510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Innova CT Option for Digital Fluoroscopic Imaging Systems Innova Models 4100, 4100IQ, 3100, 3100IQ & 2100IQ

K052157510(k) TraditionalCleared Aug 25, 200517-day reviewClass II21 CFR 892.1750Recalls 39

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · JAK System, X-Ray, Tomography, Computed

Cleared on Aug 25, 2005 as substantially equivalent to K042646 (Siemens Medical Systems, Inc., 2004) after a 17-day review. Its family tree reaches back 1 generation to 2004. It has been recalled 39 times.

Decided
Aug 25, 2005
Review
17 days
Class
Class II
Product code
JAK
Recalls
39
Predicates
1

Its longest line of descent runs back 1 generation to K042646 (2004). No later device cites it as a predicate yet.

K042646 · Dynact · Siemens Medical Systems, Inc. · 20042000200220042006
K042646 · Dynact · Siemens Medical Systems, Inc. · 20040
  1. Root · 1 generation backK042646 DynactSiemens Medical Systems, Inc. · 2004
  2. This deviceK052157 Innova CT Option for Digital Fluoroscopic Imaging Systems Innova…Ge Medical Systems, LLC · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K042646DynactSiemens Medical Systems, Inc.2004—
0 (this device)K052157Innova CT Option for Digital Fluoroscopic Imaging Systems Innova Models 4100…Ge Medical Systems, LLC2005—
FIG. 1 — Lineage of K052157, 2004–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042646Dynact
Summary page 1
…n is substantially equivalent to the currently marketed DynaCT option of Slermens Vascular Angiographic system (K042646). The Innova CT is intended for use with Innova44100 (renamed Innova. 4 10 0'0 and previously…
Siemens Medical Systems, Inc.2004—

Also cleared against K042646: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 21, 2013Class IIZ-0947-2013Nonconforming Material/ComponentA locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.
Dec 20, 2011Class IIZ-1318-2013Nonconforming Material/ComponentWhile performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries…
Oct 21, 2011Class IIZ-1142-2013Labeling designAs part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff…
Aug 24, 2011Not classifiedZ-0133-2012Nonconforming Material/ComponentA potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
May 16, 2011Class IIZ-1616-2013Software designGE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
Mar 24, 2010Not classifiedZ-1661-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1659-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1660-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Dec 22, 2009Not classifiedZ-0618-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0619-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0617-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Aug 5, 2009Not classifiedZ-0410-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0409-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0408-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Apr 16, 2009Not classifiedZ-0135-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0137-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0136-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Jul 25, 2008Not classifiedZ-2438-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jul 25, 2008Not classifiedZ-2436-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jul 25, 2008Not classifiedZ-2437-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jun 17, 2008Not classifiedZ-1475-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1476-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1474-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jan 30, 2008Not classifiedZ-0473-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0475-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0474-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 14, 2008Not classifiedZ-2041-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2039-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2040-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Nov 15, 2007Not classifiedZ-0691-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Nov 15, 2007Not classifiedZ-0692-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Mar 26, 2007Not classifiedZ-0871-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0870-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0869-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-1044-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…
Mar 26, 2007Not classifiedZ-1043-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…
Aug 17, 2006Not classifiedZ-1519-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…
Aug 17, 2006Not classifiedZ-1520-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…
Aug 17, 2006Not classifiedZ-1518-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 3 other devices also cite its predicate K042646.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. This device is not…

Read the full statement (118 words)

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. This device is not intended for mammography applications.

Innova CT is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up.

Innova CT is intended for imaging bone and soft tissues as well as other internal body structures.

Innova CT is not intended for mammography applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052157: Innova CT Option for Digital Fluoroscopic Imaging Systems Innova Models 4100, 4100IQ, 3100, 3100IQ & 2100IQ.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052157

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026