510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Inomax Dsir (Delivery System)

K131686510(k) TraditionalCleared Nov 29, 2013172-day reviewClass II21 CFR 868.5165Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ino Therapeutics/Ikaria · MRN Apparatus, Nitric Oxide Delivery

Cleared on Nov 29, 2013 as substantially equivalent to K061901 (Ino Therapeutics, 2006) and 12 other predicates after a 172-day review. Its family tree reaches back 11 generations to 2001. 5 later devices descend from it. It has been recalled 4 times.

Decided
Nov 29, 2013
Review
172 days
Class
Class II
Product code
MRN
Recalls
4
Predicates
13

Its longest line of descent runs back 11 generations to K003665 (2001). 19 ancestors in all, the oldest K000186 (2000). 5 later devices descend from it, 3 citing it directly.

  1. Root · 11 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 7 generations between
  3. 3 generations backK113272 Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria · 2012
  4. 2 generations backK121021 Inomax DS (Delivery System)Ino Therapeutics · 2012
  5. 1 generation backK130605 Inomax Dsir (Delivery System)Ino Therapeutics · 2013
  6. This deviceK131686 Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria · 2013
  7. Descendant · cites it directly · 3 of its ownK200389 INOmax DSIR PlusMallinckrodt Manufacturing, LLC · 2020
  8. Descendant · 2 generations on · 1 of its ownK212409 Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics · 2023
  9. Descendant · 3 generations on · 0 of its ownK242374 ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC · 2025
  10. Descendant · cites it directly · 0 of its ownK211153 Inomax DSIRMallinckrodt Manufacturing, LLC · 2023
  11. Descendant · cites it directly · 0 of its ownK143213 INOmax DSIR Plus MRIIno Therapeutics/Ikaria · 2015
Show all 11 generations
  1. Root · 11 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 10 generations backK011874 Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  3. 9 generations backK052663 Ino Therapeutics Inoblender, Model 10004Ino Therapeutics · 2005
  4. 8 generations backK061901 Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2006 · Recalled
  5. 7 generations backK070867 Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC · 2007
  6. 6 generations backK071516 Inomax DSIno Therapeutics, LLC · 2007
  7. 5 generations backK080484 Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2008
  8. 4 generations backK110635 Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria · 2011
  9. 3 generations backK113272 Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria · 2012
  10. 2 generations backK121021 Inomax DS (Delivery System)Ino Therapeutics · 2012
  11. 1 generation backK130605 Inomax Dsir (Delivery System)Ino Therapeutics · 2013
  12. This deviceK131686 Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria · 2013

Most influential descendants

Showing 5 of 5 · See all
  • K200389INOmax DSIR PlusMallinckrodt Manufacturing, LLC · 2020 · cites it directly3descendants
  • K212409Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics · 2023 · 2 generations on1descendant
  • K242374ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC · 2025 · 3 generations on0descendants
  • K211153Inomax DSIRMallinckrodt Manufacturing, LLC · 2023 · cites it directly0descendants
  • K143213INOmax DSIR Plus MRIIno Therapeutics/Ikaria · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K003665Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−10K011874Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−9K052663Ino Therapeutics Inoblender, Model 10004Ino Therapeutics2005—
−8K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics20063
−7K070867Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC2007—
−6K071516Inomax DSIno Therapeutics, LLC2007—
−5K080484Inomax DS (Delivery System), Model 10003Ino Therapeutics2008—
−4K110635Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria2011—
−3K113272Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria2012—
−2K121021Inomax DS (Delivery System)Ino Therapeutics2012—
−1K130605Inomax Dsir (Delivery System)Ino Therapeutics2013—
0 (this device)K131686Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria2013—
+1K200389INOmax DSIR PlusMallinckrodt Manufacturing, LLC2020—
+2K212409Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics2023—
+3K242374ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC2025—
+1K211153Inomax DSIRMallinckrodt Manufacturing, LLC2023—
+1K143213INOmax DSIR Plus MRIIno Therapeutics/Ikaria2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200310
200810
200910
201020
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
202010
202320
202510
FIG. 1 — Lineage of K131686, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRN device

