510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Syngo RT Therapist Connect

K090683510(k) SpecialCleared May 8, 200953-day reviewClass II21 CFR 892.5050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on May 8, 2009 as substantially equivalent to K072485 (Siemens Medical Solutions USA, Inc., 2007) and 2 other predicates after a 53-day review. Its family tree reaches back 10 generations to 1992. 7 later devices descend from it. It has been recalled 4 times.

Decided
May 8, 2009
Review
53 days
Class
Class II
Product code
IYE
Recalls
4
Predicates
3

Its longest line of descent runs back 10 generations to K914698 (1992). 46 ancestors in all, the oldest K862339 (1986). 7 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2005: 1 other ancestor2004: 4 other ancestors1999: 1 other ancestor2003: 2 other ancestors2001: 4 other ancestors1997: 1 other ancestor1995: 2 other ancestors1990: 2 other ancestors1988: 1 other ancestor1986: 1 other ancestor2000: 1 other ancestor1992: 1 other ancestor1991: 2 other ancestors2002: 1 other ancestor1998: 4 other ancestors1996: 1 other ancestor1994: 3 other ancestors1993: 1 other ancestor1989: 1 other ancestor2010: 1 descendant cleared2013: 1 descendant cleared2011: 1 descendant cleared2014: 2 descendants cleared2012: 1 descendant cleared2018: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK060226 · Oncor Expression · Siemens Medical Solutions USA, Inc. · 2006 · recalledK072485 · Artiste MV SA · Siemens Medical Solutions USA, Inc. · 2007 · recalledK082775 · Prescision Option · Siemens Medical Solutions USA, Inc. · 20091990200020102020
  1. Root · 10 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 6 generations between
  3. 3 generations backK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  4. 2 generations backK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · Recalled
  5. 1 generation backK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  6. This deviceK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009
  7. Descendant · cites it directly · 5 of its ownK102671 Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010
  8. Descendant · 2 generations on · 4 of its ownK103606 Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  9. Descendant · cites it directly · 3 of its ownK123812 Coherence RT Therapist Connect Workspace, V2.3, Aria Connectivity…Siemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. Descendant · 3 generations on · 3 of its ownK121295 Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc. · 2012 · Recalled
  11. Descendant · 2 generations on · 2 of its ownK132935 Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace…Siemens Medical Solutions USA, Inc. · 2014
Show all 10 generations
  1. Root · 10 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 9 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 8 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 7 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 6 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
  6. 5 generations backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
  7. 4 generations backK031764 Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  8. 3 generations backK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  9. 2 generations backK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · Recalled
  10. 1 generation backK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  11. This deviceK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12012: 12013: 12014: 22018: 1
2010high 2 (2014) · low 0 (2015)2026

Most influential descendants

Showing 5 of 7 · See all
  • K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010 · cites it directly5descendants
  • K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011 · 2 generations on4descendants
  • K123812Coherence RT Therapist Connect Workspace, V2.3, Aria Connectivity Option for Siemens…Siemens Medical Solutions USA, Inc. · 2013 · cites it directly3descendants
  • K121295Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc. · 2012 · 3 generations on3descendants
  • K132935Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace, V4.3.1 Upgrade Kit…Siemens Medical Solutions USA, Inc. · 2014 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−9K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−8K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−7K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−6K013112Focal SimComputerized Medical Systems, Inc.200110
−5K022036Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc.2002—
−4K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200317
−3K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.200672
−2K072485Artiste MV SASiemens Medical Solutions USA, Inc.200724
−1K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
0 (this device)K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.2009—
+1K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.2010—
+2K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.20117
+1K123812Coherence RT Therapist Connect Workspace, V2.3, Aria Connectivity Option for…Siemens Medical Solutions USA, Inc.20139
+3K121295Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc.20125
+2K132935Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace, V4.3.1…Siemens Medical Solutions USA, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
198910
199020
199120
199210
199310
199430
199520
199610
199710
199840
199910
200010
200140
200210
200320
200440
200510
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201210
201310
201420
201810
FIG. 1 — Lineage of K090683, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072485Artiste MV SA
Summary page 1
…P~ro~duct ;IoC C)(le.;ravlt~c ARTISTETM Solution (aka K072485 December 27, ARTISTE MV) with the 2007…
Siemens Medical Solutions USA, Inc.200724
*K060226Oncor Expression
Summary page 1
…2007 syn go®RT Therapist workspace ONCORTm Expression with K060226 March 15, COHERENCETm RT Therapist 2006…
Siemens Medical Solutions USA, Inc.200672
*K082775Prescision Option
Summary page 1
…OHERENCETm RT Therapist 2006 works pace PreScisionTM Option K082775 February 19 2009 Siemens Medical So…
Siemens Medical Solutions USA, Inc.2009—

Also cleared against K072485: 5 other devices →

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.20105
K123812Coherence RT Therapist Connect Workspace, V2.3, Aria Connectivity Option for Siemens…Siemens Medical Solutions USA, Inc.20133

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 26, 2016Class IIZ-2878-2016Software designSoftware patch installation to address several safety issues.
Aug 5, 2015Class IIZ-2812-2015Software design (manufacturing process)A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
Feb 14, 2012Not classifiedZ-1270-2012Device DesignImages used in treatment planning may belong to another patient. Risk of treatment to wrong area. Software update to upgrade Siemens installed base to the latest release of MOSAIQ version 2.00X3 SP5, for RT Therapist version 2.1A
Jun 29, 2010Not classifiedZ-1523-2011OtherA software problem that has the potential for data loss during the transfer of treatment records to the OIS, which may not be recorded in MOSAIQ, and subsequently may lead to mistreatment.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 1.1 years on median.
  • 20 of the 46 ancestors have a recall of any class.
  • Of its descendants, K102671 (Siemens Medical Solutions USA, Inc., 2010) has gone furthest, with 5 descendants of its own.
  • 5 other devices also cite its predicate K072485.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer: The syngo®Suite for Oncology Workspaces: The syngo®workspaces includes a number of syngo®based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also.permit patient data…

Read the full statement (138 words)

The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer:

The syngo®Suite for Oncology Workspaces:

The syngo®workspaces includes a number of syngo®based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also.permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.

The svngo®RT Therapist Connect Workspace:

The syngo®RT Therapist Connect Workspace contains software applications that permits patient selection/setup, patient positioning verification, treatment delivery/verification, and treatmint recording. The syngO®RT Therapist Connect Workspace is an option when third party OIS, Treatment Planning Systems and/or PACS devices conforming to the DICOM® Standard are intended to be used in conjunction with the Siemens branded Linear Accelerator system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090683: Syngo RT Therapist Connect.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K090683

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026