510(k) Report

510(k) · Cleared

Recalled · 2009

Modification to: HI-Art System

K060912510(k) SpecialCleared Apr 19, 200616-day reviewClass II21 CFR 892.5050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tomotherapy Incorporated · IYE Accelerator, Linear, Medical

Cleared on Apr 19, 2006 as substantially equivalent to K042739 (Tomotherapy Incorporated, 2004) after a 16-day review. Its family tree reaches back 7 generations to 1992. 24 later devices descend from it. It has been recalled 4 times.

Decided
Apr 19, 2006
Review
16 days
Class
Class II
Product code
IYE
Recalls
4
Predicates
1

Its longest line of descent runs back 7 generations to K914698 (1992). 22 ancestors in all, the oldest K882729 (1988). 24 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors1998: 4 other ancestors1992: 1 other ancestor1996: 1 other ancestor1994: 3 other ancestors1988: 1 other ancestor1993: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2009: 3 descendants cleared2008: 1 descendant cleared2011: 3 descendants cleared2010: 1 descendant cleared2013: 1 descendant cleared2014: 2 descendants cleared2012: 2 descendants cleared2016: 2 descendants cleared2015: 1 descendant cleared2020: 2 descendants cleared2017: 2 descendants cleared2018: 2 descendants cleared2021: 1 descendant cleared2023: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013673 · HI-Art System · Tomo Therapy, Inc. · 2002K033347 · Modification to: HI-Art System · Tomotherapy Incorporated · 2003K042739 · HI-Art System · Tomotherapy Incorporated · 20041990200020102020
Other ancestors, by generationDescendants, by generation 4 generations back: 5 other ancestors5 generations back: 6 other ancestors6 generations back: 4 other ancestors 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 5 descendants4 generations forward: 6 descendants5 generations forward: 4 descendants6 generations forward: 2 descendants K042739 · HI-Art System · Tomotherapy Incorporated · 2004K033347 · Modification to: HI-Art System · Tomotherapy Incorporated · 2003K013673 · HI-Art System · Tomo Therapy, Inc. · 2002K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−50+5
  1. Root · 7 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 6 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 5 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 4 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 3 generations backK013673 HI-Art SystemTomo Therapy, Inc. · 2002
  6. 2 generations backK033347 Modification to: HI-Art SystemTomotherapy Incorporated · 2003
  7. 1 generation backK042739 HI-Art SystemTomotherapy Incorporated · 2004
  8. This deviceK060912 Modification to: HI-Art SystemTomotherapy Incorporated · 2006
  9. Descendant · cites it directly · 13 of its ownK082005 Modification to HI-Art Treatment SystemTomotherapy Incorporated · 2008 · Recalled
  10. Descendant · 2 generations on · 11 of its ownK112776 Tomotherapy Treatment System Fast OptimizerTomotherapy Incorporated · 2011
  11. Descendant · 3 generations on · 10 of its ownK121934 Tomotherapy Treatment SystemAccuray Incorporated · 2012 · Recalled
  12. Descendant · cites it directly · 9 of its ownK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  13. Descendant · 2 generations on · 7 of its ownK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled

Showing 5 of 24 descendants. See all descendants

Descendants cleared per year

2008: 12009: 32010: 12011: 32012: 22013: 12014: 22015: 12016: 22017: 22018: 22020: 22021: 12023: 1
2008high 3 (2009) · low 0 (2019)2026

Most influential descendants

Showing 5 of 24 · See all
  • K082005Modification to HI-Art Treatment SystemTomotherapy Incorporated · 2008 · cites it directly13descendants
  • K112776Tomotherapy Treatment System Fast OptimizerTomotherapy Incorporated · 2011 · 2 generations on11descendants
  • K121934Tomotherapy Treatment SystemAccuray Incorporated · 2012 · 3 generations on10descendants
  • K082775Prescision OptionSiemens Medical Solutions USA, Inc. · 2009 · cites it directly9descendants
  • K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · 2 generations on7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−6K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−5K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−4K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−3K013673HI-Art SystemTomo Therapy, Inc.2002—
−2K033347Modification to: HI-Art SystemTomotherapy Incorporated2003—
−1K042739HI-Art SystemTomotherapy Incorporated2004—
0 (this device)K060912Modification to: HI-Art SystemTomotherapy Incorporated2006—
+1K082005Modification to HI-Art Treatment SystemTomotherapy Incorporated20087
+2K112776Tomotherapy Treatment System Fast OptimizerTomotherapy Incorporated2011—
+3K121934Tomotherapy Treatment SystemAccuray Incorporated20128
+1K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
+2K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20094

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199010
199210
199310
199430
199610
199840
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200930
201010
201130
201220
201310
201420
201510
201620
201720
201820
202020
202110
202310
FIG. 1 — Lineage of K060912, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042739HI-Art System
Summary page 1
…System Predicate Device: TomoTherapy Hi-Art System K042739 INTENDED USE The TomoTherapy HI-ART System" is intended to be used as an integrated system for the planning an…
Tomotherapy Incorporated2004—

What descends from it? (24)

See all 24

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K082005Modification to HI-Art Treatment SystemTomotherapy Incorporated200813
K082775Prescision OptionSiemens Medical Solutions USA, Inc.20099

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 10, 2009Not classifiedZ-2243-2009Software designTomoTherapy received one customer report indicating that after starting a patient procedure from the Status Console, the radiation on light and the audible indicator on the Status Console came on as expected, however the Operator Station…
Oct 14, 2008Not classifiedZ-0628-2009Software designTomoTherapy Incorporated identified a potential issue with the Hi-Art system during the course of ongoing testing. The Operator Station Calibration panel provides access to view and modify machine specific configuration settings. Access…
Sep 18, 2008Not classifiedZ-0339-2009Software designTomoTherapy Inc. discovered an issue on the Hi-Art system during the course of ongoing testing. It has been identified that treatment plans involving very small structures (volume <2cc), reading in a high dose gradient region, may exhibit…
Feb 22, 2008Not classifiedZ-1615-2008Software designRegistration adjustment field numbers may appear on printed pages of the Register Tab as nonsense characters or values different than the numbers shown on the Operator Station (OS) display. Setup will move the couch to the correct offset…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • 3 of the 22 ancestors have a recall of any class.
  • Of its descendants, K082005 (Tomotherapy Incorporated, 2008) has gone furthest, with 13 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The TomoTherapy 11I-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue. The HI-ART System's planning station or operator station is intended to be used by the physician/oncologist to prescribe a radiation therapy…

Read the full statement (220 words)

The TomoTherapy 11I-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

The HI-ART System's planning station or operator station is intended to be used by the physician/oncologist to prescribe a radiation therapy plan for a particular patient. The HI-ART System then calculates the treatment plan which the physician reviews and approves.

The [II-ART system's operator station and status console is then intended to be used by the therapist to select and implement the patient's treatment plan. The treatment process will begin by performing a TomolmageTM (MVCT) scan (a CT using the onboard linear accelerator as the radiation source). This Tomolmage (MVCT) will confirm that the patient's position is correct for the radiation therapy as well as assist in patient re-positioning when necessary. The Tomohnage (MVCT image) is not for diagnostic use.

When patient positioning is complete, the 1ll-ART System is then intended to be used by the therapist to treat the patient using the selected treatment plan. The l1-ART System delivers the radiation therapy treatment, stereotactic radiotherapy, or stereotactic radiosurgery in accordance with the physician approved plan delivered in a helical tomographic pattern.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060912: Modification to: HI-Art System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K060912

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026