510(k) Report

510(k) · Cleared

Recalled · Class III · 2012

Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and P Controls, and Urine Control L/H

K072115510(k) TraditionalCleared Nov 30, 2007121-day reviewClass II21 CFR 862.1225Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Horiba Abx · CGX Alkaline Picrate, Colorimetry, Creatinine

Cleared on Nov 30, 2007 as substantially equivalent to K973869 (Trace America, Inc., 1998) and 3 other predicates after a 121-day review. Its family tree reaches back 4 generations to 1994. It has been recalled once.

Decided
Nov 30, 2007
Review
121 days
Class
Class II
Product code
CGX
Recalls
1
Predicates
4

Its longest line of descent runs back 4 generations to K933930 (1994). 21 ancestors in all, the oldest K801297 (1980). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2007: 1 other ancestor1998: 1 other ancestor1994: 1 other ancestor2003: 1 other ancestor2002: 1 other ancestor1997: 2 other ancestors1996: 1 other ancestor1992: 1 other ancestor1990: 1 other ancestor1988: 1 other ancestor1980: 1 other ancestor1999: 1 other ancestor1993: 1 other ancestor1991: 3 other ancestorsK933930 · Puritan-Bennett Companion 550 Liquid Oxygen Portable Unit · Puritan Bennett Corp. · 1994K993220 · Puritan-Bennett Helios Portable Liquid Oxygen System · Puritan Bennett Corp. · 1999K041122 · Precision Liquid Oxygen System · Precision Medical, Inc. · 2004 · recalledK052007 · Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module) · Horiba Abx · 2005 · recalled198019851990199520002005
  1. Root · 4 generations backK933930 Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp. · 1994
  2. 3 generations backK993220 Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp. · 1999
  3. 2 generations backK041122 Precision Liquid Oxygen SystemPrecision Medical, Inc. · 2004 · Recalled
  4. 1 generation backK052007 Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)Horiba Abx · 2005 · Recalled
  5. This deviceK072115 Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and…Horiba Abx · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K933930Puritan-Bennett Companion 550 Liquid Oxygen Portable UnitPuritan Bennett Corp.1994—
−3K993220Puritan-Bennett Helios Portable Liquid Oxygen SystemPuritan Bennett Corp.1999—
−2K041122Precision Liquid Oxygen SystemPrecision Medical, Inc.20041
−1K052007Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)Horiba Abx20056
0 (this device)K072115Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and P…Horiba Abx2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198810
199010
199130
199210
199310
199410
199610
199720
199810
199910
200210
200310
200710
FIG. 1 — Lineage of K072115, 1994–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGX device

  1. #1 K242685 Atellica® CH Creatinine_3 (Crea3)score 12.1 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973869Trace Reagent Line for the Cobas Mira
Summary page 2
…Substantially equivalent Submission device Predicate device ABX PENTRA Creatinine 120 CP K973869 (serum/plasma samples) K941837 (urine samples) ABX PENTRA Total Protein 1…
Trace America, Inc.1998—
*K941837Roche Reagent for Creatinine
Summary page 2
…Predicate device ABX PENTRA Creatinine 120 CP K973869 (serum/plasma samples) K941837 (urine samples) ABX PENTRA Total Protein 100 CP K973869 ABX PENTRA Multical…
Roche Diagnostic Systems, Inc.1994—
*K052007Abx Pentra 400 Clinical Chemistry Analyzer (Option: I.S.E. Module)
Summary page 1
…Date Prepared: 2 6th July 2007 Device Names: The following reagents, controls & calibrators are for use in conjunction with the ABX PENTRA 400, cleared to market under K052007. REAGENTS: Trade/Proprietary Name: ABX PENTRA Creatinine 120 CP Common or Usual Name: Creatinine Device Class…
Horiba Abx20056
*K070249Abx Pentra 400, Calcium CP, Creatinine CP, Phosphorus CP, Amylase CP, Urine Control L/H
Summary page 2
…control material (assayed) Product Code: JJY; Multi-Analyte Controls, All Kinds (assayed) Trade/Proprietary Name: ABX PENTRA Urine Control L/H (K070249) Common or Usual Name: Urine control Device Class Class I Classification Name: §862.1660 : Q…
Horiba Abx20071

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 14, 2012Class IIIZ-0289-2013Nonconforming Material/ComponentThis recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 5.0 years on median.
  • 4 of the 21 ancestors have a recall of any class.

How does its category compare?

All CGX devices

CGX Alkaline Picrate, Colorimetry, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · CGX

1in 2025

2007: 42009: 12012: 12013: 12014: 22016: 12022: 12024: 22025: 12026 (partial year): 1
2007high 4 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGX

25in 2007

2007: 25 days median
2007high 25 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGX

2in 2016

2008: 12009: 22010: 22011: 12012: 22016: 22026 (partial year): 1
2007high 2 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGX

34in 2025

2007: 152008: 272009: 52010: 52011: 322012: 132013: 242014: 132015: 132016: 32017: 212018: 142019: 252020: 192021: 132022: 232023: 192024: 322025: 34
2007high 34 (2025) · low 3 (2016)2025

MAUDE reports naming product code CGX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007425015
20080—127
20091—25
20100—25
20110—132
20121—213
20131—024
20142—013
20150—013
20161—23
20170—021
20180—014
20190—025
20200—019
20210—013
20221—023
20230—019
20242—032
20251—034
20261—1—

What is it indicated for?

Creatinine and Total Protein reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure a variety of analytes. ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine…

Read the full statement (224 words)

Creatinine and Total Protein reagents, with associated calibrators and controls, are intended for use on ABX PENTRA 400 Clinical Chemistry Analyzer to measure a variety of analytes.

ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffé method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

ABX PENTRA Total Protein 100 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in-vitro determination of Total Proteins in serum and plasma by colorimetry.

Measurements obtained by this device are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders.

The ABX PENTRA Multical is a calibrator for use in the calibration of quantitative Horiba ABX methods on Horiba ABX clinical chemistry analyzers.

The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision.

The ABX PENTRA P Control is for use in quality control by monitoring accuracy and precision.

The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072115: Abx Pentra Creatinine 120 CP, Total Protein 100 CP, Multical, N and P Controls, and Urine Control L/H.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072115

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026