 22017: 42018: 12019: 42020: 22021: 42023: 12024: 22026: 2
2005high 11 (2011) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 65 · See all
  • K062710GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc. · 2006 · cites it directly47descendants
  • K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008 · 2 generations on37descendants
  • K093886Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C · 2010 · 3 generations on21descendants
  • K080181Modification to Galileo GoldHamilton Medical AG · 2008 · 2 generations on21descendants
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations on16descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K933930Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp.1994—
−1K993220Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp.1999—
0 (this device)K041122Precision Liquid Oxygen SystemPrecision Medical, Inc.2004—
+1K062710GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.20062
+2K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc.2008—
+3K093886Engstrom Carestation, Engstrom ProDatex-Ohmeda Inc., A Division of Instrumentarium C2010—
+2K080181Modification to Galileo GoldHamilton Medical AG2008—
+2K081521Modification to Hamilton-G5Hamilton Medical AG2008Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200630
200720
200831
200910
201020
2011114
201262
201373
201410
201561
201622
201742
201810
201941
202021
202141
202311
202420
202620
FIG. 1 — Lineage of K041122, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

19 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993220Puritan-Bennett Helios Portable Liquid Oxygen System
Summary page 1
…868.5655 Class ILDevice Predicate Device Equivalence Precision Medical, Inc. is claiming substantial equivalence to the Puritan-Bennett Helios Liquid Oxygen System K 993220. 4 51 Okliquidsystem42204 Device Descriptio…
Puritan Bennett Corp.1999—

Also cleared against K993220: 3 other devices →

What descends from it? (65)

See all 65

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K062710GE Datex-Ohmeda Engstrom CarestationDatex-Ohmeda, Inc.200647
K052007Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)Horiba Abx200515
K103324Precision Liquid Oxygen SystemPrecision Medical, Inc.20110

Has anything in its family been recalled? (19 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 4, 2006Not classifiedZ-0153-2007OtherComponent assembly; The fill chuck is not assembled properly which may allow the device to come apart over time releasing its contents.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K131774Hamilton-G5Hamilton Medical AG · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K142679Carescape R860Datex-Ohmeda, Inc. · 2015 · 5 generations on · ◆ in the 2015 descendants bar of FIG. 13 Class I recalls
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 5 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K123125GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc. · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 19 shown.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 5.6 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K062710 (Datex-Ohmeda, Inc., 2006) has gone furthest, with 47 descendants of its own.
  • 3 other devices also cite its predicate K993220.

How does its category compare?

All BYJ devices

BYJ Unit, Liquid-Oxygen, Portable: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BYJ

1in 2013

2007: 22008: 32010: 12011: 12012: 12013: 1
2007high 3 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BYJ

164in 2008

2008: 164 days median
2007high 164 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BYJ

4in 2022

2007: 12022: 4, 4 Class I
2007high 4 (2022) · low 0 (2008)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · BYJ

1in 2025

2007: 242008: 332009: 122010: 242011: 152012: 112013: 222014: 252015: 342016: 192017: 182018: 212019: 242020: 392021: 72022: 42023: 102024: 102025: 1
2007high 39 (2020) · low 1 (2025)2025

MAUDE reports naming product code BYJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—124
20083164033
20090—012
20101—024
20111—015
20121—011
20131—022
20140—025
20150—034
20160—019
20170—018
20180—021
20190—024
20200—039
20210—07
20220—4 (4)4
20230—010
20240—010
20250—01
20260—0—

What is it indicated for?

The Precision Medical, Inc. Liquid Oxygen System is intended tQ provide supplemental oxygen to oxygen patients who may have difficulty extracting oxygen from the air that they breathe. The patients would normally receive the oxygen via a nasal cannula. The system delivers 100% oxygen at 4 different flow settings. It is intended to be used as ambulatory source of oxygen both inside and out side of the patient's home. It is not intended as a life supporting device. The device has no contraindications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041122: Precision Liquid Oxygen System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041122

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026