510(k) Report

510(k) · Cleared

Cleared

Inomax DS (Delivery System), Model 10003

K070867510(k) SpecialCleared Apr 27, 200729-day reviewClass II21 CFR 868.5165

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ino Therapeutics, LLC · MRN Apparatus, Nitric Oxide Delivery

Cleared on Apr 27, 2007 as substantially equivalent to K061901 (Ino Therapeutics, 2006) after a 29-day review. Its family tree reaches back 4 generations to 2001. 15 later devices descend from it.

Decided
Apr 27, 2007
Review
29 days
Class
Class II
Product code
MRN
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K003665 (2001). 7 ancestors in all, the oldest K000186 (2000). 15 later devices descend from it, 10 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2000: 2 other ancestors2008: 2 descendants cleared2007: 1 descendant cleared2011: 2 descendants cleared2012: 2 descendants cleared2013: 2 descendants cleared2009: 1 descendant cleared2015: 1 descendant cleared2023: 2 descendants cleared2020: 1 descendant cleared2025: 1 descendant clearedK003665 · Fadassi Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K011874 · Fadasis Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K052663 · Ino Therapeutics Inoblender, Model 10004 · Ino Therapeutics · 2005K061901 · Ino Therapeutics Inomax DS (Delivery System), Model 10003 · Ino Therapeutics · 2006 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 3 other ancestors 1 generation forward: 10 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K061901 · Ino Therapeutics Inomax DS (Delivery System), Model 10003 · Ino Therapeutics · 2006 · recalledK052663 · Ino Therapeutics Inoblender, Model 10004 · Ino Therapeutics · 2005K011874 · Fadasis Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K003665 · Fadassi Medical FM-1 NO Blender · Fadasis Medical, Inc. · 20010
  1. Root · 4 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 3 generations backK011874 Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  3. 2 generations backK052663 Ino Therapeutics Inoblender, Model 10004Ino Therapeutics · 2005
  4. 1 generation backK061901 Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2006 · Recalled
  5. This deviceK070867 Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC · 2007
  6. Descendant · cites it directly · 14 of its ownK071516 Inomax DSIno Therapeutics, LLC · 2007
  7. Descendant · cites it directly · 11 of its ownK090958 Inomax DS (Delivery System)Ino Therapeutics, LLC · 2009
  8. Descendant · cites it directly · 11 of its ownK081691 Inomax DS (Delivery System)Ino Therapeutics, LLC · 2008
  9. Descendant · cites it directly · 11 of its ownK080484 Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2008
  10. Descendant · cites it directly · 9 of its ownK110344 Inomax Dsir (Delivery System)Ino Therapeutics · 2011

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2007: 12008: 22009: 12011: 22012: 22013: 22015: 12020: 12023: 22025: 1
2007high 2 (2008) · low 0 (2010)2026

Most influential descendants

Showing 5 of 15 · See all
  • K071516Inomax DSIno Therapeutics, LLC · 2007 · cites it directly14descendants
  • K090958Inomax DS (Delivery System)Ino Therapeutics, LLC · 2009 · cites it directly11descendants
  • K081691Inomax DS (Delivery System)Ino Therapeutics, LLC · 2008 · cites it directly11descendants
  • K080484Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2008 · cites it directly11descendants
  • K110344Inomax Dsir (Delivery System)Ino Therapeutics · 2011 · cites it directly9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K003665Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−3K011874Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−2K052663Ino Therapeutics Inoblender, Model 10004Ino Therapeutics2005—
−1K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics20063
0 (this device)K070867Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC2007—
+1K071516Inomax DSIno Therapeutics, LLC2007—
+1K090958Inomax DS (Delivery System)Ino Therapeutics, LLC2009—
+1K081691Inomax DS (Delivery System)Ino Therapeutics, LLC2008—
+1K080484Inomax DS (Delivery System), Model 10003Ino Therapeutics2008—
+1K110344Inomax Dsir (Delivery System)Ino Therapeutics2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200820
200910
201120
201220
201320
201510
202010
202320
202510
FIG. 1 — Lineage of K070867, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRN device

  1. #1 K242374 ULSPIRA TS™ Nitric Oxide Therapy Systemscore 12.6 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003
Summary page 1
…MRQ Panel: Anesthesiology Predicate Device(s): ]N40 Therapeutics' INOmax DS (Delivery System) (primary) 5 10 (k) Number K061901 Datex-Ohmeda INOvent Delivery System (now GE Healthcare) 5 10(k) Number K974…
Ino Therapeutics20063
Kept off the family treeK974562Datex-Ohmeda INOvent Delivery System
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
Summary page 1
…5 10 (k) Number K061901 Datex-Ohmeda INOvent Delivery System (now GE Healthcare) 5 10(k) Number K974562 Patient Population: The targeted patient population is neonates. This is derived from the INOmax drug…

Also cleared against K061901: 13 other devices →

What descends from it? (15)

See all 15

10 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071516Inomax DSIno Therapeutics, LLC200714
K090958Inomax DS (Delivery System)Ino Therapeutics, LLC200911
K081691Inomax DS (Delivery System)Ino Therapeutics, LLC200811
K080484Inomax DS (Delivery System), Model 10003Ino Therapeutics200811
K110344Inomax Dsir (Delivery System)Ino Therapeutics20119
K110635Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria20119
K113272Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria20128
K121021Inomax DS (Delivery System)Ino Therapeutics20127
K130605Inomax Dsir (Delivery System)Ino Therapeutics20136
K131686Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria20135

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K071516 (Ino Therapeutics, LLC, 2007) has gone furthest, with 14 descendants of its own.
  • 13 other devices also cite its predicate K061901.

How does its category compare?

All MRN devices

MRN Apparatus, Nitric Oxide Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · MRN

2in 2025

2007: 22008: 22009: 12010: 22011: 22012: 32013: 22015: 12018: 12020: 32022: 12023: 42024: 22025: 2
2007high 4 (2023) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRN

566in 2023

2012: 117 days median2020: 84 days median2023: 566 days median
2007high 566 (2023) · low 84 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MRN

2in 2025

2010: 22014: 12023: 12025: 2, 2 Class I2026 (partial year): 2
2007high 2 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MRN

155in 2025

2007: 32008: 92009: 102010: 1552011: 552012: 242013: 322014: 792015: 1562016: 942017: 1062018: 1382019: 1442020: 1992021: 2212022: 1152023: 992024: 1382025: 155
2007high 221 (2021) · low 3 (2007)2025

MAUDE reports naming product code MRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—03
20082—09
20091—010
20102—2155
20112—055
20123117024
20132—032
20140—179
20151—0156
20160—094
20170—0106
20181—0138
20190—0144
20203840199
20210—0221
20221—0115
20234566199
20242—0138
20252—2 (2)155
20260—2—

What is it indicated for?

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be…

Read the full statement (207 words)

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired 02, NO 2, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K070867: Inomax DS (Delivery System), Model 10003.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K070867

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026