510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Codman Eds 3 CSF External Drainage System, Models 82-1730, 82-1731, 82-1732, 82-1733, 82-1735, 82-1738

K061568510(k) SpecialCleared Sep 29, 2006115-day reviewClass II21 CFR 882.5550Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Codman & Shurtleff, Inc. · JXG Shunt, Central Nervous System And Components

Cleared on Sep 29, 2006 as substantially equivalent to K902257 (Codman & Shurtleff, Inc., 1990) and 2 other predicates after a 115-day review. Its family tree reaches back 1 generation to 1990. 4 later devices descend from it. It has been recalled 3 times.

Decided
Sep 29, 2006
Review
115 days
Class
Class II
Product code
JXG
Recalls
3
Predicates
3

Its longest line of descent runs back 1 generation to K902257 (1990). 3 ancestors in all. 4 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1991: 1 other ancestor2016: 1 descendant cleared2013: 1 descendant cleared2015: 1 descendant cleared2017: 1 descendant clearedK902257 · Codman* External Drainage System II (Eds-II) · Codman & Shurtleff, Inc. · 19901990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 2 descendants K902257 · Codman* External Drainage System II (Eds-II) · Codman & Shurtleff, Inc. · 19900
  1. Root · 1 generation backK902257 Codman* External Drainage System II (Eds-II)Codman & Shurtleff, Inc. · 1990
  2. This deviceK061568 Codman Eds 3 CSF External Drainage System, Models 82-1730, 82-1731…Codman & Shurtleff, Inc. · 2006
  3. Descendant · cites it directly · 1 of its ownK162437 Codman EDS3 CSF External Drainage SystemCodman & Shurtleff, Inc. · 2016 · Recalled
  4. Descendant · cites it directly · 1 of its ownK121248 Liquoguard CSF Liquoguard CSF Tubing Set (1500MM) Liquoguard CSF…Moller Medical GmbH · 2013
  5. Descendant · 2 generations on · 0 of its ownK172537 Codman EDS 3 CSF External Drainage System, Codman EDS 3 CSF External…Codman & Shurtleff, Inc. · 2017
  6. Descendant · 2 generations on · 0 of its ownK150898 LiquoGuard 7, LiquoGuard 7 Drainage Set, Lenght 2000mmMoller Medical GmbH · 2015

Most influential descendants

Showing 4 of 4 · See all
  • K162437Codman EDS3 CSF External Drainage SystemCodman & Shurtleff, Inc. · 2016 · cites it directly1descendant
  • K121248Liquoguard CSF Liquoguard CSF Tubing Set (1500MM) Liquoguard CSF Tubing Set (2000MM)…Moller Medical GmbH · 2013 · cites it directly1descendant
  • K172537Codman EDS 3 CSF External Drainage System, Codman EDS 3 CSF External Drainage System…Codman & Shurtleff, Inc. · 2017 · 2 generations on0descendants
  • K150898LiquoGuard 7, LiquoGuard 7 Drainage Set, Lenght 2000mmMoller Medical GmbH · 2015 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K902257Codman* External Drainage System II (Eds-II)Codman & Shurtleff, Inc.1990—
0 (this device)K061568Codman Eds 3 CSF External Drainage System, Models 82-1730, 82-1731, 82-1732…Codman & Shurtleff, Inc.2006—
+1K162437Codman EDS3 CSF External Drainage SystemCodman & Shurtleff, Inc.20163
+1K121248Liquoguard CSF Liquoguard CSF Tubing Set (1500MM) Liquoguard CSF Tubing Set…Moller Medical GmbH2013—
+2K172537Codman EDS 3 CSF External Drainage System, Codman EDS 3 CSF External Drainage…Codman & Shurtleff, Inc.2017—
+2K150898LiquoGuard 7, LiquoGuard 7 Drainage Set, Lenght 2000mmMoller Medical GmbH2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201510
201610
201710
FIG. 1 — Lineage of K061568, 1990–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K902257Codman* External Drainage System II (Eds-II)
Summary page 1
…Common Name: External CSF Drainage System Classification Name: Central nervous system fluid shunt and components Predicate Device: K902257 External Drainage System 11 K910938 External Drainage System 1I K9540…
Codman & Shurtleff, Inc.1990—
*K910938Codman External Drainage System II (Eds II)
Summary page 1
…ication Name: Central nervous system fluid shunt and components Predicate Device: K902257 External Drainage System 11 K910938 External Drainage System 1I K954021 External Drainage System 1I Device Description: The EDS 3 syst…
Codman & Shurtleff, Inc.1991—
*K954021External Drainage System (Eds and Eds II)
Summary page 1
…onents Predicate Device: K902257 External Drainage System 11 K910938 External Drainage System 1I K954021 External Drainage System 1I Device Description: The EDS 3 system is designed to collect cerebral spinal fluid (CSF) and other…
Johnson & Johnson Professionals, Inc.19951

Also cleared against K902257: 3 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162437Codman EDS3 CSF External Drainage SystemCodman & Shurtleff, Inc.20161
K121248Liquoguard CSF Liquoguard CSF Tubing Set (1500MM) Liquoguard CSF Tubing Set (2000MM)…Moller Medical GmbH20131

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 15, 2014Class IIZ-0089-2015Device Designtubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
Sep 15, 2014Class IIZ-0090-2015Device Designtubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
Sep 15, 2014Class IIZ-0091-2015Device Designtubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K162437 (Codman & Shurtleff, Inc., 2016) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K902257.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

Use of the CODMAN EDS 3 CSF External Drainage System is indicated for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061568: Codman Eds 3 CSF External Drainage System, Models 82-1730, 82-1731, 82-1732, 82-1733, 82-1735, 82-1738.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K061568

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026