Becker External Drainage and Monitoring System

K984053510(k) TraditionalCleared Feb 4, 199990-day reviewClass II21 CFR 882.5550Recalls 6 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic PS Medical · JXG Shunt, Central Nervous System And Components

Cleared on Feb 4, 1999 as substantially equivalent to K902257 (Codman & Shurtleff, Inc., 1990) and 1 other predicate after a 90-day review. Its family tree reaches back 1 generation to 1990. 2 of the 16 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 6 times, including 5 Class I recalls.

Decided
Feb 4, 1999
Review
90 days
Class
Class II
Product code
JXG
Recalls
6 (5 Class I)
Predicates
2

Its longest line of descent runs back 1 generation to K902257 (1990). 2 ancestors in all. 13 later devices descend from it, 5 citing it directly. 2 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1991: 1 other ancestor2016: 1 descendant cleared2018: 2 descendants cleared2017: 2 descendants cleared2020: 2 descendants cleared, 1 with a Class I recall12000: 1 descendant cleared2019: 1 descendant cleared2025: 2 descendants cleared2022: 1 descendant cleared2023: 1 descendant clearedK902257 · Codman* External Drainage System II (Eds-II) · Codman & Shurtleff, Inc. · 19901990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 5 descendants, 1 with a Class I recall12 generations forward: 6 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K902257 · Codman* External Drainage System II (Eds-II) · Codman & Shurtleff, Inc. · 19900
  1. Root · 1 generation backK902257 Codman* External Drainage System II (Eds-II)Codman & Shurtleff, Inc. · 1990
  2. This deviceK984053 Becker External Drainage and Monitoring SystemMedtronic PS Medical · 1999
  3. Descendant · cites it directly · 4 of its ownK171880 IRRAflow CNS SystemIrras, Inc. · 2018
  4. Descendant · cites it directly · 4 of its ownK161605 Smart External Drain (SED) SystemAqueduct Critical Care, Inc. · 2016
  5. Descendant · 2 generations on · 3 of its ownK192289 IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Ltd. · 2019
  6. Descendant · 2 generations on · 3 of its ownK171586 Smart External Drain (SED) SystemAqueduct Critical Care, Inc. · 2017
  7. Descendant · 2 generations on · 2 of its ownK200807 IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Inc. · 2020

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2000: 12016: 12017: 22018: 22019: 12020: 22022: 12023: 12025: 2
2000high 2 (2017) · low 0 (2001)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K171880IRRAflow CNS SystemIrras, Inc. · 2018 · cites it directly4descendants
  • K161605Smart External Drain (SED) SystemAqueduct Critical Care, Inc. · 2016 · cites it directly4descendants
  • K192289IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Ltd. · 2019 · 2 generations on3descendants
  • K171586Smart External Drain (SED) SystemAqueduct Critical Care, Inc. · 2017 · 2 generations on3descendants
  • K200807IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Inc. · 2020 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K902257Codman* External Drainage System II (Eds-II)Codman & Shurtleff, Inc.1990—
0 (this device)K984053Becker External Drainage and Monitoring SystemMedtronic PS Medical1999—
+1K171880IRRAflow CNS SystemIrras, Inc.2018—
+1K161605Smart External Drain (SED) SystemAqueduct Critical Care, Inc.2016—
+2K192289IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Ltd.2019—
+2K171586Smart External Drain (SED) SystemAqueduct Critical Care, Inc.2017—
+2K200807IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control UnitIrras USA, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
201610
201720
201820
201910
202021
202210
202310
202520
FIG. 1 — Lineage of K984053, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K902257Codman* External Drainage System II (Eds-II)
Summary page 1
…ronic PS Medical has, with input from neurosurgeons, developed a device design, which is considered to be equivalent to the predicate Codman External Drainage System [l (K902257, The Medtronic PS Medical Becker EDMS is a complete closed system irfdicated for the draining and monitoring of CSF flow from the…
Codman & Shurtleff, Inc.1990—
*K910938Codman External Drainage System II (Eds II)
Our OCR of the scanned summary read K920938, FDA’s number for VJW2919 Pneumatic Compression Upgrade (Varian Canada, Inc.). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Codman External Drainage System II”.
Codman & Shurtleff, Inc.1991—

Also cleared against K902257: 3 other devices →

What descends from it? (13)

See all 13

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171880IRRAflow CNS SystemIrras, Inc.20184
K161605Smart External Drain (SED) SystemAqueduct Critical Care, Inc.20164
K172759Smart External Drain (SED) SystemAqueduct Critical Care, Inc.20172
K200456Becker External Drainage and Monitoring System Class IMedtronic, Inc.20201
K992226XDS External CSF Drainage and Monitoring SystemRadionics, Inc.20000

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2023Class IIZ-2412-2023Labeling Change ControlProduct labeling contains incorrect expiration date
Jun 9, 2014Class IZ-1920-2014Device DesignMedtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
Jun 9, 2014Class IZ-1923-2014Device DesignMedtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
Jun 9, 2014Class IZ-1921-2014Device DesignMedtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
Jun 9, 2014Class IZ-1922-2014Device DesignMedtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.
Jun 9, 2014Class IZ-1924-2014Device DesignMedtronic Neurosurgery is recalling the Medtronic Duet External Drainage and Monitoring Systems because the patient line tubing in the Duet External Drainage and Monitoring Systems may become disconnected from the patient line stopcock.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K200456Becker External Drainage and Monitoring SystemMedtronic, Inc. · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K171880 (Irras, Inc., 2018) has gone furthest, with 4 descendants of its own.
  • 3 other devices also cite its predicate K902257.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to reduce intracranial pressure (ICP), e.g. pre-intra- or postoperative; monitor CSF chemistry, cytology and physiology; provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with severe head injury; subarachnoid…

Read the full statement (105 words)

Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to reduce intracranial pressure (ICP), e.g. pre-intra- or postoperative; monitor CSF chemistry, cytology and physiology; provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with severe head injury; subarachnoid hemorrhage graded lii, IV or V preoperatively; Reyes syndrome or similar encephalopathies; hydrocephalus; intracranial hemorrhage or miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre-and postoperatively for space occupying lesions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K984053: Becker External Drainage and Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K984053

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026