K052863510(k) TraditionalCleared Jan 6, 200687-day reviewClass II21 CFR 868.5895Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hamilton Medical AG · CBK Ventilator, Continuous, Facility Use

Cleared on Jan 6, 2006 as substantially equivalent to K982910 (Hamilton Medical, Inc., 1999) and 2 other predicates after an 87-day review. Its family tree reaches back 6 generations to 1991. 3 of the 27 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 3 times.

Decided
Jan 6, 2006
Review
87 days
Class
Class II
Product code
CBK
Recalls
3
Predicates
3

Its longest line of descent runs back 6 generations to K902859 (1991). 25 ancestors in all, the oldest K902506 (1991). 1 later device descends from it, 1 citing it directly. 3 of the 27 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors, 1 with a Class I recall11999: 3 other ancestors2002: 1 other ancestor2001: 4 other ancestors1998: 1 other ancestor1997: 3 other ancestors1996: 1 other ancestor1991: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor1992: 1 other ancestor2007: 1 descendant clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK022679 · Raphael · Hamilton Medical AG · 2003 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 20041990200020102020
  1. Root · 6 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 5 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 4 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 3 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
  5. 2 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
  6. 1 generation backK040574 Galileo GoldHamilton Medical, Inc. · 2004
  7. This deviceK052863 Raphael ColorHamilton Medical AG · 2006
  8. Descendant · cites it directly · 0 of its ownK071194 Raphael Color Asv VentilatorHamilton Medical AG · 2007

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K071194Raphael Color Asv VentilatorHamilton Medical AG · 2007 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K902859Servo Ventilator 300Siemens Life Support Systems19913
−5K961687EVITA 4Drager, Inc.1996—
−4K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−3K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp.1998Class I
−2K022679RaphaelHamilton Medical AG20033
−1K040574Galileo GoldHamilton Medical, Inc.2004—
0 (this device)K052863Raphael ColorHamilton Medical AG2006—
+1K071194Raphael Color Asv VentilatorHamilton Medical AG2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199210
199410
199510
199610
199730
199810
199930
200021
200140
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
FIG. 1 — Lineage of K052863, 1991–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982910Hamilton Medical Galileoventilator
Summary page 1
…being claimed Predicate Device Manufacturer 510(k) number(s) Classification NIV mode and TRC feature GALILEO Gold HAMILTON MEDICAL K982910, K001686, Ventilator, Continuous, AG K040574 Faci…
Hamilton Medical, Inc.1999—
*K001686Galileo Ventilator
Summary page 1
…imed Predicate Device Manufacturer 510(k) number(s) Classification NIV mode and TRC feature GALILEO Gold HAMILTON MEDICAL K982910, K001686, Ventilator, Continuous, AG K040574 Facility Use…
Hamilton Medical, Inc.2000—
*K040574Galileo Gold
Summary page 1
…mode and TRC feature GALILEO Gold HAMILTON MEDICAL K982910, K001686, Ventilator, Continuous, AG K040574 Facility Use per 21 CFR 868.6895…
Hamilton Medical, Inc.2004—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071194Raphael Color Asv VentilatorHamilton Medical AG20070

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2006Not classifiedZ-1250-06OtherALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
May 24, 2006Not classifiedZ-1249-06OtherALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
May 24, 2006Not classifiedZ-1251-06OtherALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 2 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 5 companies; a new generation was cleared every 1.4 years on median.
  • 9 of the 25 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K071194 (Hamilton Medical AG, 2007) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The RAPHAEL Color ventilator is a continuous ventilator designed for ventilation of adult, pediatric, and infant patients weighing between 5 and 200 kg. The RAPHAEL Color ventilator is intended for use by properly trained personnel under direct supervision of a licensed physician. The RAPHAEL Color is intended for use in a hospital or hospital-type facility, including use at a patient bedside or for intra-facility transport, provided compressed gas is supplied.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052863: Raphael Color.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K052863

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026