  1. #1 K242374 ULSPIRA TS™ Nitric Oxide Therapy Systemscore 12.6 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (13)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003
Summary page 1
…Device Classification: Class 11 -21 CFR 868.5165 Product Code: MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635,…
Ino Therapeutics20063
*K070867Inomax DS (Delivery System), Model 10003
Summary page 1
…assification: Class 11 -21 CFR 868.5165 Product Code: MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635,…
Ino Therapeutics, LLC2007—
*K071516Inomax DS
Summary page 1
…ion: Class 11 -21 CFR 868.5165 Product Code: MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635,…
Ino Therapeutics, LLC2007—
*K080484Inomax DS (Delivery System), Model 10003
Summary page 1
…Class 11 -21 CFR 868.5165 Product Code: MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635,…
Ino Therapeutics2008—
*K081691Inomax DS (Delivery System)
Summary page 1
…11 -21 CFR 868.5165 Product Code: MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, K…
Ino Therapeutics, LLC2008—
*K090958Inomax DS (Delivery System)
Summary page 1
…MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Devic…
Ino Therapeutics, LLC2009—
*K092545Inomax DS (Delivery System), Model 10003
Summary page 1
…MRN (Primary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Device Th…
Ino Therapeutics, LLC2010—
*K093922Inomax DS
Summary page 1
…mary), MRO, MRP, MRQ Predicate Device(s) K06 190 1, K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Device The INOmax…
Ino Therapeutics20101
*K110344Inomax Dsir (Delivery System)Ino Therapeutics2011—
*K110635Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria2011—
*K113272Inomax DS (Delivery Systems
Summary page 1
…K070867, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Device The INOmax DSIR®I uses a "dual-channel' design to ensure the safe…
Ino Therapeutics, LLC/Dba Ikaria2012—
*K121021Inomax DS (Delivery System)
Summary page 1
…, K07 1516, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Device The INOmax DSIR®I uses a "dual-channel' design to ensure the safe…
Ino Therapeutics2012—
*K130605Inomax Dsir (Delivery System)
Summary page 1
…, K080484, K08 169 1, K090958, K092545, K093922, KI 10344, KI 10635, KI 113272, K12102 1, K130605 Description of Device The INOmax DSIR®I uses a "dual-channel' design to ensure the safe delivery…
Ino Therapeutics2013—

Also cleared against K061901: 13 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200389INOmax DSIR PlusMallinckrodt Manufacturing, LLC20203
K211153Inomax DSIRMallinckrodt Manufacturing, LLC20230
K143213INOmax DSIR Plus MRIIno Therapeutics/Ikaria20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 9, 2018Class IIZ-2733-2018Process controlINOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020).
Nov 21, 2016Class IIZ-1076-2017Equipment maintenanceMallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.
Oct 17, 2014Class IIZ-0575-2015Process change controlAn issue has been identified in the INOmax DSIR system that could result in monitored Nitric Oxide (NO) concentration reporting lower than expected. This issue only pertains to those devices manufactured using a specific version of the…
Mar 27, 2014Class IIZ-1528-2014Software Design ChangeAn anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 5 companies; a new generation was cleared every 7 months on median.
  • 3 of the 19 ancestors have a recall of any class.
  • Of its descendants, K200389 (Mallinckrodt Manufacturing, LLC, 2020) has gone furthest, with 3 descendants of its own.
  • 13 other devices also cite its predicate K061901.

How does its category compare?

All MRN devices

MRN Apparatus, Nitric Oxide Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MRN

2in 2025

2007: 22008: 22009: 12010: 22011: 22012: 32013: 22015: 12018: 12020: 32022: 12023: 42024: 22025: 2
2007high 4 (2023) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRN

566in 2023

2012: 117 days median2020: 84 days median2023: 566 days median
2007high 566 (2023) · low 84 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MRN

2in 2025

2010: 22014: 12023: 12025: 2, 2 Class I2026 (partial year): 2
2007high 2 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MRN

155in 2025

2007: 32008: 92009: 102010: 1552011: 552012: 242013: 322014: 792015: 1562016: 942017: 1062018: 1382019: 1442020: 1992021: 2212022: 1152023: 992024: 1382025: 155
2007high 221 (2021) · low 3 (2007)2025

MAUDE reports naming product code MRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—03
20082—09
20091—010
20102—2155
20112—055
20123117024
20132—032
20140—179
20151—0156
20160—094
20170—0106
20181—0138
20190—0144
20203840199
20210—0221
20221—0115
20234566199
20242—0138
20252—2 (2)155
20260—2—

What is it indicated for?

The INOmax* DS delivery system delivers INOMvAXO (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be…

Read the full statement (204 words)

The INOmax* DS delivery system delivers INOMvAXO (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmaxDS Orovides continuous integrated monitoring of inspired 02, N02, and NO, and a comprehensive alarm system.

The iNOmaxv DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmaxo DS includes a backup NO delivery capability that provides a fixe'dflow of 250 niL/min of NO which along with user supplied 10 Llnin of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the lNOblendero for backup.

The target patient population is controlled by the drug labeling for INOMAXO and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131686: Inomax Dsir (Delivery System).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131686

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026